Eingeladener Vortrag
- nein (2)
The accurate knowledge of the properties of polycapillary X-ray semi-lenses has a significant influence on the quantitative results in a widespread field of applications involving microfocus X-ray beams. A routine procedure for the characterisation of a polycapillary X-ray semi-lens with a scanning electron microscope (SEM) having attached an energy dispersive spectrometer (EDS) is presented in this paper. A key issue of the procedure consists of fitting the semi-lens in front of the EDS for spectra acquisition. Relevant semi-lens parameters such as focal distance, full width at half maximum (FWHM) of the acceptance area, and transmission are determined in parallelising mode of the semi-lens. Special attention has been paid to the calculation of the transmission.
A test material for routine performance evaluation of energy-dispersive X-ray spectrometers (EDS) is presented. It consists of a synthetic, thick coating of C, Al, Mn, Cu, and Zr, in an elemental composition that provides interference-free characteristic X-ray lines of similar intensities at 10 kV scanning electron microscope voltage. The EDS energy resolution at the C-K, Mn-Lα, Cu-Lα, Al-K, Zr-Lα, and Mn-Kα lines, the calibration state of the energy scale, and the Mn-Lα/Mn-Kα intensity ratio as a measure for the low-energy detection efficiency are calculated by a dedicated software package from the 10 kV spectrum. Measurements at various input count rates and processor shaping times enable an estimation of the operation conditions for which the X-ray spectrum is not yet corrupted by pile-up events. Representative examples of EDS systems characterized with the test material and the related software are presented and discussed.
An energy dispersive X-ray spectrometer operating with a semiconductor detector should be specified in compliance with the ISO standard 15632 [1]. Requirements for specification are: a) a general description of the spectrometer to evaluate its performance, b) the energy resolution with corresponding dead time, c) the P/B ratio in the Fe55 spectrum and, finally, d) the L/K intensity Ratio in a Ni or Cu spectrum to estimate spectrometer efficiency at low energies. Items b) to d) can be easily checked by the user. Related procedures are recommended in the annexes of the standard.