Filtern
Dokumenttyp
Schlagworte
- PCB (2)
- Analysis (1)
- Bodenluft (1)
- Chrom(VI) (1)
- Cyanide (1)
- Evaluierung von Analyseverfahren (1)
- ICP-OES (1)
- Messunsicherheit (1)
- Method validation (1)
- OCP (1)
Evaluation of methods for the analysis of soils according to § 8 of the German Federal Soil Protection Act (BBodSchG)
The German Federal Soil Protection Act establishes the conditions for effective soil protection and remediation of contaminated sites. In the German Federal Soil Protection and Contaminated Sites Ordinance (BBodSchV) limiting values and methods for the determination of environmentally harmful substances in soils are fixed. The aim of this project was to evaluate selected methods by comparative investigations and proficiency testing schemes. New methods for the determination of inorganic and organic pollutants should be developed and compared with the existing standard
methods. Despite the widespread use of ICP-OES up to now no national or international standard is available for the determination of heavy metals in soil extracts with this analytical method. The influence of spectral and non spectral interferences on the accuracy of the results was systematically investigated, especially based on aqua regia soil extracts. Analyte concentration equivalents were determined, too. Based on these results a proposal for a standard was submitted to the corresponding working group of ISO TC 190.[...]
A procedure for the quantification of the PCB marker congeners 28, 52, 101, 138, 153, 180 in plastic waste form the recycling of cables in the range of content between 0.1 mg kg-1 and 20 mg kg-1 was validated. Representative reference materials for laboratory analyses were produced using granulates from different recycling processes. The method development included aspects of grinding, sieving, homogenising as well as the comparison of extraction methods and chromatographic conditions. The validated procedure foresees grinding of the technical sample to an upper particle size limit of 0.50 mm followed by extraction with an unpolar solvent such as n-hexane and gas chromatographic determination. LOD and LOQ for individual congeners ranged between 0.01 mg kg-1 and 0.08 mg kg-1. Electron capture detection (ECD) and mass selective detection (MS) were equivalent if chromatographic conditions were appropriately selected. The application of MS was advantageous in presence of significant amounts of interfering components which have been found in some cables.
Three reference materials containing the marker congeners in the range of 0.1 mg kg-15 mg kg-1 were submitted to a final validation intercomparison with 23 selected field laboratories using the developed method. Relative reproducibility standard deviations varied between 15% and 33%.