The ISO 16000-6 standard describes a method for the determination of volatile organic compounds (VOCs) in indoor and test chamber air by sorbent-based active sampling, thermal desorption and gas chromatography coupled with mass spectrometry (GC/MS). It also gives directions to adapt this methodology to very volatile organic compounds (VVOCs). Indeed, toxicologically based guideline values are being implemented for these compounds and it becomes necessary to measure them. But a comprehensive and robust measurement method is lacking. This work highlights the points that still need to be explored towards the standardisation of a suitable procedure: investigations on sorbent combinations, the suitability of chromatography columns and the use of gaseous standards are required. The biggest challenge remains in the fact that strong sorbents adsorb water together with VVOCs. Water may impair the analysis and the optimal approach to eliminate it is still to be found and integrated into the sampling strategy.
Formaldehyde is considered as carcinogenic and is emitted from particleboards and plywood used in toy manufacturing. Currently, the flask method is frequently used in Europe for market surveillance purposes to assess formaldehyde release from toys, but its concordance to levels measured in emission test chambers is poor. Surveillance laboratories are unable to afford laborious and expensive emission chamber testing to comply with a new amendment of the European Toy Directive; they need an alternative method that can provide reliable results. Therefore, the application of miniaturised emission test chambers was tested. Comparisons between a 1 m3 emission test chamber and 44 mL microchambers with two particleboards over 28 days and between a 24 L desiccator chamber and the microchambers with three puzzle samples over 10 days resulted in a correlation coefficient r2 of 0.834 for formaldehyde at steady state. The correlation between the results obtained in microchambers vs. flask showed a high variability over 10 samples (r2: 0.145), thereby demonstrating the error-proneness of the flask method in comparison to methods carried out under ambient parameters. An exposure assessment was also performed for three toy puzzles: indoor formaldehyde concentrations caused by puzzles were not negligible (up to 8 μg/m3), especially when more conservative exposure scenarios were considered.
Formaldehyd gilt als krebserregend und wird von Spanplatten und Sperrholz emittiert, die bei der Spielzeugherstellung verwendet werden. Die Untersuchungsänter benötigen eine wirtschaftliche und zuverlässige Methode. Die Anwendung miniaturisierter Emissionsprüfkammern wurde getestet und mit größeren Emissionsprüfkammern und die Flaschenmethode verglichen. Eine Expositionsabschätzung wird auch vorgestellt.
Several aspects were explored towards the standardization of a suitable procedure. The use of gaseous standards is necessary and a standard gas mixture containing 60 substances was successfully generated and employed for further investigations. The suitability of different chromatography columns was addressed: The use of PLOT (Porous Layer Open Tubular) columns is well suitable for VVOC analysis. The recoveries of the 60 analytes on several adsorbents and their combinations were determined: A combination of a graphitized carbon black and a carbon molecular sieve showed great results for all analytes. Carbon molecular sieves adsorb water which can impair the analysis. Different options such as purging, the use of a drying system or splitting were investigated for water removal. This contribution will present experimental results supporting the standardization of a method for VVOC analysis.
Mehrere Aspekte wurden im Hinblick auf die Standardisierung eines geeigneten Verfahrens für die Analyse sehr flüchtiger organischer Verbindungen (VVOCs) untersucht. Die Verwendung gasförmiger Standards ist notwendig und es wurde ein stabiles Standardgasgemisch aus 60 Substanzen erfolgreich erzeugt und auf Stabilität und Reaktivität überprüft. Die Eignungsuntersuchung verschiedener Chromatographiesäulen ergab, dass PLOT-Säulen (Porous Layer Open Tubular) für die VVOC-Analytik gut geeignet sind. Für die 60 Analyten wurden Wiederfindungsraten auf mehreren Adsorbenzien und deren Kombinationen bestimmt. Dabei zeigte eine Kombination aus einem graphitierten Ruß und einem Kohlenstoff-Molekularsieb hervorragende Ergebnisse für alle Zielkomponenten. Nachteilig an Kohlenstoff-Molekularsieben ist, dass sie Wasser adsorbieren, was die Analyse beeinträchtigen kann. Zur Wasserentfernung wurden verschiedene Optionen wie das Trockenspülen oder der Einsatz eines Trocknungssystems untersucht.
Since 1997, the Committee for Health-related Evaluation of Building Products (AgBB) has been developing the basis for building regulations for protection against indoor health risks that may arise when building products are used. In 2009, the AgBB decided to include relevant VVOCs in the assessment procedure. The ISO 16000-6 (2021) states that the use of the thermal desorption gas chromatography technique is appropriate for VVOCs, if adaptations are considered. However, a suitable method for the trustworthy quantification of VVOC emissions from building products and in the indoor air is still missing.
This webinar will present the results of a research project on the development of a procedure for VVOC analysis:
• identification of the gaps towards standardization
• investigations on gas standards and the suitability of chromatography columns
• investigations on sorbent combinations and water management
• validation of the method and screening of VVOC emissions from building products.