Trace elements of ancient gold such as Pt, give fundamental information on the circulation of the metal in the past. In the case of objects from the cultural heritage, the determination of trace elements requires non-destructive point analysis in general. These conditions and the need of good detection limits restrain the number of applicable analytical techniques.
After the development of a PIXE set-up with a selective Cu or Zn filter of 75 µm and of a PIXE-XRF set-up using a primary target of As, we tested the possibilities of spatially resolved Sy-XRF to determine Pt in gold alloys. With a Zn filter, PIXE showed a detection limit of 1000 ppm in gold while PIXE-XRF lowers this detection limit down to 80 ppm. This last value being constrained by the resonant Raman effect produced on gold. In order to improve the detection limit of Pt keeping the non-destructiveness and access to point analysis, we developed an analytical protocol for XRF with synchrotron radiation at BESSY II, using the BAMline set-up. The L-lines of Pt were excited by a beam of energy above and below 11.564 keV and measured using a Si(Li) detector with a 50 µm Cu filter. A µ-beam of 100250 µm2 was used according to the size of the sample. The determination of the Pt content in the samples was carried out by Monte-Carlo simulation and subtraction of Au and Pt spectra obtained on pure standards. The limit of detection for Pt of 20 ppm was determined by using certified standards. The detection limits of a small set of other characteristic elements of gold were also measured using an incident energy of 33 keV.
Naturwissenschaftler arbeiten mit Archäologen, Kunsthistorikern und Restauratoren zusammen, um Materialinformationen über Kunstwerke mit den Erkenntnissen aus Altertumsforschung, Kunstgeschichte und Denkmalpflege zu verbinden. Röntgenfluoreszenz und Röntgenemission sind die Methoden der Wahl, wenn wertvolle und fragile Objekte zerstörungsfrei zu untersuchen sind.
New analyses of a series of very rare silverpoint drawings that were executed by Rembrandt Harmensz. van Rijn (16061669) which are kept today in the Kupferstichkabinett (Museum of Prints and Drawings) of the State Museums of Berlin are reported here. Analysis of these drawings requires particular attention because the study has to be fully non-destructive and extremely sensitive. The metal alloy on the paper does not exceed some hundreds of µg/cm². Therefore, synchrotron radiation induced X-ray fluorescence (SR-XRF) is together with external micro-proton-induced X-ray emission the only well-suited method for the analyses of metalpoint drawings. In some primary work, about 25 German and Flemish metalpoint drawings were investigated using spatially resolved SR-XRF analysis at the BAMline at BESSY. This study enlarges the existing FrenchGerman database of metalpoint drawings dating from the 15th and 16th centuries, as these Rembrandt drawings originate from the 17th century where this graphical technique was even rarer and already obsolete. It also illustrates how SR-XRF analysis can reinforce art historical assumptions on the dating of drawings and their connection.
We performed spatially resolved synchrotron-induced x-ray fluorescence analyses (SY-XRF) of five silverpoint drawings from the famous `Karlsruhe sketchbook´ and a silverpoint, which is assumed to be an original one used by the Renaissance artist Hans Baldung Grien (1484-1545) himself. The silverpoint, one out of a very small number of ancient styli conserved nowadays, is fixed on the so-called `Karlsruhe sketchbook´. Measurements were executed at the FLUO-TOPO beamline at the electron storage ring ANKA. It has been shown, that SY-XRF at this beamline is nondestructive and sensitive enough for the analysis of minor and major elements in the silver strokes. This study provides new insights into this graphical technique and ancient drawing instruments: results are that three drawings have been created with a common type of silverpoint containing about (10.4 ± 0.7) wt% of Cu. In all the cases, Hg at a 10% level by weight was identified in the strokes. Presence of Hg is not due to the original silver alloy, but is a result of a general alteration phenomenon of silverpoint drawings. This fact is confirmed by the analysis of the original silverpoint, which contains only small traces of Hg. The result for the copper content of the silverpoint (10.4 ± 1.7) wt% is in good agreement with that of these three drawings. It is supposable that the silverpoint found attached today served for the creation of the corresponding drawings. Moreover, the graphical technique of Baldung is found to be continual over three decades. Only one very late drawing dated to 1543 was probably realized with another silverpoint, containing about 3 wt% of Cu. This difference in the chemical composition of drawings from the late period might be used as a more precise dating indicator of Baldung drawings that are not clearly attributed to a certain period.
