There is a high international priority attached to activities which reduce NOx in the atmosphere. The current level of permitted emissions is typically between 50 µmol/mol and 100 µmol/mol, but lower values are expected in the future. Currently, ambient air quality monitoring regulations also require the measurement of NOx mole fractions as low as 0.2 µmol/mol. The production of accurate standards at these levels of mole fractions requires either dilution of a stable higher concentration gas standard or production by a dynamic technique, for example one based on permeation tubes.
The CCQM-K74 key comparison was designed to evaluate the level of comparability of National Metrology Institutes' measurement capabilities and standards for nitrogen dioxide (NO2) at a nominal mole fraction of 10 µmol/mol.
The measurements of this key comparison took place from June 2009 to May 2010.
Seventeen laboratories took part in this comparison coordinated by the BIPM and VSL. The key comparison reference value was based on BIPM measurement results, and the standard measurement uncertainty of the reference value was 0.042 µmol/mol.
This key comparison demonstrated that the results of the majority of the participants agreed within limits of ±3% relative to the reference value. The results of only one laboratory lay significantly outside these limits. Likewise this comparison made clear that a full interpretation of the results of the comparison needed to take into account the presence of nitric acid (in the range 100 nmol/mol to 350 nmol/mol) in the cylinders circulated as part of the comparison, as well as the possible presence of nitric acid in the primary standards used by participating laboratories.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
DatesIssue 1A (Technical Supplement 2012)
This pilot study compares the performance of participants in analyzing gas mixtures of nitrogen dioxide in nitrogen by comparison with in-house gravimetric standards using Fourier transformed infrared spectroscopy (FT-IR). In this study the same gas mixtures were used as in the key comparison CCQM-K74, which was designed to evaluate the level of comparability of National Metrology Institutes' measurement capabilities for nitrogen dioxide (NO2) at a nominal mole fraction of 10 µmol/mol. In the comparison CCQM-K74 most of the participants used chemiluminescence, with a small number using UV absorption or FT-IR spectroscopy, and thus it is of interest to improve understanding of the comparative performance of these techniques because they do not exhibit any cross-sensitivity to nitric acid (HNO3), which was known to be present in the mixtures used for the comparison.
The results of this pilot study indicate good consistency and a level of agreement similar to that reported in the comparison CCQM-K74, demonstrating that FT-IR can be operated as a comparison method when calibrated with appropriate gas standards and can achieve similar measurement uncertainties to chemiluminescence and UV absorption techniques.
An additional pilot study, CCQM-P110-B2, was conducted on the same gas mixtures in parallel with this pilot study. The second study addressed FT-IR spectroscopy when used to measure the gas mixtures with respect to reference spectra. The results of this second study will be reported elsewhere.
Main text. To reach the main text of this paper, click on Final Report.
The final report has been peer-reviewed and approved for publication by the CCQM-GAWG.
DatesIssue 1A (Technical Supplement 2012)
Under the auspices of the Organic Analysis Working Group (OAWG) of the Comité Consultatif pour la Quantité de Matière (CCQM) a key comparison, CCQM K55.b, was coordinated by the Bureau International des Poids et Mesures (BIPM) in 2010/2011. Nineteen national measurement institutes and the BIPM participated. Participants were required to assign the mass fraction of aldrin present as the main component in the comparison sample for CCQM-K55.b which consisted of technical grade aldrin obtained from the National Measurement Institute Australia that had been subject to serial recrystallization and drying prior to sub-division into the units supplied for the comparison.
Aldrin was selected to be representative of the performance of a laboratory's measurement capability for the purity assignment of organic compounds of medium structural complexity [molar mass range 300 Da to 500 Da] and low polarity (pKOW < -2) for which related structure impurities can be quantified by capillary gas phase chromatography (GC).
