For the first time polyethylene (PE) frits were used to quantify sulphur in copper metal and its alloys by isotope dilution combined with LA-ICP-MS: an alternative approach for sample preparation. The properties of the PE frit meet the requirements for isotope dilution LA-ICPMS which are porous material, thermal and chemical resistance and high absorption efficiency. The breakthrough, however, as a support material, is the low sulphur blank, which is only two times of the gas flow blank (2.3-4.0 x 10⁴ cps). Additionally, the porosity of the frit was considered, as it directly affects the absorption efficiency for the sample solution, which is present in the cavities of the frit.
The absorption efficiency was studied by loading sulphur standards with varying sulphur amounts (0 - 80 µg S) onto the frits. The remaining sulphur which was not absorbed by the frit was rinsed off and was measured by ICP-MS. This indirect method shows that more than 99.5 % of the loaded sulphur was absorbed by the frit. The so prepared frits with increasing sulphur amount were measured by LA-ICP-MS showing a good linearity between 0 µg S and 40 µg S with a coefficient of determination, r2 of 0.9987 and a sensitivity of 3.4x10⁴ cps µgˉ¹ for 32S.
Three copper reference materials produced by BAM (BAM-M376a, BAM-228 and BAM-227) were selected to develop and validate the LA-ICP-IDMS procedure. The IDMS technique was applied to these samples as follows: the samples were spiked, dissolved, digested and then the digested solution was absorbed on the frits. The dried frit samples were then analyzed by LA-ICP-IDMS and it could be demonstrated that the sample solution dispersed on the frits did not influence the 32S/34S ratio significantly even though the sulphur intensities were fluctuating along the scan lines. Relative standard deviations of the isotope ratios were below 5 % in average between 3 lines (except for the pure spike solution and procedure blank). The measurement results were validated by comparing them with the results obtained by conventional ICP-IDMS. Plotting the mass fraction of sulphur in copper obtained by LA-ICP-IDMS versus those obtained by ICP-IDMS yields a linear curve with a correlation coefficient of 0.9999 showing a strong agreement between both techniques.
The metrological traceability to the SI for the mass fraction of sulphur in copper is established by an unbroken chain of comparisons, each accompanied by an uncertainty budget. Thus, the measurement results are considered reliable, acceptable and comparable within the stated measurement uncertainty. The metrological traceability chain from the kg down to the final mass fraction in the samples obtained by LA-ICP-IDMS is illustrated in this presentation.
This is the first time that PE frits were used to quantify sulfur in copper and its alloys by isotope dilution combined with LA-ICP-MS: an alternative approach for sample preparation. The following properties of the PE frit meet the requirements for isotope dilution LA-ICPMS:: porous material, thermo plastic (melting point >100oC), chemical resistance (nitric acid >70%) and high adsorption efficiency. The breakthrough, however, as a support material, especially when comparing the PE frit with other materials such gelatin or sodium silicate is the low blank, which is only two times of the gas flow blank (2.3-4.0 x 104 cps). Additionally, the porosity of the frit was considered, as it directly affects the adsorption efficiency for the sample solution, which is present in the cavities of the frit.
Adsorption efficiency was studied by depositing sulfur standards with varying sulfur amounts (0, 2, 5, 10, 20, 40 and 80 µg S) on the frits. The remaining sulfur which was not absorbed by the frit was rinsed off and was measured by ICP-MS. This indirect method shows that more than 99.5 % of the loaded sulfur was absorbed by the frit. Such high absorption efficiency is completely sufficient for a support material to be used in LA-ICP-IDMS. The so prepared frits whith increasing sulfur amount were measured by LA-ICP-MS showing a good linearity between 0 µg S and 40 µg S whit a correlation coefficient r2 of 0.9987 and sensitivy of 3.4x104 cpsµg-1 for 32S.
Three copper reference materials produced by BAM were selected to develop and validate the LA-ICP-IDMS procedure. The IDMS technique was applied to these samples as follows: the samples were spiked, dissolved, digested and then the digest was adsorbed on the frits.
This is the first time that PE frits were used to quantify sulfur in copper and its alloys by isotope dilution combined with LA-ICP-MS: an alternative approach for sample preparation. The following properties of the PE frit meet the requirements for isotope dilution LA-ICPMS: porous material, thermo plastic (melting point >100°C), chemical resistance (nitric acid >70%) and high adsorption efficiency. The breakthrough, however, as a support material, especially when comparing the PE frit with other materials such gelatin or sodium silicate is the low blank, which is only two times of the gas flow blank (2.3-4.0 x 10⁴ cps). Additionally, the porosity of the frit was considered, as it directly affects the adsorption efficiency for the sample solution, which is present in the cavities of the frit.