The propagation of synchrotron radiation through nanocapillary structures with channel sizes of 200 nm and periods in the micrometer size has been studied experimentally. It was shown that the propagation through individual capillary channels has a mode formation character. Furthermore it was shown that during the propagation through capillary channels the coherence of synchrotron radiation is partially conserved. Interference of beams propagating through different capillary channels is observed which leads to a periodically modulated distribution of the radiation intensity in a plane far from the exit of the structure. These investigations are of high relevance for the understanding of X-ray transmission through nanocapillaries and the appearance of wave properties at this size scale.
Measuring in a non-destructive way the characteristic trace elements of the metal allows following the circulation of gold in the past. The aim of this work is to probe the possibilities of X-ray fluorescence with high energy synchrotron radiation (SR-XRF) at the BAMline at BESSY II to determine the concentration of Pt in ancient gold alloys. A HP-Ge detector was used to measure the Pt K-lines excited with an incident energy of 79.5 keV. Data processing was done by subtraction of a Pt free gold standard spectrum from the spectrum of the sample. Depending on the sample composition, the MDL ranges between 40 and 90 ppm. The first results obtained for a small set of gold alloys of different thickness, size and composition showed that high energy SR-XRF is a significant method for the non-destructive determination of Pt in gold.
Trace-elements are more significant for provenancing archaeological metallic artifacts than the main components. For gold, the most promising elements are platinum group elements (PGE), Sn, Te, Sb, Hg and Pb. Several small fragments of natural Transylvanian gold placer and primary were studied by using micro-PIXE technique at the Legnaro National Laboratory AN2000 microbeam facility, Italy and at the AGLAE accelerator, C2RMF, Paris, France and by using micro synchrotron radiation X-ray fluorescence (micro-SR-XRF) at BESSY synchrotron, Berlin, Germany. The goal of the study was to identify the trace-elements, especially Sn, Sb and Te. A spectacular application to five Dacian gold bracelets authentication is presented (Sn and Sb traces).
A set of compound refractive lenses (CRLs), mounted on a silicon wafer, was obtained from the Institut für Mikrostrukturtechnik Karlsruhe. The CRLs were characterized at the BAMline at the synchrotron BESSYII in Berlin. It could be demonstrated that beam diameters below 1 µm can be obtained. The beam size accepted by the lenses is (140 × 140) µm2, thus the intensity gain in the micrometer spot, including absorption, is about 15,000. It is possible to switch between different beam energies within a few minutes. First results of scans and the application of the lenses for MicroXANES (X-ray absorption near-edge structure) measurements are presented in this paper. Also, the possibility of MicroEXAFS (extended X-ray absorption fine structure) measurements is discussed.
The content of chromium in different oxidation states in chromium-treated wood was studied with XANES (X-ray absorption near-edge structure) measurements at the Cr K absorption edge. It could be shown that wood samples treated with Cr(VI) (pine and beech) did still contain a measurable content of Cr(VI) after four weeks conditioning. If such wood samples were heat exposed for 2 h with 135 °C prior conditioning, Cr(VI) was no longer detected by XANES, indicating a complete reduction to chromium (III).
Synchrotron radiation induced X-ray fluorescence for the characterization of ancient gold objects
(2009)
A set of ten calibration materials based on acrylonitrile-butadiene-styrene terpolymer (ABS) containing the elements Br, Pb, Cd, Cr, and Hg was produced in order to control polymer samples in regard to the restriction on the use of certain hazardous substances (RoHS) directive. The materials were produced with respect to the special demands of X-ray fluorescence analysis (XRF) and inductively coupled plasma mass spectrometry in combination with laser ablation (LA-ICP-MS).
The mass fractions of all elements were 0-1500 mg/kg and the materials were produced as granulates and solid discs with a diameter of 40 mm and thicknesses of 1, 2, and 6 mm. It could be shown that the signals measured by XRF strongly depend on the thickness of polymeric samples.
Macroscopic homogeneity was assessed with XRF measuring a total of 180 samples. The microscopic homogeneities were determined for a material with average elemental mass fractions with the aid of synchrotron radiation (SR) µ-XRF and LA-ICP-MS. Sufficient macroscopic and microscopic homogeneities for all elements could be achieved. It was observed that organic additives show a better homogeneity than oxides.
