Datei im Netzwerk der BAM verfügbar ("Closed Access")
Filtern
Dokumenttyp
- Zeitschriftenartikel (12)
- Vortrag (8)
- Posterpräsentation (8)
- Sonstiges (1)
Schlagworte
- Mikroplastik (10)
- TED-GC-MS (10)
- Microplastics (9)
- Microplastic (8)
- TED-GC/MS (6)
- Analyse (2)
- Gas chromatography (2)
- Internal Standard (2)
- Thermoanalytik (2)
- Tire wear (2)
Organisationseinheit der BAM
Eingeladener Vortrag
- nein (8)
A harmonization of sampling, sample preparation and detection is pivotal in order to obtain comparable data on microplastics (MP) in the environment. This paper develops and proposes a suitable sampling concept for waterbodies that considers different plastic specific properties and influencing factors in the environment.
Both artificial water including defined MP fractions and the discharge of a wastewater treatment plant were used to verify the derived sampling procedure, sample preparation and the subsequent analysis of MP using thermal extraction-desorption gas chromatography - mass spectrometry (TED-GC-MS).
A major finding of this paper is that an application of various particle size classes greatly improves the practical handling of the sampling equipment. Size classes also enable the TED-GC-MS to provide any data on the MP size distribution, a substantial sampling property affecting both the necessary sampling volume and the optimal sampling depth.
In the artificial water with defined MP fractions, the recovery rates ranged from 80 to 110%, depending on the different MP types and MP size classes. In the treated wastewater, we found both Polyethylene and polystyrene in different size classes and quantities.
There is growing interest in quantifying microplastics in environmental samples. This application note presents a thermal extraction desorption-gas chromatography/mass spectrometry (TED-GC/MS) method that is well suited to automation and increased sample throughput. The method is also able to detect all particle sizes in the sample as long as the limit of detection (LOD) is reached and allows analysis of larger samples of 15 to 25 mg or more. Samples were decomposed by thermogravimetric analysis (TGA), and the gaseous decomposition products were trapped on a solid-phase sorbent, followed by thermal desorption‑gas chromatography/mass spectrometry (TD-GC/MS) using an Agilent 5977B GC/MSD coupled to an Agilent 7890B GC. Target microplastic particle (MP) polymers were identified in environmental samples including surface water, finished compost, house dust, and drinking water. Quantification of MP polymers in environmental samples provided LODs of 0.06 to 2.2 μg, allowing the detection of MPs in trace amounts with sample weights of up to 1 g. Method repeatability was adequate for reliable quantification with RSDs of approximately 6 to 12%.
Sulphurous additives for polystyrene: Influencing decomposition behavior in the condensed phase
(2015)
The thermal decomposition behaviour of polystyrene (PS) containing sulphur and phosphorus additives was investigated, using thermogravimetry coupled with Fourier transform infrared spectroscopy (TGA-FTIR). It was found that the additives influence the decomposition process of the polymer in the condensed phase, resulting in a decrease in styrene monomer formation and an increase in styrene oligmer derivatives. Via reference measurements with binary mixtures it was found that the presence of sulphur additives influences the radicalic decomposition process of PS. In combination with quantum chemical calculations it was concluded that this is due to the formation of radicals that abstract hydrogen from the polymer matrix at lower temperatures, disfavouring the radicalic decomposition pathway leading to styrene.
n order to determine the relevance of microplastic particles in various environmental media, comprehensive investigations are needed. However, no analytical method exists for fast identification and quantification. At present, optical spectroscopy methods like IR and RAMAN imaging are used. Due to their time consuming procedures and uncertain extrapolation, reliable monitoring is difficult. For analyzing polymers Py-GC-MS is a standard method. However, due to a limited sample amount of about 0.5 mg it is not suited for analysis of complex sample mixtures like environmental samples. Therefore, we developed a new thermoanalytical method as a first step for identifying microplastics in environmental samples. A sample amount of about 20 mg, which assures the homogeneity of the sample, is subjected to complete thermal decomposition. The specific degradation products of the respective polymer are adsorbed on a solid-phase adsorber and subsequently analyzed by thermal desorption gas chromatography mass spectrometry. For certain identification, the specific degradation products for the respective polymer were selected first. Afterwards real environmental samples from the aquatic (three different rivers) and the terrestrial (bio gas plant) systems were screened for microplastics. Mainly polypropylene (PP), polyethylene (PE) and polystyrene (PS) were identified for the samples from the bio gas plant and PE and PS from the rivers. However, this was only the first step and quantification measurements will follow.
