In many scientific fields, isotopic analysis can offer valuable information. Up to date, typically bulk analysis is aimed at measuring the isotopic composition of the entire elemental content of the sample. However, the analyte element is usually present under the form of different species. Thus, separating species of interest from one another and from matrix components prior to isotope ratio measurements can provide species-specific isotopic information, which could be used for tracing the origin of environmental pollutants and elucidation of (environmental) speciation. Using on-line hyphenations of separation techniques with multicollector-ICP-MS (MC-ICP-MS) can save time and effort and enables the analysis of different species during a single measurement. Whereas some works hyphenating GC and IC with MC-ICP-MS have already been reported, LC and CE hyphenations are still inadequately represented based on the capabilities of these separation techniques.
In this work, we developed an on-line hyphenation of CE with multicollector-ICP-MS (CE/MC-ICP-MS) for isotopic analysis of sulfur species using a multiple-injection approach for instrumental mass bias correction by standard-sample bracketing. With this method, the isotopic composition of sulfur in sulfate originating from river water could be analyzed without sample preparation. The results were compared to data from off-line analysis of the same samples to ensure accuracy. The precision of the results of the on-line measurements was promising regarding the differentiation of the river systems by the isotopic signature of river water sulfate. The great potential of this method is based on the versatility of the applied separation technique, not only in the environmental field but also for, e.g., biomolecules, as sulfur is the only covalently bound constituent of proteins that can be analyzed by MC-ICP-MS.
In many scientific fields, isotopic analysis can offer valuable information, e.g., for tracing the origin of food products, environmental contaminants, forensic and archaeological samples (provenance determination), for age determination of minerals (geochronological dating) or for elucidating chemical processes. Up to date, typically bulk analysis is aimed at measuring the isotopic composition of the entire elemental content of the sample. However, the analyte element is usually present under the form of different species. Thus, separating species of interest from one another and from matrix components prior to isotope ratio measurements can provide species-specific isotopic information, which could be used for tracing the origin of environmental pollutants and elucidation of (environmental) speciation. Using on-line hyphenations of separation techniques with multicollector-ICP-MS (MC-ICP-MS) can save time and effort and enables the analysis of different species during a single measurement.
In this work, we developed an on-line hyphenation of CE with multicollector-ICP-MS (CE/MC-ICP-MS) for isotopic analysis of sulfur species. With this method, the isotopic composition of sulfur in sulfate originating from river water could be analyzed without sample preparation. The results were compared with data from off-line analysis of the same samples to ensure accuracy. The precision of the results of the on-line measurements was high enough to distinguish the rivers from one another by the isotopic signature of the river water sulfate. Next to environmental applications, a future field could be species-specific isotopic analysis of biomolecules, as sulfur is the only covalently bound constituent of proteins which can be analyzed by MC-ICP-MS.
Inductively coupled plasma mass spectrometry (ICP-MS) is a powerful method for the matrix-independent quantitative analysis of target elements. Developed for the use in inorganic trace analysis, ICP-MS is nowadays a valuable tool for bioanalytical questions. Especially the use of ICP-MS for quantitative proteomics by measuring heteroatoms has gained recognition in the last decade, considering that established quantification methods like organic mass spectrometry depend on labelling of the target protein or the existence of matched protein and peptide standards. The need for reliable quantification of proteins is continuously growing, but only a limited number of well-characterized and quantified protein standards are available so far. Accurately quantified, traceable protein standards are necessary to ensure comparability of measurements between laboratories, not only in basic research but also in a clinical context. One example of this is the Alzheimer’s disease biomarker tau protein. However, existing tau standards lack comparability, emphasizing the need for a well-quantified protein standard.
Therefore, we developed a method for the quantification of pure proteins via sulfur isotope dilution ICP-MS (IDMS). As sulfur is present in two amino acids, cysteine and methionine, it exists in nearly all proteins and can be used for the quantification of proteins of known stoichiometry. We employed simple offline strategies for the separation of non-protein bound sulfur species. Quantification of these contaminations by IDMS allows for correction of the protein content and enables reliable protein quantification. We report the protein mass fractions of a standard reference material and commercially available proteins determined by sulfur IDMS, including the expanded uncertainties. The developed method can be applied for the reliable and traceable quantification of pure proteins for use as in-house standards. Here, we successfully used this method for the quantification of the tau protein.
Diatoms are located at the bottom of the food chain. Toxicological relevant metals can possibly accumulate within the food web and cause harmful effects. Diatoms are a test system in ecotoxicology. Toxicological effects weaken the growth of algae which is by default investigated by means of fluorescence detection. On basis of the expose concentration as well as obtained fluorescence data potential threshold exceedance in e.g. surface waters is assessed.
