A fast method for microplastic detection is thermal extraction desorption‐gas chromatography/mass spectrometry (TED‐GC/MS), which uses polymer‐specific thermal decomposition products as marker compounds to determine polymer mass contents in environmental samples. So far, matrix impacts of different environmental matrices on TED‐GC/MS performance had not yet been assessed systematically. Therefore, three solid freshwater matrices representing different aquatic bodies with varying organic matter contents were spiked with a total of eight polymers. Additionally, for the first time, the two biodegradable polymers polybutylene adipate terephthalate (PBAT) and polylactide (PLA) were analysed using TED‐GC/MS. The methodological focus of this work was on detectability, quality of signal formation as well as realisation of quantification procedures and determination of the limit of detection (LOD) values. Overall, TED‐GC/MS allowed the unambiguous detection of the environmentally most relevant polymers analysed, even at low mass contents: 0.02 wt% for polystyrene (PS), 0.04 wt% for the tyre component styrene butadiene rubber (SBR) and 0.2 wt% for polypropylene (PP), polyethylene (PE) and PBAT. Further, all obtained LOD values were increased in all matrices compared to the neat polymer without matrix. The LOD of the standard polymers were increased similarly (PS: 0.21–0.34 μg, SBR: 0.27–0.38 μg, PP: 0.32–0.36 μg, PMMA: 0.64–1.30 μg, PET: 0.90–1.37 μg, PE: 3.80–6.99 μg) and their decompositions by radical scission processes were not significantly influenced by the matrices. In contrast, matrixspecific LOD increases of both biodegradable polymers PBAT (LOD: 1.41–7.18 μg) and PLA (0.84–20.46 μg) were observed, probably due to their hetero‐functional character and interactions with the matrices. In conclusion, the TED‐GC/MS performance is not solely determined by the type of the polymers but also by the composition of the matrix.
Introduction to microplastics, definition, relevance, analytical approaches & challenges. TED-GC/MS working principle, features & performance. Optimisation in TED-GC/MS methodology, new validation data. Replacement of Internal Standard, outlook towards upcoming publication and application within EU´s revised drinking water directive.
Die Thermo Extraktion Desorption Gas Chromatograpie Massen Spektrometrie (TED-GC/MS) ist ein schnelles thermoanalytisches Verfahren zur Bestimmung von Gehalten an Mikroplastik (MP). Die Kombination aus teils schwerflüchtigen Analyten und Kältebrücken führt zu Memory-Effekten, denen u.a. durch häufige Blindwertmessungen begegnet werden muss. Reduzierte Probedurchsätze sind die Folge. Um Messergebnisse abzusichern und den Gerätestatus bei gleichzeitig hoher Messleistung überwachen zu können, wurde die sequentielle Standardmethode überarbeitet.
Deuterated polystyrene (D-PS) improves microplastic detection with TED-GC / MS, among other things through higher reproducibility. It is inexpensive (0.1 ct / analysis) and can be dosed quickly and reproducibly thanks to its good solubility (e.g. in toluene). However, certain matrices lead to an exchange of the deuterium with hydrogen. The poster shows the measuring principle, in which matrices H/D-exchange takes place, the results of kinetic studies and ways of preventing or circumventing the H/D-exchange.
Deuteriertes Polystyrol (D-PS) verbessert die Mikroplastikdetektion mit der TED-GC/MS, u.a. durch höhere Reproduzierbarkeit. Es ist kostengünstig (0,1 ct/Analyse) und durch die gute Löslichkeit (z.B. in Toluol) schnell und reproduzierbar zu dosieren. Bestimmte Matrices führen jedoch zu einem Austausch des Deuteriums durchWasserstoff. Das Poster zeigt das Messprinzip, bei welchen Matrices H/D-Austausch erfolgt, Ergebnisse von Kinetikuntersuchungen sowie Möglichkeiten, den Austausch zu verhindern oder zu umgehen.
