Measurements of aerosol particles are vital for enforcing EU air quality regulations to protect human health, and for research on climate change effects. Although metrics such as PM10 and PM2.5 are currently in use, the level of uncertainty of aerosol metrics is too high and the traceability is insufficient. The project AEROMET, which has been started in June 2017 aims at implementing improvements in a) the uncertainty of particle mass, size and number concentration measurements and b) in the characterization of regulated components in airborne particles. Both are demanded by existing networks within the EU as well as by global atmospheric research.
On-site measurement campaigns
One of the objects is the application of mobile x-ray spectroscopy techniques combined with aerosol sampling techniques for quantifying particle compositions in the field for real time analysis. During two in-field measurement campaigns in Budapest, Hungary in May 2018 and Cassino, Italy in September 2018 the size dependent mass concentrations of specific elements in ambient aerosols were monitored under dynamic conditions.
Typically, airborne particles are sampled on filter substrates. During this project new sampling methods with specially designed substrate holders for an in-situ TXRF analysis were developed and applied for the first time. This approach allows a direct time and size resolved analysis without laborious digestion steps and a reduced risk of contamination.
Aerosol particles were sampled in a 13-stage DLPI impactor - size range from 0,03 µm to 10 µm - which was equipped with special adapters for acrylic discs of 30 mm diameter, serving as substrates. TXRF analysis was performed on site with the transportable spectrometer S2 PICOFOX (Bruker Nano GmbH) equipped with a Mo X-ray tube and a 30 mm² Silicon Drift Detector (SDD). Excitation conditions were 50 kV, 600 µA, measurement time 1000 s. Quantification was based on internal standardization using 50 ng of Y in solution, which was pipetted into the centre of the discs prior to sampling.
At moderate air pollution levels, i.e. PM10 ~ 20 µg/m³, sampling times of less than 2 hours were enough for the detection of elements in different particle size bins. The in-situ approach and the high sensitivity of TXRF enables the observation of rather quick changes in the quantity and distribution of elements in an ambient aerosol on the day of sampling, as the below example from the Cassino field campaign on 11 Sept. 2018 shows: The analysis of the morning and afternoon sampling shifts reveals the occurrence of the elements Fe, Ca and Si in different size bins as well as their significant temporal change in respective mass concentrations over the day while the distributions of several other elements in the aerosol remain unchanged.
The validation of these results by backup measurements is planned.
Simple particulate matter sensors are gaining popularity due to their low price, easy handling and good temporal resolution. In this presentation, we report on the performance of a Sharp optical PM sensor GP2Y1010AU0F, which costs less than 15 €.
The sensor is built around an infrared emitting diode (ILED) and a phototransistor detecting the light scattered from the aerosol particle. An electronic circuit shapes the detected light in a pulsed signal. The manufacturer advises sampling the output signal 280 microseconds after the ILED pulse. The measured output voltage is an indicator of dust concentration.
We have built two identical simple PM monitoring devices using Raspberry Pi 3 computer interfacing the PM sensor with Microchip’s MCP3002 ADC via SPI. The ADC is capable of more than 100 ksamples/s at 10-bit resolution. The Rpi3 was pulsing the sensor at 10Hz, digitizing and saving the data and sending the results wirelessly.
Sensor’s output pulse shape was sampled with 10 microsecond time steps and saved, thus making offline analysis possible. A time jitter of output pulses can be observed and suggests a peak fitting as a better approach to the signal readout compared to the single sampling at a fixed time after pulse triggering We compared both methods.
A mobile Bruker S2 Picofox TXRF spectrometer has been used in two field campaigns within the EMPIR env07 AEROMET project for the on-site analysis of cascade impactor aerosol samples.The results show that even at moderate air pollution levels – i.e.PM10 fairly below 20 μg/m³ - element mass concentrations in air in the range of 100 pg/m³could be measured in up to 13 size bins after sampling times of less than only 0.5 days.