The project “Intelligence for Soil” (I4S) aims at the design of an integrated system for improvement of soil functions and fertilizer recommendations. This system is composed by different sensors that will provide a detailed assessment of soil properties and processes, which are prerequisites for a site-specific, resource-saving and ecofriendly soil management, considering the soil as a sustainable resource for the bioeconomy. One of these sensors will be an energy-dispersive X-Ray Fluorescence spectrometer. It is a non-destructive technique suitable for in-situ measurements due to a minimum sample preparation and it allows fast multielement analyses. In this work, an automatic system has been developed using a handheld equipment from Olympus (Vanta C series). A polypropylene (PP) film was used to protect the measuring window of the device from dust and possible cross-contamination. To control the stepper motor that unrolls the PP film, a microcontroller was used to ensure that a piece of clean PP is in front of the measuring window for each new analysis.
Preliminary calibration studies using pre-defined methods, Geochem and Soil Methods, were performed with the following Certified Reference Materials (CRMs): NRC Till 1-3, NIST 2710, BAM U110, ERM CC141, BCR 142R, IAEA Soil 7. The CRMs were measured 10 times at different spots. The spot size was 10 mm in diameter, irradiation time was 60 ? s. The averaged data from X ray characteristic emission line intensities for Al, P, Si, Ca, Fe, Mn, Zn, Cu and Ni Kα were in close agreement with certified mass fraction data. The linear correlation coefficients (r) ranged from 0.852 for P to 0.999 for Mn. A second round of calibration studies were performed with the following CRMs: GBW07402, GBW 07405, NCS DC 73023, NCS DC 73030, NCS DC 85109, NCS DC 87104, NIST 1646a, NIST 2704, NIST 2710, NRC Till 1-3, VS 2498-83, and the same pre-defined methods and experimental parameters were employed. The results of the 13 CRMs were now compared with the values obtained by a validated WDXRF method. The linear correlation coefficients (r) ranged from 0.998 for Ca to 0.999 for Zn. Other elements such as S, K and Ti can also be properly determined, but validation still requires more robust calibration models.
Further calibration studies will be performed in order to circumvent matrix effects and to guarantee reliable results. Besides that, the automatic system will be placed on a mobile sensor platform and the system will be tested in the field. A guidance for on site-specific fertilization integrating the results obtained from different sensors placed on the platform is expected.
Die vom BMBF unterstütze Forschungsinitiative BonaRes entwickelt Strategien um Boden als nachhaltige Ressource für die Bioökonomie zu nutzen. Das Projekt Intelligence for Soil (I4S) ist ein Teil dieser Initiative und beschäftigt sich mit der Entwicklung eines integrierten Systems zur ortsspezifischen Düngung. Ein Ziel dieses Projektes ist die Etablierung einer mobilen Sensorplattform zur Bodenkartierung. Aufgabe der BAM ist dabei das Installieren und die Optimierung eines RFA Sensors für die online Messung direkt auf dem Feld. Die RFA ist eine zerstörungsfreie Technik, die kaum Probenvorbereitung benötigt und einen schnelle Multielement Analyse ermöglicht. Das schnelle Kartieren des Bodens ermöglicht später eine ortsspezifische Düngung mit Nährstoffen wie z.B. K, Ca und P.
Within the framework of precision agriculture, the determination of various soil properties is moving into focus, especially the demand for sensors suitable for in-situ measurements. Energy-dispersive X-ray fluorescence (EDXRF) can be a powerful tool for this purpose. In this study a huge diverse soil set (n = 598) from 12 different study sites in Germany was analysed with EDXRF. First, a principal component analysis (PCA) was performed to identify possible similarities among the sample set.
Clustering was observed within the four texture classes clay, loam, silt and sand, as clay samples contain high and sandy soils low iron mass fractions. Furthermore, the potential of uni- and multivariate data evaluation with partial least squares regression (PLSR) was assessed for accurate Determination of nutrients in German agricultural samples using two calibration sample sets. Potassium and iron were chosen for testing the performance of both models. Prediction of these nutrients in 598 German soil samples with EDXRF was more accurate using PLSR which is confirmed by a better overall averaged deviation and PLSR should therefore be preferred.
As part of the BonaRes research initiative, funded by the BMBF, strategies are being developed to use soil as a sustainable resource in the bio economy. The interdisciplinary subproject I4S - “Intelligence for soil” - is responsible for the development of an integrated system for site-specific management of soil fertility. For this purpose, a platform is constructed and various sensors are installed. Real-time data will be summarised in models and decision-making algorithms will be used to control fertilisation and accordingly improve soil functions. Aim of the BAM in the frame of I4S is the characterisation of an energy-dispersive X-ray fluorescence (EDXRF) based sensor for robust determination of plant essential nutrients in soil.
First a principal component analysis (PCA) was used to identify outliers and to observe the largest variance within the German soil samples. It could be monitored that splitting of the samples was due to their iron content. Given that clay samples contain high amounts of iron and sandy samples low amounts, a classification of the samples by their soil texture according to VD LUFA was possible. Considering the complex composition of soil, a matrix-specific calibration was carried out by univariate and multivariate data analysis. The figures of merit demonstrated that a more robust calibration model with negligible matrix effects can be obtained by a multivariate approach using partial least squares regression (PLSR). A better correlation between predicted values compared to reference values for German soil samples was observed for the chemometric calibration model than for the univariate one.
Different factors can affect the received calibration models such as moisture and particle size distribution which is especially important due to later online analysis.
In first studies the influence of moisture on the detection of plant essential nutrients was investigated. With increasing water content, the characteristic fluorescence peaks decrease and start to increase again at a water content of 15 %. With lower moisture content the soil agglomerates which leads to lower packing of the sample, resulting in a rougher surface which negatively influence the signals. Whereas, agglomerates are not formed at higher water content. This allows the sample to be packed more tightly thus a smoother surface and a better homogeneity is obtained.
Furthermore, particle size distribution leads to significantly higher uncertainties and lower signals when comparing grounded (< 500 μm) and not grounded (< 2 mm) samples. This can be explained by amplifying of the already known inhomogeneity of soils.
Both factors must be included in the chemometric PLSR to obtain robust calibration models for each macro and micro nutrient.