Carbamazepine (CBZ), caffeine and cetirizine were monitored by enzyme-linked immunosorbent assays (ELISAs) in surface and wastewaters from Berlin, Germany. This fast and cost-efficient method enabled to assess the spatial and temporal variation of these anthropogenic markers in a high-throughput screening. CBZ and cetirizine were detected by the same antibody, which selectively discriminates between both compounds depending on the pH value used in the incubation step. To our best knowledge, this is the first dual-analyte immunoassay working with a single antibody.
The frequent sampling with 487 samples being processed allowed for the repeated detection of unusually high concentrations of CBZ and caffeine. ELISA results correlate well with the ones obtained by liquid chromatography tandem mass spectrometry (LC-MS/MS). Caffeine concentrations found in surface waters were elevated by combined sewer overflows after stormwater events. During the hay fever season, the concentrations of the antihistamine drug cetirizine increased in both surface and wastewaters.
Caffeine was almost completely removed during wastewater treatment, while CBZ and cetirizine were found to be more persistent. The maximum concentrations of caffeine, CBZ and cetirizine found in influent wastewater by LCMS/MS were 470, 5.0 and 0.49 µg L-1, while in effluent wastewater the concentrations were 0.22, 4.5 and 0.51 µg L-1, respectively. For surface waters, concentrations up to 3.3, 4.5 and 0.72 µg L-1 were found, respectively.
Antibodies are a promising tool for the fast and selective trace detection of explosives. Unfortunately, the production of high-quality antibodies is not trivial and often expensive. Therefore, excellent antibodies are a rare and limiting resource in fields such as biosensing, environmental analysis, diagnostics, cancer therapy, and proteomics. Here, we report the synthesis, bioconjugation, and application of the structurally optimized hapten 6-(2,4,6-trinitro)-phenylhexanoic acid to improve the selectivity and sensitivity of antibodies for the detection of one of the most important explosives, trinitrotoluene. With a conjugate of bovine serum albumin and a highly purified N-hydroxy-succinimide (NHS)activated hapten, two rabbits were immunized to obtain polyclonal antibodies. The immunization process was monitored by enzyme-linked immunosorbent assay to gain information about the progress of antibody titer and affinity. Finally, the polyclonal antibodies reached an affinity constant of (5.1 ± 0.6) × 109l/mol (rabbit R1) and (2.3 ± 0.2) × 109l/mol (rabbit R2). The respective assays show a minimum test midpoint (IC50 value) of 0.1 ± 0.01 µg/l (R1) and 0.2 ± 0.02 µg/l (R2) and a working range of 0.005 to 150 µg/l (R1) and 0.007 to 200 µg/l (R2), which corresponds to more than four orders of magnitude for both. This is quite remarkable for a competitive immunoassay, which is often believed to have a narrow dynamic range. The limit of detection was calculated to 0.6 ng/l (R1) and 1.5 ng/l (R2), which is up to 100 times improvement in relation to the assay of Zeck et al. (1999) on the basis of a monoclonal antibody. The excellent selectivity of the polyclonal antibodies was comprehensively examined by determining the cross-reactivity to common explosives and other nitroaromatics including nitro musk components. The widely held belief that polyclonal antibodies generally display higher cross-reactivities than monoclonals could be disproved.