In order to determine the relevance of microplastic particles in various environmental media, comprehensive investigations are needed. However, no analytical method exists for fast identification and quantification. At present, optical spectroscopy methods like IR and RAMAN imaging are used. Due to their time consuming procedures and uncertain extrapolation, reliable monitoring is difficult. For analyzing polymers Py-GC-MS is a standard method. However, due to a limited sample amount of about 0.5 mg it is not suited for analysis of complex sample mixtures like environmental samples. Therefore, we developed a new thermoanalytical method as a first step for identifying microplastics in environmental samples. A sample amount of about 20 mg, which assures the homogeneity of the sample, is subjected to complete thermal decomposition. The specific degradation products of the respective polymer are adsorbed on a solid-Phase adsorber and subsequently analyzed by thermal Desorption gas chromatography mass spectrometry. For certain identification, the specific degradation products for the respective polymer were selected first. Afterwards real environmental samples from the aquatic (three different rivers) and the terrestrial (bio gas plant) systems were screened for microplastics. Mainly polypropylene (PP), polyethylene (PE) and polystyrene (PS) were identified for the samples from the bio gas plant and PE and PS from the rivers. However, this was only the first step and quantification measurements will follow.
The TED-GC-MS analysis is a two-step method. A sample is first decomposed in a thermogravimetric analyzer (TGA) and the gaseous decomposition products are then trapped on a solid-phase adsorber. Subsequently, the solid-phase adsorber is analyzed with thermal desorption gas chromatography mass spectrometry (TDU-GC-MS). This method is ideally suited for the analysis of polymers and their degradation processes. Here, a new entirely automated System is introduced which enables high sample throughput and reproducible automated fractioned collection of decomposition products. Strengths and limitations of the system configuration are elaborated via three examples focused on practical challenges in materials analysis and identification: i) separate analysis of the components of a wood-plastic-composite material, ii) quantitative determination of weight concentration of the constituents of a polymer blend and iii) quantitative analysis of model samples of microplastics in suspended particulate matter.
Deuterated polystyrene (D-PS) improves microplastic detection with TED-GC / MS, among other things through higher reproducibility. It is inexpensive (0.1 ct / analysis) and can be dosed quickly and reproducibly thanks to its good solubility (e.g. in toluene). However, certain matrices lead to an exchange of the deuterium with hydrogen. The poster shows the measuring principle, in which matrices H/D-exchange takes place, the results of kinetic studies and ways of preventing or circumventing the H/D-exchange.
Deuteriertes Polystyrol (D-PS) verbessert die Mikroplastikdetektion mit der TED-GC/MS, u.a. durch höhere Reproduzierbarkeit. Es ist kostengünstig (0,1 ct/Analyse) und durch die gute Löslichkeit (z.B. in Toluol) schnell und reproduzierbar zu dosieren. Bestimmte Matrices führen jedoch zu einem Austausch des Deuteriums durchWasserstoff. Das Poster zeigt das Messprinzip, bei welchen Matrices H/D-Austausch erfolgt, Ergebnisse von Kinetikuntersuchungen sowie Möglichkeiten, den Austausch zu verhindern oder zu umgehen.
The presence of plastic debris and especially small plastic particles in marine ecosystems has been recognized in the 1970`s. Since then various other environmental systems all over the world, like rivers, lakes and biota have been found to be polluted by plastics. Lately, agriculturally used land, waste-water treatment and biogas facilities come into focus as possibly relevant distribution vectors. To monitor the situation and investigate distribution pathways, the need for fast and reliable analytical procedures is high.
Up to now, there is no standard procedure for sampling, concentrating and analyzing plastic particles in environmental samples. Our group recently reported the development of a new analyzing method for microscale plastic particles, thermal extraction desorption gas chromatography mass spectrometry (TED-GC-MS). This method consists of thermal extraction of the sample via thermogravimetric analysis, collection of the evolved gaseous decomposition products on a solid-phase adsorber and analysis of the loaded adsorber using Thermal-Desorption-GC-MS. This technique utilizes the characteristic decomposition products of polymers for identification and even quantification of plastics in environmental samples.
In the present work we investigated samples of fermentation residues taken from a biogas plant. The results of sample preparation, concentration and analysis are presented as well as a comparison with alternative methods.