Filtern
Dokumenttyp
Sprache
- Englisch (2)
Referierte Publikation
- nein (2)
Schlagworte
- Indentation (2) (entfernen)
A new calibration machine according to the
requirements given by ISO 14577-3 [1] was described in [2].
The results of the verification are reported in this paper. It is
shown that the calibration machine can be used as a national
standard for the materials parameter of the instrumented
indentation test.
The instrumented indentation test provides a versatile method to determine several materials parameter. In order to provide reference specimens, a calibration machine according to ISO 14577-3 is designed for the macro range. Using innovative technology, the new machine design may lead to a standard solution not only for calibration machines.