Isotope analysis plays a critical role in various disciplines, including environmental science, archaeology, and forensic investigations. Traditional methods such as mass spectrometry provide precise isotopic data but often require complex, costly setups and extensive sample preparation. As an alternative, optical spectrometry has emerged as a versatile and less invasive technique. This presentation explores the advancements and applications of optical spectrometry methods in isotope analysis, emphasizing their benefits and challenges.
Lithium (Li), Boron (B), Nitrogen (N), Magnesium (Mg), and Calcium (Ca) are pivotal elements across various spheres such as the hydrosphere, biosphere, and lithosphere, significantly impacting (bio-) geochemical and physiological processes. These elements exhibit stable isotopes with substantial roles in geological, environmental, and biological studies. The traditional method for measuring isotope amount ratios has been through mass spectrometry, which, despite its accuracy, comes with high operational costs, the need for skilled operators, and time-consuming sample preparation processes.
Combining optical spectroscopy with chemometrics introduces an innovative, cost-effective approach by the hand of high-resolution continuum source atomic and molecular absorption spectrometry (HR-CS-AAS and HR-CS-MAS) for the analysis of isotope ratios in Li, B, N, Mg, and Ca. By analyzing the atomic or molecular absorption spectrum of the in-situ generated cloud of atoms of diatomic molecules (e.g., Li, BH, NO, MgF, CaF) during the electronic transition from the fundamental state, this method allows for the rapid determination of isotope ratios directly from sample solutions without the need for complex sample preparation.
For each element, the respective atomic or molecule's absorption spectrum was deconvoluted into its isotopic components using partial least squares regression or machine learning algorithms. Robust calibration models were developed, calibrated with enriched isotope, and validated against certified reference materials. Spectral data underwent preprocessing to optimize the modeling to determine the optimal number of latent variables.
The findings showcase that this optical spectrometric method yields results that agree with those obtained via inductively coupled plasma mass spectrometry (ICP-MS), offering a promising, cost-effective, and rapid alternative for isotope analysis with precisions as low as ± 0.2‰. This approach is a significant advancement in analytical chemistry, providing a new way to study isotope variations in biological, environmental, and geological samples.
An alternative method for lithium isotope analysis by using high-resolution atomic absorption spectrometry (HR-CS-AAS) is proposed herein. This method is based on monitoring the isotope shift of approximately 15 pm for the electronic transition 22P←22S at around the wavelength of 670.8 nm, which can be measured by state-of-the-art HR-CS-AAS. Isotope analysis can be used for (i) the traceable determination of Li concentration and (ii) isotope amount ratio analysis based on a combination of HR-CS-AAS and spectral data analysis by machine learning (ML).
In the first case, the Li spectra are described as the linear superposition of the contributions of the respective isotopes, each consisting of a spin-orbit doublet, which can be expressed as Gaussian components with constant spectral position and width and different relative intensity, reflecting the isotope ratio in the sample. Precision was further improved by using lanthanum as internal spectral standard. The procedure has been validated using human serum-certified reference materials. The results are metrologically comparable and compatible with the certified values.
In the second case, for isotope amount ratio analysis, a scalable tree boosting ML algorithm (XGBoost) was employed and calibrated using a set of samples with 6Li isotope amount fractions ranging from 0.06 to 0.99 mol mol−1. The training ML model was validated with certified reference materials. The procedure was applied to the isotope amount ratio determination of a set of stock chemicals and a BAM candidate reference material NMC111 (LiNi1/3Mn1/3Co1/3O2), a Li-battery cathode material. These determinations were compared with those obtained by MC-ICP-MS and found to be metrologically comparable and compatible. The residual bias was −1.8‰, and the precision obtained ranged from 1.9‰ to 6.2‰. This precision was sufficient to resolve naturally occurring variations. The NMC111 cathode candidate reference material was analyzed using high-resolution continuum source atomic absorption spectrometry with and without matrix purification to assess its suitability for technical applications. The results obtained were metrologically compatible with each other.
Ein Weg zur wissensbasierten Produktion führt über die Betrachtung der wesentlichen Apparate-, Prozess- und Freigabedaten aus Betrieben und Labors. Das Potenzial dieser Daten wird heute vielfach noch nicht konsequent für ein umfassendes Verständnis der Produktion genutzt.
Neben Fragen zur Datenerfassung, Datenkonnektivität und Datenintegrität müssen solche Daten für eine ganzheitliche Prozessanalyse zunächst mit Kontextinformationen zusammengebracht werden. Datenquellen enthalten vor allem Zeitwertpaare, die numerisch vorbehandelt und möglichst vollautomatisch ausgewertet werden müssen. Am Beispiel NMR-spektroskopischer Daten wird der Stand der Auswertung mit physikalisch motivierten Modellen, wie z. B. dem IHM erläutert.