Compared to the intended EU reference level of 300 Bq m−3 for indoor radon concentrations, the contribution of building materials appears to be low. Considering the recommended limit of 100 Bq m−3 by WHO, their contribution is supposed to be relevant, especially at low air exchange rates. This study as part of a two-part research project investigated the suitability of direct low level 222Rn measurement under simulated indoor conditions with commercial radon monitors and dynamically operated emission test chambers. Active measuring devices based on ionisation or scintillation chambers with 1-σ uncertainties below 8.6% at 20 Bq m−3 were found to be best suitable for a practical test procedure for the determination of radon exhalation rates of building materials. For the measurement of such low concentrations, the knowledge of the accurate device background level is essential.
Compared to the designated European upper reference level of 300 Bq/m³ for indoor radon concentrations, the contribution of building materials appears to be low. Taking into account the recommended limit of 100 Bq/m³ by WHO as well as the increased dose conversion factor, the contribution of building products is quite relevant, especially at low air change rates. Therefore in Germany a maximum contribution of building materials to the indoor radon concentration of 20 Bq/m³ is discussed.
Because measurements of low concentrations are associated with high uncertainties, radon exhalation rates are often determined in chambers without air exchange and indoor conditions are extrapolated.
This study investigated the suitability of direct low level radon measurement under simulated indoor conditions with commercially available instruments, based on the preliminary EU standard for the determination of the emission of volatile organic compounds (VOC) from building materials in test chambers.
Measurements are possible with a relative uncertainty ranging from 7 to 18%. Four major problems were identified:
1. Detector sensitivity: Two of four instruments were identified to determine low radon concentrations in a sufficiently short measuring time.
2. Thoron discrimination: Thoron was discriminated by aging the sampled air in a hose before the detector.
3. Sampling setup: An external chamber measurement was feasible as long as all main system parts were under slight overpressure.
4. Blank value: For VOC-emission testing ambient air, conditioned by filtration and sorption, is used to generate an air exchange. Regarding radon, this treatment is non-effective. If all instrument background levels have been characterized, the blank value must be measured with a second detector in parallel.
Compared to the designated European upper reference level of 300 Bq m-3 for indoor radon concentrations, the contribution of building materials appears to be low. Taking into account the recommended limit of 100 Bq m-3 by WHO as well as the increased dose conversion factor, the contribution of building products is quite relevant, especially at low air change rates. Therefore in Germany a maximum contribution of building materials to the indoor radon concentration of 20 Bq m-3 is discussed.
Because measurements of low concentrations are associated with high uncertainties, radon exhalation rates are often determined in chambers without air exchange and indoor conditions are extrapolated.
This study investigated the suitability of direct low level radon measurement under simulated indoor conditions with commercially available instruments, based on the preliminary EU standard for the determination of the emission of volatile organic compounds (VOC) from building materials in test chambers.
Measurements are possible with a relative uncertainty ranging from 7 to 18 %. Four major problems were identified:
1. Detector sensitivity: Two of four instruments were identified to determine low radon concentrations in a sufficiently short measuring time.
2. Thoron discrimination: Thoron was discriminated by aging the sampled air in a hose before the detector.
3. Sampling setup: An external chamber measurement was feasible as long as all main system parts were under slight overpressure.
4. Blank value: For VOC-emission testing ambient air, conditioned by filtration and sorption, is used to generate an air exchange. Regarding radon, this treatment is non-effective. If all instrument background levels have been characterized, the blank value must be measured with a second detector in parallel.