Validation report on NMR
(2017)
The departure from the current automation landscape to next generation automation concepts for the process industry has already begun. Smart functions of sensors simplify their use and enable plug and play integration, even though they may appear to be more complex at first sight. Smart sensors enable concepts like self-diagnostics, self-calibration, and self-configuration/ parameterization whenever our current automation landscape allows it.
Here we summarize the currently discussed general requirements for process sensors 4.0 and introduce a smart online NMR sensor module as example, which was developed for an intensified industrial process funded by the EU’s Horizon 2020 research and innovation programme (www.consensspire.eu).
The recent developments of inovative compact NMR spectrometers are therefore remarkable due to their possible application in process analytical technology (PAT) in an industrial environment without high maintenance requirements. spectroscopic methods, NMR spectroscopy offers some unique features for online reaction monitoring. However, those benefits have not been exploited for PAT applications so far. Within the CONSENS Project, the challenge to adapt a commercially available benchtop NMR spectrometer to the full requirements of an automated chemical production environment was tackled successfully. was provided in an explosion proof housing and involves a compact NMR spectrometer together with an automated data acquisition and evaluation unit, flow control, as well as data communication.
In case study one of the CONSENS project, two aromatic substances were coupled by a lithiation reaction, which is a prominent example in pharmaceutical industry. The two aromatic reactants (Aniline and o-FNB) were mixed with Lithium-base (LiHMDS) in a continuous modular plant to produce the desired product (Li-NDPA) and a salt (LiF). The salt precipitates which leads to the formation of particles. The feed streams were subject to variation to drive the plant to its optimum.
The uploaded data comprises the results from four days during continuous plant operation time. Each day is denoted from day 1-4 and represents the dates 2017-09-26, 2017-09-28, 2017-10-10, 2017-10-17.
In the following the contents of the files are explained.
Monitoring chemical reactions is the key to chemical process control. Today, mainly
optical online methods are applied. NMR spectroscopy has a high potential for direct
loop process control. Compact NMR instruments based on permanent magnets
are robust and relatively inexpensive analysers, which feature advantages like low
cost, low maintenance, ease of use, and cryogen-free operation. Instruments for
online NMR measurements equipped with a flow-through cell, possessing a good
signal-to-noise-ratio, sufficient robustness, and meeting the requirements for
integration into industrial plants (i.e., explosion safety and fully automated data
analysis) are currently not available off the rack.
Intensified continuous processes are in focus of current research. Flexible (modular)
chemical plants can produce different products using the same equipment with short
down-times between campaigns and quick introduction of new products to the
market. In continuous flow processes online sensor data and tight closed-loop control
of the product quality are mandatory. If these are not available, there is a huge risk of
producing large amounts of out-of-spec (OOS) products. This is addressed in the
European Unionʼs Research Project CONSENS (Integrated Control and Sensing)
by development and integration of smart sensor modules for process monitoring and
control within such modular plant setups.
The presented NMR module is provided in an explosion proof housing of 57 x 57 x
85 cm module size and involves a compact 43.5 MHz NMR spectrometer together
with an acquisition unit and a programmable logic controller for automated data
preparation (phasing, baseline correction) and evaluation. Indirect Hard Modeling
(IHM) was selected for data analysis of the low-field NMR spectra. A set-up for
monitoring continuous reactions in a thermostated 1/8” tubular reactor using
automated syringe pumps was used to validate the IHM models by using high-field
NMR spectroscopy as analytical reference method.
Process analytical techniques are extremely useful tools for chemical production and manufacture and are of particular interest to the pharmaceutical, food and (petro-) chemical industries.
Today, mainly optical online methods are applied. NMR spectroscopy has a high potential for direct loop process control. Compact NMR instruments based on permanent magnets are robust and relatively inexpensive analysers, which feature advantages like low cost, low maintenance, ease of use, and cryogen-free operation. Instruments for online NMR measurements equipped with a flow-through cell, possessing a good signal-to-noise-ratio, sufficient robustness, and meeting the requirements for integration into industrial plants (i.e., explosion safety and fully automated data analysis) are currently not available off the rack.
A major advantage of NMR spectroscopy is that the method features a high linearity between absolute signal area and sample concentration, which makes it an absolute analytical comparison method which is independent of the matrix. This is an important prerequisite for robust data evaluation strategies within a control concept and reduces the need for extensive maintenance of the evaluation model over the time of operation. Additionally, NMR spectroscopy provides orthogonal, but complimentary physical information to conventional, e.g., optical spectroscopy. It increases the accessible information for technical processes, where aromatic-toaliphatic conversions or isomerizations occur and conventional methods fail due to only minor changes in functional groups.
As a technically relevant example, the catalytic hydrogenation of 2-butyne-1,4-diol and further pharmaceutical reactions were studied using an online NMR sensor based on a commercially available low-field NMR spectrometer within the framework of the EU project CONSENS (Integrated Control and Sensing).
The goal of this work is to identify the optimal operating input for a lithiation reaction that is performed in a highly innovative pilot scale continuous flow chemical plant in an industrial environment, taking into account the process and safety constraints. The main challenge is to identify the optimum operation in the absence of information about the reaction mechanism and the reaction kinetics. We employ an iterative real-time optimization scheme called modifier adaptation with quadratic approximation (MAWQA) to identify the plant optimum in the presence of plant-model mismatch and measurement noise. A novel NMR PAT-sensor is used to measure the concentration of the reactants and of the product at the reactor outlet. The experiment results demonstrate the capabilities of the iterative optimization using the MAWQA algorithm in driving a complex real plant to an economically optimal operating point in the presence of plant-model mismatch and of process and measurement uncertainties.