Adsorption efficiency was studied by depositing sulfur standards with varying sulfur amounts (0, 2, 5, 10, 20, 40 and 80 µg S) on the frits. The remaining sulfur which was not absorbed by the frit was rinsed off and was measured by ICP-MS. This indirect method shows that more than 99.5 % of the loaded sulfur was absorbed by the frit. Such high absorption efficiency is completely sufficient for a support material to be used in LA-ICP-IDMS. The so prepared frits with increasing sulfur amount were measured by LA-ICP-MS showing a good linearity between 0 µg S and 40 µg S with a correlation coefficient r2 of 0.9987 and sensitivity of 3.4x10⁴ cps µgˉ¹ for 32S.
Three copper reference materials produced by BAM were selected to develop and validate the LA-ICP-IDMS procedure. The IDMS technique was applied to these samples as follows: the samples were spiked, dissolved, digested and then the digest was adsorbed on the frits.
An immunohistochemical method is described to visualize the distribution of metallothioneins 1/2 (MT 1/2) and metallothionein 3 (MT 3) in human ocular tissue. It is making use of (a) antibodies conjugated to gold nanoclusters (AuNCs) acting as labels, and (b) laser ablation (LA) coupled to inductively coupled plasma – mass spectrometry (ICP-MS). Water-soluble fluorescent AuNCs (with an average size of 2.7 nm) were synthesized and then conjugated to antibody by carbodiimide coupling. The surface of the modified AuNCs was then blocked with hydroxylamine to avoid nonspecific interactions with biological tissue. Immunoassays for MT 1/2 and MT 3 in ocular tissue sections (5 μm thick) from two post mortem human donors were performed. Imaging studies were then performed by fluorescence using confocal microscopy, and LA-ICP-MS was performed in the retina to measure the signal for gold. Signal amplification by the >500 gold atoms in each nanocluster allowed the antigens (MT 1/2 and MT 3) to be imaged by LA-ICP-MS using a laser spot size as small as 4 μm. The image patterns found in retina are in good agreement with those obtained by conventional fluorescence immunohistochemistry which was used as an established reference method.
Nearly all disease processes are associated with variations of components of the extracellular matrix (ECM) that are typically observed during the development of inflammation. This concerns for example proteoglycans and their associated glycosaminoglycans (GAG), which have been shown to bind to cationic metal imaging probes due to their strong complexing activity. The complexing activity largely depends on the degree of GAG sulfation and/or carboxylation as well as on the GAG isomericity. In this central project, we investigate GAG structures from inflammatory disorders (namely cardiovascular diseases, inflammatory intestinal diseases and neuroinflammation) provided by researchers of the Collaborative Research Center at the molecular disaccharidic level using chromatographic and mass spectrometric methods. In parallel, the spatial localization and quantification of metal-based imaging probes are evaluated by LA-ICP-MS imaging.
Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is nowadays an established multi-elemental analysis and mapping technique. It was shown that LA-ICP-MS can visualize the elemental distribution within tissue thin sections or cell samples. Quantification is possible by using appropriate matrix-matched calibration samples. Besides naturally occurring elements and metals from contrast agents, biomolecules using metal-tagged antibodies were detected in different bio-medical samples. By combining the results with findings from histology, magnetic resonance imaging (MRI) and other techniques disease related changes like alterations of the extracellular matrix can be investigated.
In recent years, elemental imaging of biological samples like tissue thin sections using laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is gaining more and more importance. Improvements concerning spatial resolution as well as signal-to-background ratio due to low-dispersion sample chambers make LA-ICP-MS also interesting for single cell analysis.
To evaluate the interaction of nanoparticles (NPs) with cells LA-ICP-MS was applied for the imaging of individual cells. Our findings show, that NP aggregates can be localized within cellular compartments. The uptake efficiency depends strongly on the physico-chemical properties of the nanostructures (size, chemical composition, surface modification), as well as on the incubation conditions (concentration, time).
