Chemische Charakterisierung und Spurenanalytik
Filtern
Erscheinungsjahr
- 2023 (18) (entfernen)
Dokumenttyp
- Vortrag (8)
- Posterpräsentation (5)
- Zeitschriftenartikel (3)
- Beitrag zu einem Tagungsband (1)
- Handbuch (1)
Schlagworte
- Process Analytical Technology (6)
- Benchtop-NMR (5)
- Benchtop NMR (2)
- Hydrogen Storage (2)
- MefHySto (2)
- NMR spectroscopy (2)
- Online-NMR-Spektroskopie (2)
- Prozessanalytik (2)
- qNMR (2)
- ANAKON (1)
Organisationseinheit der BAM
- 1.4 Prozessanalytik (18) (entfernen)
Eingeladener Vortrag
- nein (8)
Volatile organic compounds (VOCs) emitted by furniture and
building materials can cause health issues. For an
improvement of indoor air quality low emitting materials
should be used. Quality assurance and –control (QA/QC)
measures require an emission reference material (ERM) with a
predictable emission rate of VOCs. The idea is to use porous
materials as ERM, which store the VOCs inside their pores and
emit them constantly.
qNMR - The handbook
(2023)
Quantitative NMR (qNMR) has been around for a long time, but also has great potential to solve future problems in any quantitative analysis. As a primary method, it differs fundamentally from chromatographic methods: it is better described as a quantum mechanical balance. Succesful implementation of qNMR requires certain attention to detail. "qNMR - the handbook" is intended to be a guide for analysts to help understand the fundamental principles of NMR and the significant points relating to its implementation for quantitation. Regulatory considerations of qNMR adoption are explained. NMR fundamentals are explained to provide understanding. Together with many useful examples, the book is a compelling addition to the laboratory's reference library, providing all the tools that any practitioner should know to successfully implement qNMR.
The authors are qNMR pioneers and come from a variety of backgrounds including business, government and academia.
The application of compact NMR instruments to continuously flowing samples at higher temperatures or exothermically reacting mixtures is limited by the temperature sensitivity of permanent magnets. Typically, such temperature effects directly influence the achievable magnetic field homogeneity and hence measurement quality. The internal-temperature control loop of the magnet and instruments is typically not designed for compensation of such effects. Passive insulation is restricted by the small dimensions within the magnet borehole.
Here, we present a design approach for active heat shielding with the aim of variable temperature control of NMR samples for benchtop NMR instruments using a compressed airstream which is variable in flow and temperature. Based on the system identification and surface temperature measurements using an optical thermo-graphy setup, a model predictive control was set up to minimize any disturbance effect on the permanent magnet from the probe or sample temperature.
This methodology will facilitate the application of variable-temperature shielding and, therefore, potentially extend the application of compact NMR instruments to flowing samples at higher temperature than the permanent magnet.
Die NMR-Spektroskopie ist eine der zentralen nicht-invasiven Analysenmethoden in der organischen Chemie und aus dem Laboralltag nicht mehr wegzudenken. Die direkte Proportionalität zwischen der Anzahl der Atomkerne im Messvolumen und der Signalfläche im Spektrum ist vergleichbar mit einem „Zählen der Kernspins“. Der Kalibrieraufwand für die Quantifizierung ist minimal und aus dem NMR-Spektrum sind Informationen zu Struktur und Identität zugänglich.
Online-NMR-Spektroskopie wird unter Zuhilfenahme von speziellen Durchflussproben-köpfen und Messzellen bereits seit Jahrzehnten erforscht und eingesetzt. Durch die hohen Anforderungen an die Aufstellung und den Betrieb von klassischen Hochfeld-NMR-Spektrometern konnte sie allerdings nie den Sprung vom Labor in den Prozess schaffen. Dies hat sich durch die Entwicklungen im Bereich mobil einsetzbarer Benchtop-NMR-Spektrometer grundlegend verändert. Geringere Investitions- und Betriebskosten, sowie die Robustheit und einfache Bedienbarkeit dieser Systeme sind entscheidende Faktoren. Damit rückt eine Anwendung als Online-PAT-Methode technisch in greifbare Nähe. Designierte Prozess-NMR-Spektrometer sucht man heutzutage allerdings auf dem Markt meist noch vergebens.
