Hepcidin-25 concentrations measured by various methods differ considerably, complicating interpretation. Here, a previously identified plasma-based candidate secondary reference material (csRM) was modified into a serum-based two-leveled sRM. We validated its functionality to increase the equivalence between methods for international standardization. We applied technical procedures developed by the International Consortium for Harmonization of Clinical Laboratory Results. The sRM, consisting of lyophilized serum with cryolyoprotectant, appeared commutable among nine different measurement procedures using 16 native human serum samples in a first round robin (RR1). Harmonization potential of the sRM was simulated in RR1 and evaluated in practice in RR2 among 11 measurement procedures using three native human plasma samples. Comprehensive purity analysis of a candidate primary RM (cpRM) was performed by state-of-the-art procedures. The sRM was value assigned with an isotope dilution mass spectrometry-based candidate reference method calibrated using the certified pRM. The inter-assay CV without harmonization was 42.1% and 52.8% in RR1 and RR2, respectively. In RR1, simulation of harmonization with sRM resulted in an inter-assay CV of 11.0%, whereas in RR2 calibration with the material resulted in an inter-assay CV of 19.1%. Both the sRM and pRM passed international homogeneity criteria and showed long-term stability. We assigned values to the low (0.95 ± 0.11 nmol/L) and middle concentration (3.75 ± 0.17 nmol/L) calibrators of the sRM. Standardization of hepcidin is possible with our sRM, which value is assigned by a pRM. We propose the implementation of this material as an international calibrator for hepcidin-25.
Mass spectrometry-based methods play a crucial role in the quantification of the main iron metabolism regulator hepcidin by singling out the bioactive 25-residue peptide from the other naturally occurring N-truncated isoforms (hepcidin-20, -22, -24), which seem to be inactive in iron homeostasis. However, several difficulties arise in the MS analysis of hepcidin due to the sticky character of the peptide and the lack of suitable standards. Here, we propose the use of amino- and fluoro-silanized autosampler vials to reduce hepcidin interaction to laboratory glassware surfaces after testing several types of vials for the preparation of stock solutions and serum samples for isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS). Furthermore, we have investigated two sample preparation strategies and two chromatographic separation conditions with the aim of developing an LC-MS/MS method for the sensitive and reliable quantification of hepcidin-25 in serum samples. A chromatographic separation based on usual acidic mobile phases was compared with a novel approach involving the separation of hepcidin-25 with solvents at high pH containing 0.1% of ammonia. Both methods were applied to clinical samples in an intra-laboratory comparison of two LC-MS/MS methods using the same hepcidin-25 calibrators with good correlation of the results. Finally, we recommend an LC-MS/MS-based quantification method with a dynamic range of 0.5–40 μg/L for the assessment of hepcidin-25 in human serum that uses TFA-based mobile phases and silanized glass vials.
We developed a rapid and robust HPLC-MS/MS (QqQ) method for the quantification of hepcidin-25, a promising new biomarker in iron metabolism, in human samples. The novelty of the method is the use of special HPLC vials to avoid adsorptive losses due to the basic character of the peptide that causes interaction with the silanol groups of the vial’s glass surface. Up to 90% decrease in the MS/MS signal was observed, when commercial HPLC vials were used, while vials treated with 3-(2-aminoethylamino)propylmethyldimethoxysilane or 1H,1H,2H,2H-perfluorooctyltriethoxysilane, leading to no losses in the range of physiological hepcidin-25 mean serum levels (10-20 µg/L).
Hepcidin-25 has attracted much attention ever since its discovery in 2001. It is widely recognized that this liver produced peptide hormone plays a major role in the regulation of iron levels in mammals and can reveal important clinical information about several pathological states in patients suffering from iron-related disorders.
With the aim to tackle the current difficulties in hepcidin quantification and improve the status of this promising biomarker in the clinical field, we developed a rapid and robust analytical strategy for the quantification of hepcidin-25 in human samples based on HPLC-MS/MS (QqQ) to be implemented in routine laboratories. The novelty of the method is the use of special HPLC vials to avoid adsorptive losses due to the basic character of the peptide that causes interaction with the silanol groups of the vial’s glass surface. Up to 90% decrease in the MS/MS signal was observed, when commercial HPLC vials were used, while vials treated with 3-(2-aminoethylamino)propylmethyl-dimethoxysilane or 1H,1H,2H,2H-perfluorooctyltriethoxysilane, leading to no significant losses in the dynamic range of physiological hepcidin-25 mean serum levels (10-20 µg/L).
Careful analytical validation was performed for determining the reproducibility, repeatability, limit of quantification (0.5 µg/L) and linearity (0.5-40 µg/L) of the method. Serum samples from 9 healthy volunteers were analyzed with a median hepcidin-25 level of 3.3 µg/L, comparable to results reported in the literature.
Hepcidin-25 has attracted much attention ever since its discovery in 2001. It is widely recognized that this peptide hormone plays a major role in the regulation of iron levels in mammals and can reveal important clinical information about several iron-related disorders. However, the development of a reliable assay to quantify hepcidin proved to be problematic and serum hepcidin-25 concentrations determined by various assays differ substantially. Challenges arise in the MS analysis of hepcidin due to the “sticky” character of the peptide and the lack of suitable standards.
With the aim to tackle the current difficulties in hepcidin quantification and improve the status of this promising biomarker in the clinical field, we developed a rapid and robust analytical strategy for the quantification of hepcidin-25 in human samples based on HPLC-MS/MS (QqQ) as a reference method candidate to be implemented in routine laboratories. The novelty of the method is the use of amino- and fluoro-silanized autosampler vials to reduce hepcidin interaction to laboratory glassware surfaces. Furthermore, we have investigated two sample preparation strategies and two chromatographic separation conditions where the use of acidic mobile phases was compared with a novel approach involving solvents at high pH containing 0.1% of ammonia. Both methods were carefully validated and applied to clinical samples in an intra-laboratory comparison of two LC-MS/MS methods using the same hepcidin-25 calibrators with very good correlation of the results.
HPLC-MS/MS quantification of hepcidin-25, a new promising iron biomarker, in human serum samples
(2017)
Hepcidin-25 has attracted much attention ever since its discovery in 2001. It is widely recognized that this liver produced peptide hormone plays a major role in the regulation of iron levels in mammals. Hepcidin levels could reveal important clinical information about several pathological states in patients suffering from iron-related disorders such as chronic kidney disease (CKD) or hereditary haemochromatosis (HH). Despite significant efforts though, the development of a reliable assay to quantify hepcidin proved to be problematic for the last 15 years.
With the aim to tackle the current difficulties in hepcidin quantification and improve the status of this promising biomarker in the clinical field, we developed a rapid and robust analytical strategy for the quantification of hepcidin-25 in human samples based on HPLC-MS/MS (QqQ) to be implemented in routine laboratories. The novelty of the method is the use of special HPLC vials to avoid adsorptive losses due to the basic character of the peptide that causes interaction with the silanol groups of the vial’s glass surface. Up to 90% decrease in the MS/MS signal was observed, when commercial HPLC vials were used, while vials treated with 3-(2-aminoethylamino)propylmethyl-dimethoxysilane or 1H,1H,2H,2H-perfluorooctyltriethoxysilane, leading to no significant losses in the dynamic range of physiological hepcidin-25 mean serum levels (10-20 µg/L).
Careful analytical validation was performed for determining the reproducibility, repeatability, limit of quantification (0.5 µg/L) and linearity (0.5-40 µg/L) of the method. Serum samples from 9 healthy volunteers were analyzed with a median hepcidin-25 level of 3.3 µg/L, comparable to results reported in the literature.