The introduction of HR-CS-MAS instrumentation, as it is shown in Fig. 1, provides a new way for the determination of non-metals including fluorine via molecular absorption. We present a sensitive and fluorine-specific method for total fluorine determination in fifteen consume care products in which perfluordecalin were declared (nail polishes, eye shadows and body creams). After pyrohydrolytic digestion of samples, fluorine content was determined by GF-HR-CS-MAS. Calcium was used as a molecule forming element and the molecular absorption of calcium monofluoride (CaF) at the wavelength 606.440 nm was used for its quantification. No matrix effect was observed. Analytical signal depends only on the fluorine concentration and therefore pure sodium fluoride NaF was used as calibration standard. The use of zirconium oxide nanoparticles as permanent modifier improves the limit of detection (from 0.2 to 0.1 ng) and characteristic mass (from 0.10 to 0.05 ng) from a previous reported method. These results were also compared with ISE method (standard DIN 51-084) and no significant difference was found
Variations in the isotope amount composition of some elements like lithium, boron, magnesium, calcium and copper have been used as proof of provenance of a sample and to describe geological processes. Routinely, isotope compositions are determinate by mass spectrometry; the working horse of the isotope analysis. However, mass spectrometric methods are expensive, time consuming and they require a high qualified analysist.
Here an alternative faster and low cost optical method for isotope ratio determination is proposed: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Stable isotope amount composition of X= Li, B, Mg, Ca, Cu and Sr have been determined by monitoring the absorption spectrum of their monohydride (XH) in graphite furnace HR-CS-MAS.
Analysis of stable isotopes has been used as proof of provenance of mineral and biological samples, to estimate a contamination source and to determine geological processes. This kind of analysis needs high accuracy and precision for reliable conclusions. Currently, stable isotope analysis is dominated by mass spectrometric techniques that are time consuming and expensive. Here we present a fast and low cost alternative for isotope analysis of boron and magnesium: high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS). Two stable isotope systems were evaluated separately: boron (10B:11B) and magnesium 24Mg:25Mg:26Mg). Their isotope amount ratios were estimated by monitoring their absorption spectra in-situ generated monohydrides. The molecular absorption spectrum of a XH molecule (X= B or Mg) with n isotopes would be a linear combination of n isotopologue spectra and the amount of each component (isotope) could be calculated by a multivariate regression (n= 2 and 3 for B and Mg respectively). For the Analysis of boron certified reference materials (CRM), the band 1→1 for the electronic transition X1Σ+ → A1Π was measured around wavelength 437.1 nm. Since boron has a Memory effect in graphite furnaces, a combination of 2 % (v/v) hydrogen gas in argon, 1 % trifluoromethane in argon, an acid solution of calcium chloride and mannitol as chemical modifiers were used during the BH vaporization at 2600 °C. Partial least square regression (PLS) for analysis of samples was applied. For this, a spectral library with different isotope ratios for PLS regression was created. Magnesium does not have memory effect. Therefore, only 2 % of hydrogen in argon as gas modifier during vaporization at 2500 °C was employed for analysis of magnesium CRM. Absorption spectra of MgH for the X2Σ→A2Π electronic transition (band 0→0) were recorded around wavelength 513.45 nm. A similar PLS procedure to the BH was applied. Results for B and Mg CRM are metrologically compatible with those reported by mass spectrometric methods. An accuracy of 0.08 ‰ for B and 0.1 ‰ Mg was obtained as the average deviation from the isotope CRM. Expanded uncertainties with a coverage factor of k = 2 range between 0.10 - 0.40 ‰.
Heavy alcohol induces several chronic diseases, including cirrhosis. Consequently, cirrhotic patients need liver transplantation. However, alcoholic patients are not suitable for this. How can be physicians sure of non-alcohol consumption of a potential organ receptor? Monitoring of biomarkers by mass spectrometric techniques is today the most suitable option. This subject was discussed at the Graduate School of Analytical Science Adlershof (SALSA) in Berlin, as a part of a lecture series in analytical science.
Variations in the isotope amount composition of some elements like lithium, boron, magnesium, calcium, copper and strontium have been used as proof of provenance of a sample and to describe geological processes. Routinely, isotope compositions are determinate by mass spectrometry; the working horse of the isotope analysis. However, mass spectrometric methods are expensive, time consuming and they require a high qualified analysist.
Here an alternative faster and low cost optical method for isotope ratio determination is proposed: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Stable isotope amount composition of X= Li, B, Mg, Ca, Cu and Sr have been determined by monitoring the absorption spectrum of their monohydride (XH) in graphite furnace HR-CS-MAS. For example, for the three Mg isotopes (24Mg, 25Mg and 26Mg) band (0→0) for the electronic transition X1Σ+ → A1Π was evaluated around wavelength 513.4 nm (Fig. 1). Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression were built. Results are metrological compatible with those reported by mass spectrometric methods.
