Filtern
Dokumenttyp
- Vortrag (13)
- Posterpräsentation (1)
Sprache
- Englisch (14)
Referierte Publikation
- nein (14) (entfernen)
Schlagworte
- Absolute isotope ratio (14) (entfernen)
Organisationseinheit der BAM
Eingeladener Vortrag
- nein (13)
Isotope reference materials
(2022)
The variation of isotope ratios is increasingly used to unravel natural and technical questions. In the past, the investigation and interpretation of such variations was the field of a limited number of experts. With new upcoming techniques and research topics in the last decades, such as provenance or food authenticity studies, the number of published isotope data strongly increased. Instrumental developments such as the enhancement of inductively coupled plasma mass spectrometers (ICP-MS) from an instrument for simple quantitative analysis to highly sophisticated isotope ratio machines influenced this process significantly. While in former times only experts in mass spectrometry were able to produce reliable isotope data, nowadays many laboratories, never been in touch with mass spectrometry before, produce isotope data with an ICP-MS. Isotope reference materials (iCRM) are indispensable to enable a reliable method validation or in rare cases even SI-traceability. The fast development and the broad availability of ICP-MS also lead to an expansion of the classical research areas and new elements are under investigation. Irrespective of the investigated element or the knowledge of the user all isotope ratio applications require reference materials either for correction of instrumental isotope fractionation, for method validation or to provide a common accepted basis as needed for delta measurements. This presentation will outline the basic principles and illustrate the urgent need for new iCRMs. Consequently, the production and certification of iCRMs will be discussed and illustrated by examples of already completed certification projects. Finally, plans for future iCRMs to be produced at BAM will be presented.
This presentation provides a short introduction to the Commission on Isotopic Abundances and Atomic Weights (CIAAW). It describes the role of the Commission and provides an insight into its work and the corresponding principal tasks. Finally, it provides the reader with the latest achievements and with the most recent publications.
The reference method for obtaining absolute isotope ratios still is the isotope mixture approach. Due the huge efforts required the full isotope mixture approach is applied only by a few institutes worldwide. To enable an IRWG key comparison with a sufficiently large number of participants a proposal for absolute cu isotope ratios is presented where participants will be provided with the enriched isotopes, the isotope mixtures and the samples. In parallel a pilot study will be organized where alternative approaches for obtaining absolute Cu isotope ratios can be applied.
In the past, δ26/24Mg measurements were referenced to NIST SRM 980, the initial zero of the δ26/24Mg scale. With the development of MC-ICPMS, the detection of small but measurable isotopic differences in different chips of SRM 980 became apparent. To solve this problem a suite of magnesium isotope reference materials, ERM-AE143, -AE144 and -AE145, has been certified in a first study by applying an ab initio calibration for absolute Mg isotope ratios without any a priori assumptions, a procedure which fulfils all requirements of a primary method of measurement. We could achieve for the first time measurement uncertainties for isotope amount ratios close to the typical precision of magnesium delta values, δ26/24Mg, which are at the 0.1 ‰ level (2SD). In addition, it was demonstrated that commonly used fractionation laws are invalid for correcting Mg isotope ratios in multi-collector ICPMS as they result in a bias which is not covered by its associated uncertainty. Depending on their type, fractionation laws create a bias up to several per mil, with the exponential law showing the smallest bias between 0.1 ‰ to 0.7 ‰.
With these isotope reference materials, it is possible to establish SI-traceability for magnesium delta measurements. To realize this, we organized a second study within which five expert laboratories participated to cross-calibrate all available magnesium isotope standards, which are NIST SRM 980, IRMM-009, ERM-AE143, ERM-AE144, ERM-AE145 and the standards DSM3 and Cambridge-1. The mean δ26/24Mg values for the individual iRMs, calculated from the laboratory means show 2 SD reproducibilities varying between 0.025 and 0.093 ‰. Propagated measurement uncertainties suggest a standard uncertainty of about 0.1‰ for δ26/24Mg determinations (2SD). Thus, SI traceability for magnesium isotope amount ratios and delta values is demonstrated to be established.
Determination of absolute (SI‐traceable) isotope ratios: The use of Gravimetric Isotope Mixtures
(2023)
The presentation is brief overview on how to use gravimetric isotope mixtures to determine SI traceable isotope ratios. There is no mass spectrometer on earth that directly measures isotope ratios. Mass spectrometers will always measure signal intensity ratios instead. The actual problem is that the measured intensity ratios differ more or less from the isotope ratios. The difference can be up to more than 10 % in case of lithium while it‘s below 1 % in case of the heavier elements like lead or uranium. Consequently, the signal intensity ratios are expressed for example in V/V depending on the type of mass spectrometer you are using, while the isotope ratios are expressed in mol/mol. This phenomenon is called Instrumental Isotopic Fractionation (or short IIF) but the more common name is still mass bias (even though this name is not entirely correct). To convert the measured into the isotope ratio usually a simple multiplication with a so-called correction (or short K) factor is done. Therefore, the problem is to determine the K factor. In absence of isotope reference materials the golden route is via gravimetric isotope mixtures, which will be explained within the presentation.
The calibration of isotope ratio measurements is an ongoing challenge since instrumental isotope fractionation (IIF) has been detected in mass spectrometry (MS). There is a variety of approaches which either bypass IIF such as delta measurements or refer to reference materials (RMs) and thus shifting the problem of calibration to somebody else: the RM producer. For certifying isotope RMs with absolute isotope ratios only a few approaches are available, namely the isotope mixture approach, the double spike approach, the mass bias regression model and total evaporation in TIMS. All of them require either enriched isotopes, isotope RMs of another element or an RM for correcting residual error. As the enriched isotopes required for the isotope mixture and the double spike approach need to be fully characterized beforehand, all mentioned calibration approaches require a standard.
Here, a new and standard-free calibration approach for obtaining absolute isotope ratios of multi-isotopic elements has been developed. The underlying principle is that each MS suffers from IIF and thus yields a specific isotope fractionation line in a three-isotope diagram. When applying a second MS featuring a different ionization mechanism, we obtain a second isotope fractionation line with a different slope in the same three-isotope diagram. In both cases the absolute isotope ratios range somewhere on the isotope fractionation line. Consequentially, the intersect of both lines yield the absolute isotope ratios of the measured sample. This theory has been tested by measuring Cd and Pb isotope ratios of suitable isotope RMs with a TIMS and an ICP-MS, both equipped with multi-collector array. During the measurements the ionization conditions were changed such that different extent of the isotope fractionation has been achieved. With the resulting data set the theory described above could be verified. The obtained absolute isotope ratios were metrologically compatible with the certified isotope ratios. The remaining average bias of -5 ‰ can be reduced with further improvements. The calibration approach is universal and can be applied to any multi-isotopic element and it is not limited by the type of the mass spectrometer.
The presentations gives a short introduction into isotope analysis illustrated by examples from archaeology, food provenancing and metrology. In the main part current projects on the provenance determination of cement, the intercalibration of Mg isotope reference materials and the development of a new calibration approach for isotope measurements are presented.