- Calibration (1) (entfernen)
Gas sensors in linear form based on the measuring principle of gas selective permeability through a membrane were developed and introduced for the detection and quantification of gas concentrations. A current field study focuses on measuring CO2 concentrations for a spatially distributed monitoring of subsurface CO2 gas storage sites in near real time. A 400 m² test site and a corresponding laboratory system were built up to characterize, validate, and optimize the sensor. A calibration routine was developed, which can be applied subsequently to underground installation. First measurement results indicate the potential of the method.