Die Reaktionsverfolgung im Labormaßstab konnte an unterschiedlichen Systemen erfolgreich demonstriert werden. Dabei zeigten sich allerdings Limitierungen in Hinblick auf einen möglichen Prozesseinsatz. Insbesondere die Temperaturempfindlichkeit der Magnetsysteme ist hier zu nennen. Anhand eines Prototyps wurde eine aktive Temperaturisolation mittels temperierter Luftströmungen entwickelt und erprobt.
Eine Feldintegration von Laborsystemen geht sowohl technisch als auch regulatorisch mit zahlreichen Herausforderungen einher. Das raue Umfeld von Produktionsanlagen, sowie die Anforderungen an Explosionsschutz erfordern eine zugelassene Einhausung. Auf Basis von Erfahrungen mit einem ersten Prototyp erfolgte die gemeinsame Entwicklung eines möglichst flexibel einsetzbaren Analysenschranks. Neben der Hardware-Integration sind die automatisierte modellbasierte Auswertung der NMR-Spektren, sowie die Einbindung in die Prozessleittechnik essenziell für einen zuverlässigen Betrieb als PAT-Analysengerät.
The application of compact NMR instruments to hot flowing samples or exothermically reacting mixtures is limited by the temperature sensitivity of permanent magnets. Typically, such temperature effects directly influence the achievable magnetic field homogeneity and hence measurement quality. The internal-temperature control loop of the magnet and instruments is not designed for such temperature compensation. Passive insulation is restricted by the small dimensions within the magnet borehole.
Here, we present a design approach for active heat shielding with the aim of variable temperature control of NMR samples for benchtop NMR instruments using a compressed airstream which is variable in flow and temperature. Based on the system identification and surface temperature measurements through thermography, a model predictive control was set up to minimise any disturbance effect on the permanent magnet from the probe or sample temperature.
This methodology will facilitate the application of variable-temperature shielding and, therefore, extend the application of compact NMR instruments to flowing sample temperatures that differ from the magnet temperature.
Global demand for pigments is expected to increase to around 13.4 million tons per year by 2030, representing approximately $22 billion – 5% of which will be azo pigments as the main organic pigments. Azo pigments, i.e., especially the azo bridge required for them, are almost exclusively produced in technology by the so-called azo coupling reaction in large-scale approaches on the scale of 50–100 cubic meters. Among other things, one of the reasons why large-scale production itself is still essentially discontinuous is the lack of analytical methods for a continuous azo pigment synthesis.
In an azo coupling process, an aromatic amine is diazotized to form the diazonium compound (diazo compound), which then reacts with a coupling component, i.e., "couples". A slight excess of the coupler is desired for the end of the reaction, which stoichiometrically consumes the remaining (harmful) diazo compound, i.e., below 500 ppm. This is carried out with a manual spotting reaction of a fast-binding coupler.
To carry out continuous azo coupling, care must be taken to ensure that the actual coupling reaction proceeds as rapidly as possible, i.e., preferentially, compared with possible side reactions. Reliable measuring methods are thus a decisive criterion for the useful performance of a continuous azo pigment synthesis. An analytical online procedure is difficult due to several limitations at once: solid dispersions extremely limit the range of reproducible analytical procedures. In addition, the appearance of a signal (e.g., excess of a starting material) in the ppm range must be reliably detected against the matrix background.
The back-titration method and benchtop NMR spectroscopy were used to overcome these problems. We present a feasibility study for a suitable process analysis with benchtop NMR spectroscopy on disperse systems for equimolar dosing of the "diazotization" or "coupling" sub-step in azo coupling, which was carried out in cooperation with a chemical company. This could replace tedious manual spotting
Compact NMR Spectroscopy in the field: A Versatile PAT Tool for Production of Specialty Chemicals
(2023)
Companies of chemical industry find themselves more often in a rapidly changing environment, e.g., due to variability of raw material quality or energy costs and efficiency. Process optimization and new process concepts become more and more important. For example, flexible and modular chemical plants can produce various high-quality products using multi-purpose equipment with short downtimes between campaigns and reduce time to market for new products. Intensified continuous production plants allow for difficult to produce compounds like exothermic reactions with high heat dissipation.