Analysis of stable isotopes has been used as proof of provenance of mineral and biological samples, to estimate a contamination source and to determine geological processes. This kind of analysis needs high accuracy and precision for reliable conclusions. Currently, stable isotope analysis is dominated by mass spectrometric techniques that are time consuming and expensive. Here we present a fast and low cost alternative for isotope analysis of boron and magnesium: high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS). Two stable isotope systems were evaluated separately: boron (10B:11B) and magnesium (24Mg:25Mg:26Mg). Their isotope amount ratios were estimated by monitoring their absorption spectra in-situ generated monohydrides. The molecular absorption spectrum of a XH molecule (X= B or Mg) with n isotopes would be a linear combination of n isotopologue spectra and the amount of each component (isotope) could be calculated by a multivariate regression (n= 2 and 3 for B and Mg respectively). For the analysis of boron certified reference materials (CRM), the band 1→1 for the electronic transition X1Σ+ → A1Π was measured around wavelength 437.1 nm. Since boron has a memory effect in graphite furnaces, a combination of 2 % (v/v) hydrogen gas in argon, 1 % trifluoromethane in argon, an acid solution of calcium chloride and mannitol as chemical modifiers were used during the BH vaporization at 2600 °C. Partial least square regression (PLS) for analysis of samples was applied. For this, a spectral library with different isotope ratios for PLS regression was created. Magnesium does not have memory effect. Therefore, only 2 % of hydrogen in argon as gas modifier during vaporization at 2500 °C was employed for analysis of magnesium CRM. Absorption spectra of MgH for the X2Σ→A2Π electronic transition (band 0→0) were recorded around wavelength 513.45 nm. A similar PLS procedure to the BH was applied. Results for B and Mg CRM are metrologically compatible with those reported by mass spectrometric methods. An accuracy of 0.08 ‰ for B and 0.1 ‰ Mg was obtained as the average deviation from the isotope CRM. Expanded uncertainties with a coverage factor of k = 2 range between 0.10 - 0.40 ‰.
Boron isotope amount ratios n(10B)/n(11B) have been determined by monitoring the absorption spectrum of boron monohydride (BH) in a graphite furnace using high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Bands (0→0) and (1→1) for the electronic transition X1Σ+ → A1Π were evaluated around wavelengths 433.1 nm and 437.1 nm respectively. Clean and free of memory effect molecular spectra of BH were recorded. In order to eliminate the memory effect of boron, a combination of 2% (v/v) hydrogen gas in argon and 1% trifluoromethane in argon, an acid solution of calcium chloride and mannitol as chemical modifiers was used. Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression was built. Results obtained around the 433.1 nm and 437.1 nm spectral regions are metrologically compatible with those reported by mass spectrometric methods. Moreover, for the evaluated region of 437 nm, an accuracy of 0.15‰ is obtained as the average deviation from the isotope reference materials. Expanded uncertainties with a coverage factor of k = 2 range between 0.15 and 0.44‰. This accuracy and precision are compatible with those obtained by mass spectrometry for boron isotope ratio measurements.
Currently, two main high speed processes can be investigated: electromagnetic absorption and electromagnetic scattering. At the summer university of the School of Analytical Sciences Adlershof (SALSA-HU), a series of 5 articles were presented to ilustrate the potential of ultrafast spectroscopy, from IR band assignment to monitoring time resolved structural changes. First, by studying light absorption is possible to do assignment and interpretation of IR bands in small molecules and to study the influence of ligands or labeled molecules in protein structures. Second, by studying light scattering is possible to monitoring photoinduced structural changes in solid materials.
This presentation contains a study of photoinduced structural changes monitored by light scattering in condensed matter.
High-Resolution Continuum Source Molecular Absorption Spectrometry (HR-CS-MAS) with graphite furnace atomizers has been successfully used for the analytical determination of non-metals. The use of permanent modifiers (or coatings) for molecule formation is necessary, but their mechanisms of action remain unclear. In order to understand and improve the current analytical methods, we present a comprehensive mechanistic study of molecule formation in graphite furnaces: a key step into the recovery of analytical signals. Therefore, a well-known system for fluorine determination was studied: the molecule formation of CaF, with Zr as permanent modifier. Through a kinetic approach, an Arrhenius plot showed a pseudo first order reaction respect to F-. By spectroscopic methods (XPS, XAS, EDX) it was possible to elucidate an intermediate state. Here a mechanism is proposed, where ZrO2 works as heterogeneous catalyst: after a pyrolytic step, it is activated to the intermediate state of ZrO(OCaF) and at higher temperatures, CaF(g) is released from the zirconium-coated graphite surface. With this analytical information, improved surface modifications and coatings for graphite furnaces can be designed and some of them, for example with Zr nanoparticle and Zr xerogels are here presented.
Analytical applications of high resolution continuum source molecular absorption spectrometry
(2017)
High-resolution continuum source molecular absorption spectrometry (HR-CS-MAS) has been apply for the determination of non-metals and isotope analysis, extending so the application range of atomic absorption spectrometry (AAS). This seminar is divides in two main parts. First, here is presented a comprehensive mechanistic study of molecule formation in graphite furnaces, a key step into the recovery of analytical signals in AAS and MAS. Therefore, a well-known system for fluorine determination was studied: the molecule formation of CaF, with Zr as permanent modifier. Through a kinetic approach, an Arrhenius behaved pseudo first order reaction respect to F- was observed and by spectroscopic methods (XPS, XAS, EDX) an intermediate state was possible to be elucidated. Here it is proposed a mechanism, where zirconium works as heterogeneous catalyst: after a pyrolytic step, it is activated the intermediate ZrO(OCaF) and at higher temperatures, CaF(g) is released from the zirconium-coated graphite surface. Second, we have developed analytical methods using HR-CS-MAS as detector for quantification of fluorine in consume-care products with declared per-fluorinated ingredients. Ad, the high resolution of the instrumentation allows identify isotopic shifts in some observed molecular spectra. Consequently, the molecular spectra of enriched isotopes of B were investigated and so the potential of HR-CS-MAS for the determination of isotopic ratios is established.