Die große Herausforderung in der Detektion von Mikroplastik (MP)‑Partikeln (1-1000 µm) besteht darin, sie neben Millionen von natürlichen Partikeln in Umweltproben nachzuweisen. Zum Aufspüren von Senken, Eintragspfaden und für Monitoring-Kampagnen ist ein schnelles, automatisiertes Verfahren unerlässlich. Ein solches Verfahren wurde an der Bundesanstalt für Materialforschung und -prüfung (BAM) mit der ThermoExtraktion/Desorption-Gaschromatographie-Massenspektrometrie (TED-GC/MS) in den letzten Jahren entwickelt.
Fast quantitative detection of microplastics using TED-GC/MS, an innovative thermoanalytical method
(2021)
In the talk, the BAM is introduced and the motivation for our engagement in the field of microplastic analysis is explained. The challenges in the analysis of microplastics are shown and how we deal with them. The working principle of TED-GC/MS and its performance parameters are presented, as well as the procedure for identifying and quantifying polymers in environmental samples. Interesting results are shown using selected examples. Finally, BAM's microplastic reference materials are highlighted.
Tire and road wear particles (TRWPs) are heteroagglomerates of tire rubber and other particles deposited on the road surface and one of the main contributors to non-exhaust emissions of automobile traffic. In this study, samples from road environments were analyzed for their TRWP contents and concentra- tions of eight organic tire constituents. TRWP concentrations were determined by quantifying Zn in the density fraction < 1.9 g/cm ³and by thermal extraction desorption-gas chromatography-mass spectrometry (TED-GC/MS) and the concentrations ranged from 3.7 to 480 mg TRWP/g. Strong and statistically signif- icant correlations with TRWPs were found for 2-hydroxybenzothiazole and 2-aminobenzothiazole, indi- cating that these substances may be suitable markers of TRWPs. The mass distribution of TRWPs in road dust suggests that the main mass fraction formed on roads consists of coarse particles ( > 100 μm). Data for a sedimentation basin indicate that the fine fraction ( < 50 μm) is preferentially transported by road runoffinto receiving waters. The size distribution and density data of TRWP gathered by three different quantitation approaches also suggest that aging of TRWPs leads to changes in their particle density. An improved understanding of the dynamics of TRWP properties is essential to assess the distribution and dissipation of this contaminant of emerging concern in the environment.
There is growing interest in quantifying microplastics in environmental samples. This application note presents a thermal extraction desorption-gas chromatography/mass spectrometry (TED-GC/MS) method that is well suited to automation and increased sample throughput. The method is also able to detect all particle sizes in the sample as long as the limit of detection (LOD) is reached and allows analysis of larger samples of 15 to 25 mg or more. Samples were decomposed by thermogravimetric analysis (TGA), and the gaseous decomposition products were trapped on a solid-phase sorbent, followed by thermal desorption‑gas chromatography/mass spectrometry (TD-GC/MS) using an Agilent 5977B GC/MSD coupled to an Agilent 7890B GC. Target microplastic particle (MP) polymers were identified in environmental samples including surface water, finished compost, house dust, and drinking water. Quantification of MP polymers in environmental samples provided LODs of 0.06 to 2.2 μg, allowing the detection of MPs in trace amounts with sample weights of up to 1 g. Method repeatability was adequate for reliable quantification with RSDs of approximately 6 to 12%.
The TED-GC-MS analysis is a two-step method. A sample is first decomposed in a thermogravimetric analyzer (TGA) and the gaseous decomposition products are then trapped on a solid-phase adsorber. Subsequently, the solid-phase adsorber is analyzed with thermal desorption gas chromatography mass spectrometry (TDU-GC-MS). This method is ideally suited for the analysis of polymers and their degradation processes. Here, a new entirely automated System is introduced which enables high sample throughput and reproducible automated fractioned collection of decomposition products. Strengths and limitations of the system configuration are elaborated via three examples focused on practical challenges in materials analysis and identification: i) separate analysis of the components of a wood-plastic-composite material, ii) quantitative determination of weight concentration of the constituents of a polymer blend and iii) quantitative analysis of model samples of microplastics in suspended particulate matter.