XRF and LA-ICP-MS were successfully calibrated with the new materials. Limits of detection were in the range of 0.4-22 mg/kg for XRF and 2.3-26.8 mg/kg for LA-ICP-MS. The materials are considered as candidate reference materials (RM) by the Federal Institute of Materials Research and Testing (BAM, Germany).
X-ray scattering may contribute significantly to the spectral background of X-ray fluorescence (XRF) spectra. Based on metrological measurements carried out with a scanning electron microscope (SEM) having attached a well characterised X-ray source (polychromatic X-ray tube) and a calibrated energy dispersive X-ray spectrometer (EDS) the accuracy of a physical model for X-ray scattering is systematically evaluated for representative samples. The knowledge of the X-ray spectrometer efficiency, but also of the spectrometer response functions makes it possible to define a physical spectral background of XRF spectra. Background subtraction relying on purely mathematical procedures is state-of-the-art. The results produced by the analytical model are at least as reliable as those obtained by Monte-Carlo simulations, even without considering the very challenging contribution of multiple scattering. Special attention has been paid to Compton broadening. Relevant applications of the implementation of the analytical model presented in this paper are the prediction of the limits of detection for particular cases or the determination of the transmission of X-ray polycapillary lenses.
In this study the microscopic homogeneity of the newly developed plastic reference materials BAM H001-BAM H010 was investigated. The materials consist of an acrylonitryle-butadien-styrene terpolymer, doped with different amounts of the elements Br, Cd, Cr, Hg and Pb. For the quantitative determination of the degree of homogeneity, a procedure proposed by Kempenaers et al. (Fresenius J. Anal. Chem., 2001, 369, 733-737) was used. On every sample an extensive number of different points were analyzed and standard deviation for every element mentioned above was used to calculate a minimal sampling mass that is necessary to reach a certain level of uncertainty caused by inhomogeneity (mmin,5%). The experiments were taken out with synchrotron µ-XRF (SR µ-XRF) at BESSYII in Berlin and by laser ablation inductively coupled plasma mass spectroscopy (LA-ICP-MS). The calculated values for mmin,5% of both techniques showed comparable results for all elements. It could be shown that the materials are suitable for calibration of micro analytic techniques if at least 64 µg are used.
Measurements of ion distributions at a charged solid–liquid interface using X-ray standing waves (XSW) are presented. High energy synchrotron radiation (17.48 keV) is used to produce an XSW pattern inside a thin water film on a silicon wafer. The liquid phase is an aqueous solution containing Br and Rb ions. The surface charge is adjusted by titration. Measurements are performed over a pH range from 2.2–9, using the native Si oxide layer and functional (amine) groups as surface charge. The Debye length, indicating the extension of the diffuse layer, could be measured with values varying between 1–4 nm. For functionalized wafers, the pH dependent change from attraction to repulsion of an ion species could be detected, indicating the isoelectric point. In combination with the measurement of the streaming current, the surface charge of the sample could be quantified.
The Sarmizegetusa bracelets
(2010)
We present the authentication and analysis of these beautiful Dacian bracelets of the first century BC, originally pillaged by treasure hunters and recovered thanks to an international crime chase. They were originally fashioned from gold panned from the rivers or dug from the mines of Transylvania and hammered into the form of coiled snakes. The lack of context is the greatest loss, but a votive purpose is likely given their proximity to the great sacred centre at Sarmizegetusa Regia.
We report the results of an analytical investigation
on 416 silver-copper coins stemming from the Ottoman
Empire (end of 16th and beginning of 17th centuries), using
synchrotron micro X-ray fluorescence analysis (SRXRF).
In the past, analyses had already been conducted with energy
dispersive X-ray fluorescence analysis (EDXRF), scanning
electron microscopy with energy dispersive X-ray spectrometry
(SEM/EDX) and proton induced X-ray emission spectroscopy (PIXE).With this combination of techniques it was
possible to confirm the fineness of the coinage as well as to
study the provenance of the alloy used for the coins.
For the interpretation of the data statistical analysis
(principal component analysis—PCA) has been performed.
A definite local assignment was explored and significant
clustering was obtained regarding the minor and trace elements
composing the coin alloys.