Synthetic peptide pools are used in antigen-specific T-cell assays, which are an important part in vaccine and immunotherapeutic clinical trials. As the analytical characterization is challenging due to the similarity of the single peptides or is expensive due to isotope labeled standards, usually only a pre-characterization of the single peptides is performed. However, a regular quality control of the peptide mix would be highly desirable. Therefore, a cost-efficient high performance liquid chromatography-high resolution mass spectrometry (HPLC-HRMS) method for quality control of a model peptide pool is developed.
Peptides were synthesized using peptides&elephants proprietary libraries of individual peptides (LIPS) technology and purified by reversed-phase chromatography to > 90% each. The lyophilized single peptides were combined to a model peptide pool and analyzed by reversed-phase high-performance capillary liquid chromatography coupled to an orbitrap mass spectrometer. Separation was performed on a capillary reversed phase column (2 μm, ID x L 300 μm x 150 mm) with a linear gradient of acetonitrile + 0,05% trifluoroacetic acid.
For the separation of a model peptide pool the additive, additive concentration, the gradient elution and the temperature were optimized. Different quantification approaches were tested. Identification was performed by high resolution mass spectrometry in which extracted ion chromatograms (XIC) were used to confirm exact masses.
In line with the development of a new cost-efficient quality control method for the separation and identification of complex synthetic peptide pools, varied HPLC parameters highlighted their influence on chromatographic resolution and peptides were identified with high mass accuracy.
Synthetic peptide pools are used in antigen-specific T-cell assays, which are an important part in vaccine and immunotherapeutic clinical trials. As the analytical characterization is challenging due to the similarity of the single peptides or is expensive due to isotope labeled standards, usually only a pre-characterization of the single peptides is performed. However, a regular quality control of the peptide mix would be highly desirable. Therefore, a cost-efficient high performance liquid chromatography-high resolution mass spectrometry (HPLC-HRMS) method for quality control of a model peptide pool is developed.
Peptides were synthesized using peptides&elephants proprietary libraries of individual peptides (LIPS) technology and purified by reversed-phase chromatography to > 90% each. The lyophilized single peptides were combined to a model peptide pool and analyzed by reversed-phase high-performance capillary liquid chromatography coupled to an orbitrap mass spectrometer. Separation was performed on a capillary reversed phase column (2 μm, ID x L 300 μm x 150 mm) with a linear gradient of acetonitrile + 0,05% trifluoroacetic acid.
After optimizing the injection mode, the gradient elution, the temperature and the additives a model peptide pool was separated. The extracted ion chromatogram (XIC) was studied to confirm the exact masses.
By combination of capillary HPLC and HRMS a new cost-efficient quality control method could be developed for the separation and identification of complex synthetic peptide pools.