The key comparison reference value (KCRV) for the aldrin content of the material was 950.8 mg/g with a combined standard uncertainty of 0.85 mg/g. The KCRV was assigned by combination of KCRVs assigned by consensus from participant results for each orthogonal impurity class. The relative expanded uncertainties reported by laboratories having results consistent with the KCRV ranged from 0.3% to 0.6% using a mass balance approach and 0.5% to 1% using a qNMR method.
The major analytical challenge posed by the material proved to be the detection and quantification of a significant amount of oligomeric organic material within the sample and most participants relying on a mass balance approach displayed a positive bias relative to the KCRV (overestimation of aldrin content) in excess of 10 mg/g due to not having adequate procedures in place to detect and quantify the non-volatile content–specifically the non-volatile organics content–of the comparison sample.
There was in general excellent agreement between participants in the identification and the quantification of the total and individual related structure impurities, water content and the residual solvent content of the sample.
The comparison demonstrated the utility of 1H NMR as an independent method for quantitative analysis of high purity compounds. In discussion of the participant results it was noted that while several had access to qNMR estimates for the aldrin content that were inconsistent with their mass balance determination they decided to accept the mass balance result and assumed a hidden bias in their NMR data. By contrast, laboratories that placed greater confidence in their qNMR result were able to resolve the discrepancy through additional studies that provided evidence of the presence of non-volatile organic impurity at the requisite level to bring their mass balance and qNMR estimates into agreement.
Under the auspices of the Organic Analysis Working Group (OAWG) of the Comité Consultatif pour la Quantité de Matière (CCQM) a key comparison, CCQM K55.a, was coordinated by the Bureau International des Poids et Mesures (BIPM) in 2009/2010. Eleven national measurement institutes and the BIPM participated. Participants were required to assign the mass fraction of estradiol present as the main component in the comparison sample (CCQM-K55.a) which consisted of a bulk estradiol hemihydrate material obtained from a commercial supplier that had been extensively but not exhaustively dried prior to sub-division into the units supplied for the comparison.
Estradiol was selected to be representative of the performance of a laboratory's measurement capability for the purity assignment of organic compounds of medium structural complexity [molar mass range 300500 Da] and low polarity (pKOW < -2) for which related structure impurities can be quantified by capillary gas phase chromatography (GC) or by high performance liquid chromatography (LC).
The majority of participants used a mass balance approach to determine the estradiol content. The key comparison reference value (KCRV) for estradiol in CCQM-K55.a was assigned by combination of KCRVs assigned by consensus from participant results for each orthogonal impurity class. This allowed participants to demonstrate the efficacy (or otherwise) of their implementation of the mass balance approach and to demonstrate that their assigned value for the main component agreed with the KCRV through use of internally consistent contributing methods.
The KCRV for the estradiol content of the material was 984.3 mg/g with a combined standard uncertainty of 0.42 mg/g. The individual participant results showed that a relative expanded uncertainty for the purity assignment of 0.2% is a reasonable estimate of the best achievable result by an individual laboratory for a material of this complexity available in this amount at this level of purity. The relative expanded uncertainties reported by laboratories having results consistent with the KCRV ranged from 0.2% to 0.8%.
The review of results that were biased from the KCRV showed that two major analytical challenges are posed by the material: the measurement of its water content and controlling for related substance artefact formation during the analysis process. The results displaying a positive bias relative to the KCRV (overestimation of estradiol content) were due to underestimation of the water content of the material, while those with a negative bias (underestimation of estradiol) overestimated the total related substance impurities through a failure to detect and control for artefact formation arising from in situ oxidative dimerization of estradiol in neutral solution prior to analysis. There was however good agreement between all participants in the identification and the quantification of the individual related structure impurities actually present in the sample.
The comparison also demonstrated the utility of high-field 1H NMR for both quantitative and qualitative analysis of high purity compounds. It is noted that all the participants who used qNMR as a major or contributing technique and included it as part of, combined it or confirmed it with a conventional 'mass balance' data estimate, obtained results consistent with the KCRV.