Within the framework of a German government project (initiated by the Federal Environment Agency) to check the compliance of commercially available batteries with the German Battery Ordinance concerning their heavy metal contents, 18 different types of commercially available zinc-air button cells were analysed for their cadmium, lead and mercury contents. After microwave assisted dissolution with aqua regia, Cd and Pb were determined using inductively coupled plasma mass spectrometry (ICP-MS), and Hg was determined using inductively coupled plasma optical emission spectrometry (ICP OES) and atomic absorption spectrometry. Cd contents were found to be much lower than the permitted limits; Pb contents were also found to be below the limits. Hg contents were found to be near the limits, and in one case the limit was exceeded.
The objective of this work was to test the compliance of commercially available batteries with the German Battery Ordinance, a project of the German government that was initiated by the Federal Environment Agency. Different types of commercially available dry cells were analysed for their cadmium, lead and mercury contents. The dry cells underwent mechanical pre-treatment, separation of the different components and microwave-assisted digestion before determination of the heavy metals. Mercury is sometimes added to prevent the generation of gaseous hydrogen from the electrochemical process. Lead could be present since it is sometimes used as an alloying element of zinc. Cadmium has no technical importance and is an undesirable impurity. None of the batteries contained higher heavy metal mass fractions than the permissible limits.
An intercomparison study of analytical methods for the determination of magnesium in low alloy steel
(2012)
In an intercomparison study three low alloy steel materials were analyzed on their content of the trace element Mg, and five different analytical techniques were used, namely spark-OES, inductively coupled plasma optical emission spectrometry (ICP-OES), inductively coupled plasma time of flight mass spectrometry (ICP-TOFMS), inductively coupled plasma quadrupole mass spectrometry (ICP-QMS), and glow discharge mass spectrometry (GD-MS). Solid steel discs were used for analysis with spark-OES and GD-MS. For the analyses with ICP-OES, ICP-TOFMS, and ICP-QMS steel chips were wet-digested in aqua regia, and the wet-digestion was performed either in polypropylene tubes placed in a heating block or in Teflon pressure vessels using a microwave assisted system. The Mg concentrations obtained for the three steel materials were: 2.0, 2.8, and 10.3 µg g-1, respectively, and the spread in results was acceptable, giving RSD values in the range of 2030%.
The certification of a boron carbide reference material for chemical composition is described. The mass fractions of 16 elements and two boron species are certified in an international interlaboratory comparison with 35 participating laboratories from six different countries. Beside chemical characterization the certification process includes homogeneity and stability testing of the candidate material boron carbide (type F360, 305M422). Details of the analytical methods used for chemical characterization and of the calculation of the uncertainties of the certified mass fractions are given. The new reference material ERM®-ED102 with certified mass fractions of 18 parameters (elements and species) and of the amount fraction of the isotope 10B is a valuable tool for laboratories working in the field of advanced ceramic materials analysis to improve their analytical results.
An alternative method for lithium isotope amount ratio analysis based on a combination of high-resolution atomic absorption spectrometry and spectral data analysis by machine learning (ML) is proposed herein. It is based on the well-known isotope shift of approximately 15 pm for the electronic transition 22P←22S at around the wavelength of 670.8 nm, which can be measured by the state-of-the-art high-resolution continuum source graphite furnace atomic absorption spectrometry. For isotope amount ratio analysis, a scalable tree boosting ML algorithm (XGBoost) was employed and calibrated using a set of samples with 6Li isotope amount fractions, ranging from 0.06 to 0.99 mol mol–1, previously determined by a multicollector inductively coupled plasma mass spectrometer (MC-ICP-MS). The calibration ML model was validated with two certified reference materials (LSVEC and IRMM-016). The procedure was applied toward the isotope amount ratio determination of a set of stock chemicals (Li2CO3, LiNO3, LiCl, and LiOH) and a BAM candidate reference material NMC111 (LiNi1/3Mn1/3Co1/3O2), a Li-battery cathode material. The results of these determinations were compared with those obtained by MC-ICP-MS and found to be metrologically comparable and compatible. The residual bias was −1.8‰, and the precision obtained ranged from 1.9 to 6.2‰. This precision was sufficient to resolve naturally occurring variations, as demonstrated for samples ranging from approximately −3 to +15‰. To assess its suitability to technical applications, the NMC111 cathode candidate reference material was analyzed using high-resolution continuum source atomic absorption spectrometry with and without matrix purification. The results obtained were metrologically compatible with each other.
The therapeutic dose of lithium (Li) compounds, which are widely used for the treatment of psychiatric and hematologic disorders, is close to its toxic level; therefore, drug monitoring protocols are mandatory. Herein, we propose a fast, simple, and low-cost analytical procedure for the traceable determination of Li concentration in human serum, based on the monitoring of the Li isotope dilution through the partially resolved isotope shift in its electronic transition around 670.80 nm using a commercially available high-resolution continuum source graphite furnace atomic absorption spectrometer. With this technique, serum samples only require acidic digestion before analysis. The procedure requires three measurements—an enriched 6Li spike, a mixture of a certified standard solution and spike, and a mixture of the sample and spike with a nominal 7Li/6Li ratio of 0.82. Lanthanum has been used as an internal spectral standard for wavelength correction. The spectra are described as the linear superposition of the contributions of the respective isotopes, each consisting of a spin-orbit doublet, which can be expressed as Gaussian components with constant spectral position and width and different relative intensity, reflecting the isotope ratio in the sample. Both the spectral constants and the correlation between isotope ratio and relative band intensity have been experimentally obtained using commercially available materials enriched with Li isotopes. The Li characteristic mass (mc) obtained corresponds to 0.6 pg. The procedure has been validated using five human serum certified reference materials. The results are metrologically comparable and compatible to the certified values. The measurement uncertainties are comparable to those obtained by the more complex and expensive technique, isotope dilution mass spectrometry.
In this work, we employed glow discharge optical emission spectrometry (GD-OES) depth profiling as a fast and semi-quantitative method to investigate the aluminum (Al) current collector degradation in commercial lithium cobalt oxide (LCO) pouch cells with no Al2O3 pretreatment. After battery aging, a heterogeneous deposit was found on the surface of the cathode. Gray hotspot areas within an extensive pale white region were formed. Consistent with energy dispersive X-ray (EDX) analysis of micro-cross sections milled via targeted focused ion beam (FIB), an Al-containing layer of approximately 3 µm can be observed using GD-OES. We attribute one main cause of this layer is the degradation of the Al current collector. The nonuniform growth of this layer was investigated by performing GD-OES depth profiling at different in-plane positions. We found that the gray area has a higher mass concentration of Al, probably in metallic form, whereas the white area was probably covered more homogeneously with Al2O3, resulting from the inhomogeneous distribution of the pitting positions on the current collector. Compared to FIB-EDX, GD-OES enables a faster and more convenient depth profile analysis, which allows the more productive characterization of lithium-ion batteries (LIBs), and consequently benefits the development of preferable battery materials.