A new method has been developed for the measurement of sulfur in gas oils using a high pressure asher (HPA) for sample digestion. This, combined with a calibrated enriched isotopic reference material, and high precision mass spectrometry, makes it a powerful tool for an accurate and precise determination of the sulfur content in fossil fuels. The method described here was one of the methods used in the certification of the candidate reference materials BCR-104R, BCR-671 and BCR-672. Approximately 0.2 g of the gas oil samples were spiked with the spike isotopic reference material IRMM-646, which is enriched in 34S, and combusted in an HPA using nitric acid. The isotopes of sulfur were measured as arsenic monosulfides (AsS+) using a thermal ionization mass spectrometer equipped with a Faraday detector. The total uncertainty of the sulfur content (k = 2) is about 1%. The certified reference materials NIST SRM-2724b and NIST SRM-1619b and BCR-105 were used as part of the method validation work. The mass content of sulfur in BCR-104R was measured to be 1014 µg g-1, BCR-671 has 450 µg g-1 sulfur and BCR-672 shows a sulfur mass content of 201 µg g-1. The sulfur content of the candidate reference materials is homogeneous for the sample size used.
A comparison of high accuracy methods for the analysis of total sulfur in gas oils by three metrology institutes is described. The work contains a comparison of three different aspects of the analysis: sample digestion, instrumental measurement and isotope dilution strategies.
Thermal ionisation mass spectrometry (TIMS) in combination with high pressure asher (HPA) or Carius tube digestion has traditionally been used for such high accuracy analysis. However, this paper demonstrates that the faster, less laborious technique of ICP-MS in combination with microwave digestion is equally capable. Results from the comparison experiments show excellent agreement for fuel samples containing sulfur concentrations in the range 11200 µg g-1. This excellent agreement across the different sample preparation, measurement and isotope dilution techniques has been exploited in the certification of two new diesel fuel reference materials.
Elementanalyse mit Röntgenfluoreszenzspektrometrie direkt am Materialstrom in Produktionsprozessen gewinnt zunehmend an Bedeutung. Die Röntgenfluoreszenzspektrometrie kann aufgrund der technischen Entwicklung mittlerweile ohne aufwändige Probenvorbereitung als aussagekräftige Informationsquelle für Stoffparameter im industriellen Prozess eingesetzt werden. Der Einsatz von online Röntgenfluoreszenzanalyse zahlt sich sowohl finanziell als auch zeitlich schnell aus, da sie sowohl für den ökonomischen Einsatz von Rohmaterialien, als auch z. B. im Materialrecycling zu gebrauchen ist. In diesem Beitrag werden vier unterschiedliche Einsatzmöglichkeiten beschrieben.
In this study the microscopic homogeneity of the newly developed plastic reference materials BAM H001-BAM H010 was investigated. The materials consist of an acrylonitryle-butadien-styrene terpolymer, doped with different amounts of the elements Br, Cd, Cr, Hg and Pb. For the quantitative determination of the degree of homogeneity, a procedure proposed by Kempenaers et al. (Fresenius J. Anal. Chem., 2001, 369, 733-737) was used. On every sample an extensive number of different points were analyzed and standard deviation for every element mentioned above was used to calculate a minimal sampling mass that is necessary to reach a certain level of uncertainty caused by inhomogeneity (mmin,5%). The experiments were taken out with synchrotron µ-XRF (SR µ-XRF) at BESSYII in Berlin and by laser ablation inductively coupled plasma mass spectroscopy (LA-ICP-MS). The calculated values for mmin,5% of both techniques showed comparable results for all elements. It could be shown that the materials are suitable for calibration of micro analytic techniques if at least 64 µg are used.
Knowing the exact nutrient composition of organic fertilizers is a prerequisite for their appropriate application to improve yield and to avoid environmental pollution by over-fertilization.
Traditional standard chemical analysis is cost and time-consuming and thus it is unsuitable for a rapid analysis before manure application. As a possible alternative, a handheld X-ray fluorescence (XRF) spectrometer was tested to enable a fast, simultaneous, and on-site analysis of several elements.
A set of 62 liquid pig and cattle manures as well as biogas digestates were collected, intensively homogenized and analysed for the macro plant nutrients phosphorus, potassium, magnesium, calcium, and sulphur as well as the micro nutrients manganese, iron, copper, and zinc using the standard lab procedure. The effect of four different sample preparation steps (original, dried, filtered, and dried filter residues) on XRF measurement accuracy was examined. Therefore, XRF results were correlated with values of the reference analysis. The best R2 s for each element ranged from 0.64 to 0.92. Comparing the four preparation steps, XRF results for dried samples showed good correlations (0.64 and 0.86) for all elements. XRF measurements using dried filter residues showed also good correlations with R2 s between 0.65 and 0.91 except for P, Mg, and Ca. In contrast, correlation Analysis for liquid samples (original and filtered) resulted in lower R2 s from 0.02 to 0.68, except for K (0.83 and 0.87, respectively). Based on these results, it can be concluded that handheld XRF is a promising measuring system for element analysis in manures and digestates.
Es wurde ein Online-Analysenverfahren zur Bestimmung von Elementgehalten in Klär-schlammaschen und deren Produkten aus einem thermochemischen Verfahren entwickelt. Durch das thermochemische Verfahren können Wertstoffe aus den Aschen wieder für die Kreislaufwirtschaft nutzbar gemacht und Schadstoffe sicher entfernt werden. Das Analysensystem ist eine Methodenkombination von Online-Laser induzierter Plasma Spektroskopie (LIBS) und Online-Röntgenfluoreszenzanalyse (RFA). Robustheit und modulare Bauweise des Verfahrens sind wichtige Anforderungen um auch für andere Anwendungen nutzbar zu sein.
C, N, O, F and P can be analyzed by instrumental photon activation analysis (IPAA) including decay curve analysis. The interference of 30P (T1/2 = 149.9 s) by 15O (T1/2 = 122.2 s) can be ruled out by direct positron measurement making use of the largely different maximum β+-energies of both nuclides (3.24 MeV and 1.73 MeV, respectively). Interference by carbon (11C) can be avoided by sub-threshold activation with 17 MeV bremsstrahlung. The short half-life of 30P allows a high productivity of the method. Reliability was demonstrated in the range of 0.2%2% P (detection limit = 40 µg/g). Analysis of a certified reference material (BCR-CRM 063) yielded excellent agreement with the certified data.
The Federal Institute for materials research and testing (BAM)in cooperation with the Fachhochschule Münster has developed a new certified polymeric reference material, BAM-H010. The reference material consists of acrylonitrile–butadiene–styrene terpolymer (ABS) and has been doped with different contents of the elements Br, Cd, Cr, Hg and Pb. The monitoring of these elements is demanded in the EU directive 2002/95/EG, commonly referred to as RoHS. We have shown that the uncertainty contribution of the thermal long-term stability (ustability) is negligible compared to the contribution to the uncertainty caused by the method used for the characterization of the elemental content and the batch inhomogeneity (uhomogeneity). The stability of theABS material during exposure of the samples with hard X-radiation leads to a degradation of the ABS matrix with either long- or short-chain products.