The requirement for a metrological infrastructure to ensure the interchangability of 'nonconventional’ energy gases within existing European infrastructure1 was the driving force behind the work undertaken in the three-year EMRP Characterisation of energy gases project EMRP ENG01 (June 2010 - May 2013).
As part of work package one of the project, Standards and methods were used to perform composition and impurity measurements on samples of real energy gases collected from around Europe. The aim of this study was to compare the results obtained from different labs, and thereby provide an evaluation of the labs’ capabilities and provide insight into the feasibility of different analytical methodologies for use with future measurements.
This project concerns the purity analysis of nitrogen as used in reference gas mixture preparation. This project was carried out without adding impurities to the gas used for this comparison, and is therefore more representative to evaluate the analysis of CO, CO2, CH4, O2, Ar and H2O impurities in high purity nitrogen. The analysis of the amount–of–substance fraction water was optional.
Two 50 litre high purity nitrogen cylinders were purchased from a well-qualified supplier of specialty gases. The listed components were expected to be present in the pure nitrogen at the target levels as a result of the purification of the nitrogen. From the start of this comparison it was clear that the comparison may not lead to reference values for the constituents analysed.
The results indicate that analyses of high purity gases are often limited by the limits of detection of analytical equipment used. The reports of the participating laboratories also indicate that there is no agreed method of determination of the uncertainty on a detection Limit value. The results provide useful information on the Performance of participants. For all analysed components there is reasonable agreement in results for LNE, VSL, Metas and NPL.
For BAM only the Argon result is in agreement.