Highly automated “chemical” process control along with real-time quality control are prerequisites to such concepts and thus should be based on “chemical” information. A commercially available benchtop NMR spectrometer was integrated to the full requirements of an automated chemical production environment such as explosion safety, field communication, and robust evaluation of sensor data. Field studies in modular and conventional production plant setups show promising results gaining process knowledge for further optimization. NMR appeared as preeminent online analytical method and allow using a modular data analysis approach, which can even serve as reliable reference method for further calibration-dependent PAT applications (e.g., NIR or Raman spectroscopy).
Based on experiences from earlier field studies an improved NMR analyzer enclosure setup was developed and built, including the option of a secondary method (e.g., optical spectroscopy). Integrated control systems allow for a flexible implementation based on the available automation infrastructure at the chemical plant or pilot plant setup. In the future, fully integrated and intelligently interconnecting “smart” PAT systems and processes have the potential speed up the setup of production equipment for chemicals and pharmaceuticals and therefore help to reduce the time-to-market.
Volatile organic compound (VOC) emissions from consumer products contribute to human inhalation exposure and may cause adverse health effects. Existing methods to determine long-term VOC emissions from e.g. building products need to be verified for their suitability to reliably detect initial VOC emissions from consumer products within the first hours and days of use, which would facilitate realistic inhalation exposure assessments. To investigate this issue, VOCs emitted from a test sample were determined in a large-scale emission test chamber and in two micro-chambers of different volumes, and the results were compared.
This study uses conventional 87Sr/86Sr and 143Nd/144Nd isotope and interelement ratios of Ca, Sr, K, Mn, Mg and Ti as fingerprints for provenancing ordinary Portland cements (OPC). Herein, the first database of Sr and Nd isotope ratios investigated in OPCs, stemming from 29 cement plants located worldwide, was created. The results show that the Sr isotope ratios of OPCs are higher than those of seawater from the observed geological period. The spread of 143Nd/144Nd in OPCs is not as large as the spread for 87Sr/86Sr isotope ratios. However, the combination of both Sr and Nd isotope ratios provides the potential for distinguishing between cements of different production sites. Most of the OPCs investigated have measurable differences in their 87Sr/86Sr and 143Nd/144Nd isotope ratios, which can be employed as a valuable analytical fingerprinting tool. In the case of equivocal results, divisive hierarchical clustering was employed to help overcome this issue. The construction of geochemical profiles allowed the computing of suitably defined distances between cements and clustering them according to their chemical similarity. By applying this methodology, successful fingerprinting was achieved in 27 out of the 29 ordinary Portland cements that were analysed.
Modular chemical production is a tangible implementation of the digital transformation of the specialty chemicals process industry. In particular, it enables acceleration of process development and thus faster time to market by flexibly interconnecting and orchestrating standardized physical modules and bringing them to life. For this purpose, specific (chemical) sensors of process analytics are needed, preferably without lengthy calibration or spectroscopic model development.
An excellent example of a "direct" analytical method is online nuclear magnetic resonance (NMR) spectroscopy. NMR spectroscopy meets the requirements of a direct analytical method because of the direct correlation between the signal area in the spectrum ("counting" the nuclear spins) and the analyte amount of substance concentrations. It is also extremely linear over the concentration range.
With the availability of compact benchtop NMR instruments, it is now possible to bring NMR spectroscopy directly into the field, in close proximity to specialized laboratory facilities, pilot plants, and even industrial-scale production facilities. The first systems are in TRL 8 (Qualified System with Proof of Functionality in the Field).
The presentation will discuss the many building blocks of online nuclear magnetic resonance spectroscopy, from flow cells to automated data analysis.