A layer of Sb atoms, implanted with an energy of 400 keV and a nominal dose of 5×1016 atoms/cm2 into a high purity silicon wafer, was certified for its areal density (atoms/cm2) using Rutherford backscattering spectrometry (RBS), instrumental neutron activation analysis (INAA) and inductively coupled plasma isotope dilution mass spectrometry (ICP-IDMS) and for its isotope ratio using INAA and ICP-IDMS. Excellent agreement between the results of the different independent methods was found. In the present work, the measurements of the homogeneity of the areal density of Sb, previously determined with RBS in spots having 1 mm diameter, are improved with synchrotron X-ray fluorescence analysis: Higher precision in even smaller sample spots allows to estimate a reduced inhomogeneity of the whole batch of samples of the order of only 0.4%. Thus the uncertainty of the certified value can further be reduced. Down to fractions of a chip with 0.3×0.4 mm2 area, the areal density is now certified as (4.81±0.06)×1016 Sb atoms/cm2, where the expanded uncertainty 0.06 (coverage factor k=2) corresponds to only 1.2%. The relative merits of the different analytical methods are discussed.
The first hard x-ray beamline at BESSY II has been installed by BAM and PTB at a superconducting 7 T wavelength shifter. The main optical elements of the beamline are a double-multilayer monochromator and a double-crystal monochromator. Depending on the application, the two devices are used separately or in-line. The main applications of the monochromatic radiation with photon energies up to 60 keV are x-ray fluorescence analysis, micro computed tomography, x-ray topography, detector calibration and reflectometry. Calculable undispersed radiation up to 200 keV is available for radiometric applications.
New analyses of a series of very rare silverpoint drawings that were executed by Rembrandt Harmensz. van Rijn (16061669) which are kept today in the Kupferstichkabinett (Museum of Prints and Drawings) of the State Museums of Berlin are reported here. Analysis of these drawings requires particular attention because the study has to be fully non-destructive and extremely sensitive. The metal alloy on the paper does not exceed some hundreds of µg/cm². Therefore, synchrotron radiation induced X-ray fluorescence (SR-XRF) is together with external micro-proton-induced X-ray emission the only well-suited method for the analyses of metalpoint drawings. In some primary work, about 25 German and Flemish metalpoint drawings were investigated using spatially resolved SR-XRF analysis at the BAMline at BESSY. This study enlarges the existing FrenchGerman database of metalpoint drawings dating from the 15th and 16th centuries, as these Rembrandt drawings originate from the 17th century where this graphical technique was even rarer and already obsolete. It also illustrates how SR-XRF analysis can reinforce art historical assumptions on the dating of drawings and their connection.
A set of compound refractive lenses (CRLs), mounted on a silicon wafer, was obtained from the Institut für Mikrostrukturtechnik Karlsruhe. The CRLs were characterized at the BAMline at the synchrotron BESSYII in Berlin. It could be demonstrated that beam diameters below 1 µm can be obtained. The beam size accepted by the lenses is (140 × 140) µm2, thus the intensity gain in the micrometer spot, including absorption, is about 15,000. It is possible to switch between different beam energies within a few minutes. First results of scans and the application of the lenses for MicroXANES (X-ray absorption near-edge structure) measurements are presented in this paper. Also, the possibility of MicroEXAFS (extended X-ray absorption fine structure) measurements is discussed.
The content of chromium in different oxidation states in chromium-treated wood was studied with XANES (X-ray absorption near-edge structure) measurements at the Cr K absorption edge. It could be shown that wood samples treated with Cr(VI) (pine and beech) did still contain a measurable content of Cr(VI) after four weeks conditioning. If such wood samples were heat exposed for 2 h with 135 °C prior conditioning, Cr(VI) was no longer detected by XANES, indicating a complete reduction to chromium (III).
A set of ten calibration materials based on acrylonitrile-butadiene-styrene terpolymer (ABS) containing the elements Br, Pb, Cd, Cr, and Hg was produced in order to control polymer samples in regard to the restriction on the use of certain hazardous substances (RoHS) directive. The materials were produced with respect to the special demands of X-ray fluorescence analysis (XRF) and inductively coupled plasma mass spectrometry in combination with laser ablation (LA-ICP-MS).
The mass fractions of all elements were 0-1500 mg/kg and the materials were produced as granulates and solid discs with a diameter of 40 mm and thicknesses of 1, 2, and 6 mm. It could be shown that the signals measured by XRF strongly depend on the thickness of polymeric samples.
Macroscopic homogeneity was assessed with XRF measuring a total of 180 samples. The microscopic homogeneities were determined for a material with average elemental mass fractions with the aid of synchrotron radiation (SR) µ-XRF and LA-ICP-MS. Sufficient macroscopic and microscopic homogeneities for all elements could be achieved. It was observed that organic additives show a better homogeneity than oxides.
XRF and LA-ICP-MS were successfully calibrated with the new materials. Limits of detection were in the range of 0.4-22 mg/kg for XRF and 2.3-26.8 mg/kg for LA-ICP-MS. The materials are considered as candidate reference materials (RM) by the Federal Institute of Materials Research and Testing (BAM, Germany).