qNMR - The handbook
(2023)
Quantitative NMR (qNMR) has been around for a long time, but also has great potential to solve future problems in any quantitative analysis. As a primary method, it differs fundamentally from chromatographic methods: it is better described as a quantum mechanical balance. Succesful implementation of qNMR requires certain attention to detail. "qNMR - the handbook" is intended to be a guide for analysts to help understand the fundamental principles of NMR and the significant points relating to its implementation for quantitation. Regulatory considerations of qNMR adoption are explained. NMR fundamentals are explained to provide understanding. Together with many useful examples, the book is a compelling addition to the laboratory's reference library, providing all the tools that any practitioner should know to successfully implement qNMR.
The authors are qNMR pioneers and come from a variety of backgrounds including business, government and academia.
The application of compact NMR instruments to continuously flowing samples at higher temperatures or exothermically reacting mixtures is limited by the temperature sensitivity of permanent magnets. Typically, such temperature effects directly influence the achievable magnetic field homogeneity and hence measurement quality. The internal-temperature control loop of the magnet and instruments is typically not designed for compensation of such effects. Passive insulation is restricted by the small dimensions within the magnet borehole.
Here, we present a design approach for active heat shielding with the aim of variable temperature control of NMR samples for benchtop NMR instruments using a compressed airstream which is variable in flow and temperature. Based on the system identification and surface temperature measurements using an optical thermo-graphy setup, a model predictive control was set up to minimize any disturbance effect on the permanent magnet from the probe or sample temperature.
This methodology will facilitate the application of variable-temperature shielding and, therefore, potentially extend the application of compact NMR instruments to flowing samples at higher temperature than the permanent magnet.