Silver nanoparticles (Ag NPs) are widely used in consumer products due to their excellent antibacterial properties. Their broad application has led to a variety of recent regulation on their use and labelling. Thus, a highly specific analytical method for their characterization and quantification is needed.
Due to their large separation range, field-flow fractionation (FFF) techniques are repeatedly applied for the analysis of NP. Limitations of FFF include quantification, sample loss and insufficient recovery rates. Another challenge can be non-ideal elution behavior of particles in complex and unknown matrices.
The possible sources for sample losses of Ag NP have been studied using an asymmetric flow FFF (AF4) in combination with inductively coupled plasma mass spectrometry (ICP-MS). The influence of different parameters, for example the sample concentration, on the recovery rates and sample loss has been investigated. Using laser ablation ICP-MS, the Ag deposition on the membrane was located and quantified. Our results identified ionic silver as the main sources of sample loss. These results can be useful for further method improvement.
However, when a Ag NP sample containing an unknown complex matrix is analyzed, FFF method optimization is challenging as the sample might show a shift in the retention times and lower recovery rates. In this case, ICP-MS experiment in the single particle mode (sp-ICP-MS) can be a useful addition to the FFF measurement. Here, upon assumption of spherical particles, the geometric diameters can be calculated. This fast and easy approach can be helpful in order to interpret the FFF fractograms and advice the FFF method optimization process.
Due to its simplicity, speed and ability to obtain a particle number size distribution, single particle ICP-MS (spICP-MS) has emerged as an important tool for the analysis of nanoparticles (NPs). However, when NPs are suspended in a complex, unknown solution, matrix effects can occur affecting the instrument’s sensitivity. As a result, an over- or underestimation of the particle size is possible.
In this work, a proof-of-concept study of the combination of isotopic dilution analysis (IDA) and spICP-MS compensating for possible matrix effects is presented. As an example, an isotopically enriched 109Ag standard solution was added to silver NPs (Ag NP) suspensions. Different NP suspensions with mean particle diameters between 30 and 80 nm were chosen. The mixtures were analyzed using a quadrupole ICP-MS instrument. Both Ag isotopes (107Ag and 109Ag) were monitored during one experiment. The result show a good agreement with the diameters obtained using conventional spICP-MS.
In a second step, the Ag NPs were suspended in a simulated seawater matrix. Using conventional spICP-MS, a great reduction in the signal intensities and consequently in the particle sizes, was monitored. The application of the IDA-spICP-MS approach on these samples was able to obtain similar diameters compared to the samples without matrix.