Moreover, LA-ICP-MS is increasingly becoming an important complementary technique in bioanalysis by using element-tagging strategies to determine biomolecules indirectly. Based on the specific binding between antibodies and their corresponding antigens, proteins and peptides can be detected in tissue or cells using tagged antibodies. As artificial tags metal chelates loaded with lanthanides, polymer-based elemental tags or metal-containing nanoparticles can be used. Thereby LA-ICP-MS is a sensitive detection tool for multiplexed immuno-histochemistry of tissue and cell samples.
Our results demonstrate the potential of LA-ICP-MS to investigate the distribution of naturally occurring elements, administered agents as well as biomolecules by using metal-tagged antibodies.
Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is nowadays a versatile and powerful analytical method for direct solid sample analysis. The applicability has been demonstrated for a wide variety of samples covering hard and soft materials. In an imaging mode the technique provides quantitative information on the elemental distribution within a sample. LA-ICP-MS imaging is of particular interest in biomedical research as the distribution of an element gives valuable insight on uptake and distribution of essential and toxic trace elements, administered contrast agents as well es nanoparticles. LA-ICP-MS is therefore a powerful complement to other imaging techniques. Recent instrumental improvements, especially in sample chamber design, have contributed to better sensitivity and spatial resolution enabling subcellular imaging. The book chapter provides a comprehensive overview about spatially resolved localisation and quantification of various nanoparticles in cells and tissue thin sections by LA-ICP-MS. Furthermore, different sample preparation strategies and internal standardisation and calibration approaches for bioimaging by LA-ICP-MS are summarized and discussed.
Metal-containing nanomaterials are used in numerous fields ranging from industrial applications to nanomedicine. Several studies have demonstrated that the physicochemical properties of nanoparticles have an impact on their pharmacokinetics, transfer and clearance. The high sensitivity and multielement capability of LA-ICP-MS enables the elucidation of interactions between tissue components and nanomaterials used as imaging probes or drug carriers. Potential toxic effects are investigated as well. Thus, LA imaging significantly supports the clinical translation of safe and efficient nanoparticles for diagnostic and therapeutic purposes.
Many current nanomaterials can serve as contrast agents in cellular or tissue imaging, drug delivery vehicles or therapeutics, whereas others can cause toxic effects. In order to evaluate nano-bio interactions, the number of nanoparticles (NPs) inside cells as well as their localisation within cellular substructures is of particular interest. The cellular uptake depends on the primary characteristics of the NPs (e.g. size, shape, surface coating) and on the cell type.
Laser ablation inductively coupled plasma mass spectrometry (LA‑ICP‑MS) is more and more used to study the NP pathway from uptake, via intracellular processing up to cell division. High-spatial resolution laser ablation at single cell level is achieved using novel low-dispersion LA chambers and by careful optimisation of laser energy, ablation frequency and scan speed at small laser spot sizes down to 1 µm. Different examples from BAM, Division 1.1 and cooperation partners using LA-ICP-MS to localize and quantify metal-containing nanoparticles are shown. The results demonstrate the potential of LA-ICP-MS providing insight into NP uptake, intracellular distribution and cell-to-cell variation.
We have applied laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) with subcellular resolution as an elemental mass microscope to investigate the distributions of Ag nanoparticles (NP) in a 3-dimentional multicellular spheroid (MCS) model. The production of MCS has been optimized by changing the seeding cell number (500 to 40,000 cells) and the growth period (1 to 10 days). Incubations of MCS with Ag nanoparticle suspensions were performed with a concentration of 5 µg mL-1 for 24 hours. Thin-sections of the Eosin stained MCS were analysed by elemental mass microscopy using LA-ICP-MS to image distributions of 109Ag, 31P, 63Cu, 66Zn and 79Br. A calibration using NP suspensions was applied to convert the measured Ag intensity into the number of particles being present in each measurement pixel. The numbers of NP determined ranged from 30 up to 4,000 particles in an enrichment zone. The particle distribution was clearly correlated to 31P, 66Zn and 79Br and was localized in an outer rim of proliferating cells (confirmed by DAPI) with a width of about two-single cell diameters. For the highest seeding cell number NPs were only detected in this outer rim, whereas small molecules as for instance 79Br and 109Ag ions were detected in the core of the MCS as well. Aniline blue staining demonstrated that this outer rim was rich in collagen structures in which fibroblast cells were embedded and a thin-membrane was visible which separated the core from the biological active cell layer functioning as biological barriers for NP transport. In this presentation, we will show the possibility using this 3-dimensional model for toxicological and medical applications.