Standardisation of ellipsometry is mainly based on three documents: pioneering work at NIST [1] on film standards, a publicly available specification and stan-dard testing procedure [2] for real-world samples by BAM and a standard specification for integrated metrology by SEMI [3]. Other activities refer to inter-laboratory comparisons for selected layer/substrate systems and analytical methods including ellipsometry [4] and the development of combined standards for thickness d and step height h [5].
Generally, precision of ellipsmetry in terms of raw data Ψ and Δ meets the major demands for all ellipsometric principles (RPE, RAE, PME, RCE, NE).
As a result, fingerprint capabilty is al-most perfect for various layer/substrate systems. However, this is not true for accuracy in terms of physical quantities such as thickness d, optical (n, k) and dielectric constans (ε1, ε2,) as modelling is required. Hence, the layer/substrate and material dependent validation of models is nowadays a key issue.
In contrast to the sub-nanometer sensi-tivity in vertical direction, the effective lateral resolution of almost all ellipso-metric principles except for imaging null
ellipsometry is rather poor, usually be-yond 100 µm.
Hence, nanometer-scaled homogeneity testing and the measurement of micro- and nanopatterned structures is not feasible by standard configurations.
The synthesis of polymerlike amorphous carbon (a-C:H) thin-films by microwave excited collisional hydrocarbon plasma process is reported. Stable and highly aromatic a-C:H were obtained containing significant inclusions of poly(p-phenylene vinylene) (PPV). PPV confers universal optoelectronic properties to the synthesized material. That is a-C:H with tailor-made refractive index are capable of becoming absorption-free in visible (red)-near infrared wavelength range. Production of large aromatic hydrocarbon including phenyl clusters and/or particles is attributed to enhanced coagulation of elemental plasma species under collisional plasma conditions. Detailed structural and morphological changes that occur in a-C:H during the plasma synthesis are also described.