The need for the traceable characterization of fluorescence instruments is emphasized from a chemists point of view, focusing on spectral fluorescence standards for the determination of the wavelength- and polarization-dependent relative spectral responsivity and relative spectral irradiance of fluorescence measuring systems, respectively. In a first step, major sources of error of fluorescence measurements and instrument calibration are revealed to underline the importance of this issue and to illustrate advantages and disadvantages of physical and chemical transfer standards for generation of spectral correction curves. Secondly, examples for sets of traceable chemical emission and excitation standards are shown that cover a broad spectral region and simple procedures for the determination of corrected emission spectra with acceptable uncertainties are presented. With proper consideration of the respective measurement principle and geometry, these dye-based characterization procedures can be not only applied to spectrofluorometers but also to other types of fluorescence measuring systems and even to Raman spectrometers.
The inter-instrument, inter-laboratory, and long-term comparability of fluorescence data requires the correction of the measured emission and excitation spectra for the wavelength- and polarization-dependent spectral irradiance of the excitation channel at the sample position and the spectral responsivity of the emission channel employing procedures that guarantee traceability to the respective primary standards. In this respect the traceability chain of fluorometry is discussed from a radiometrists point of view. This involves, in a first step, the realization of the spectral radiance scale, based on the blackbody radiator and electron storage ring, and the spectral responsivity scale, based on the cryogenic radiometer and their control via key comparisons of the national metrology institutes. In a second step, the characterization including state-of-the art uncertainties of the respective source and detector transfer standards such as tungsten strip lamps, integrating sphere radiators, and trap detectors used to disseminate these radiometric quantities to users of spectroscopic techniques is presented.
To improve the reliability of fluorescence data in the life and material sciences and to enable accreditation of fluorescence techniques, standardization concepts are required that guarantee and improve the comparability of fluorescence measurements. At the core of such concepts are simple and evaluated fluorescence standards for the consideration of instrument-specific spectral and intensity distortions of measured signals and for instrument performance validation (IPV). Similarly in need are fluorescence intensity standards for the quantification from measured intensities and for signal referencing, thereby accounting for excitation light-induced intensity fluctuations. These standards should be preferably certified, especially for use in regulated areas like medical diagnostics. This encouraged us to develop liquid and solid standards for different fluorescence parameters and techniques for use under routine measurement conditions in different formates. Special emphasis was dedicated to the determination and control of the spectral responsivity of detection systems, wavelength accuracy, homogeneity of illumination, and intensity referencing for e.g. spectrofluorometers, fluorescence sensors and confocal laser scanning fluorescence microscopes. Here, we will present design concepts and examples for mono- and multifunctional fluorescence standards that provide traceability to radiometric units and present a first step towards a toolbox of standards.