However, this approach does not allow for the determination of “real” accumulated metal concentration in diatoms. Common approaches are based on bulk analysis via e.g. ICP-MS, ICP-OES or AAS. But, biological variability is completely disregarded.
To tackle this problem, alternative approaches are highly needed. Within the last years, sp-ICP-MS for nanoparticle as well as single cell analysis turned out as a powerful technique to analyze metal contents as well as size distributions on broad size range. But, common ICP-MS systems do not allow for multi-element detection within single particle/cell events. Thus, simultaneous MS detection devices are needed - just recently, ICP-ToF-MS experienced a revival.
Within our previous work, we developed an automated sample introduction system based on a HPLC system on-line with single particle-ICP-MS, which allowed for ionic background separation and single algae analysis. However, for unambiguous tracing several fingerprint elements and multielement analysis in single algae (diatoms) is needed. Thus, we coupled our previous setup on-line to ICP-ToF-MS. Test diatom species were exposed to test substances (Zn) as well as nanoparticles (FeNPs).
The developed setup allowed for a fast, automated and multielement analysis in single diatoms. Furthermore, we combined our approach with multivariate data assessment - multielement detection of characteristic fingerprint elements allowed for an unambiguous diatom tracing. Clustering of diatoms according to metal exposure concentration levels was enabled. Our approach is a new potential tool in ecotoxicological testing.
An introduction into absolute isotope ratios is given, with application fields ranging from fundamental science to geochronology and forensics. This is followed by a proposal for developing new calibration approaches for obtaining absolute isotope ratios at unrivaled uncertainty levels. This new developments will set the basis for improvements in all scientific fields.
The introduction of fluorine in organic molecules leads to new chemical/physical properties. Especially in the field of pharmaceuticals, fluorinated organic molecules are becoming more and more popular and at present amount up to 25% of market share, with an upward trend. The main benefits of fluorinated pharmaceuticals are: (i) enhanced fat solubility; (ii) enhanced interaction of catalytic-center of enzymes with fluorine-drugs; (iii) a delayed metabolism within the human body. Highly fluorinated organic substances are also used in technical applications (e.g. coatings, fire-extinguishing agents).
Due to the broad variety of fluorinated substances and increasing production volumes numerous and up to date unknown fluorine-species are most likely to be present in the (aquatic) environment. Analytical methods to assess the degree of contamination of surface waters with organically bound fluorine are highly needed and up to now only combustion ion chromatography based method is available, which is relatively laborious.
Since a few years’ high resolution-continuum source-graphite furnace atomic absorption spectrometers (HR-CS-GFAAS) are commercially available from Analytik Jena. By means of this technique, the detection of high resolution molecular absorption spectra (MAS) is enabled. Thus, fluoride is detectable upon the addition of a modifier and the formation of a diatomic molecule (e.g. GaF). Just recently, we applied this technique for total fluorine (mainly dissolved fluoride) analysis in river water samples.
In the present work a HR-CS-GFMAS method for extractable organically bound fluorine (EOF) analysis in surface water samples was developed by us. The method is based on SPE extraction of organically bound fluorine even in the presence of high fluoride concentrations followed by HR-CS-GFMAS analysis upon elution. Due to high enrichment factors, LODs in the low ng/L range were achieved. We successfully applied our SPE HR-CS-GFMAS method to Rhine water samples and EOF in the range of about 50-300 ng/L was detectable.
Stable isotope ratios and trace element concentrations of fossil bones and teeth are important geochemical proxies for the reconstruction of diet and past environment in archaeology and palaeontology. However, since diagenesis can significantly alter in vivo incorporated isotope signatures and elemental compositions, it is important to understand alteration processes. Here, we present the results of in vitro alteration experiments of dental tissues from a modern African elephant molar in aqueous solutions at 30 °C and 90 °C for 4 to 63 days each. Dental cubes with ≈3 mm edge length, comprising both enamel and dentin, were placed into 2 ml of an acidic (HNO3) aqueous solution (pH ≈1) enriched with different isotopes (25Mg, 44Ca, 67Zn, 86Sr). Element and isotope distribution profiles across the different dental cubes were measured with LA-(MC-)ICP-MS and EMPA, while potential changes of the bioapatite crystal structure were characterised by Raman spectroscopy. Isotope ratios measured by LA-(MC-)ICP-MS revealed an alteration of the outer ≈200-300 μm of the enamel in all experiments. Dentin was fully altered after one week (at 90 °C) and the tracer solution started to penetrate through the dentin even into the innermost enamel. However, the central part of the enamel remained unaltered. The Raman spectra suggest a strong recrystallization in the dentin and in the outer ≈40 μm of the enamel and a partial demineralisation of the outer rim of the cubes. Our results indicate that independent of time, temperature or low initial pH, enamel apatite shows a high resistance against the experimental alteration in structure and isotopic composition, in contrast to dentin apatite.