The meaning of calibration in general and in the field of isotopic measurements in particular is described, stressing the fundamental difference between calibration as an aim and tools to achieve calibration. The role of proper uncertainty budgeting as a prerequisite for establishing a calibrated measurement is explained. It leads to the recommendation that the quality of the uncertainty statement should be heavily weighed when ranking or judging calibrated isotopic measurement results evaluated by the Commission on Atomic Weight and Isotopic Abundances.
Two synthetic isotope mixtures for the calibration of sulfur isotope amount ratio measurements were gravimetrically prepared from high purity Ag2S materials enriched in 32S, 33S, and 34S. The mixtures were made so as to closely resemble the (natural) isotopic composition of the materials to be calibrated. This allowed a totally independent evaluation, on the same samples, of the relative combined uncertainty of: (a) the procedure to perform direct measurements of the amount of substance ratios of gas isotopes in the redetermination of the Avogadro constant and (b) the gravimetric preparation procedure. The result of both procedures, mass spectrometry and gravimetry, agree to a relative uncertainty of 3 × 10-4 for sulfur amount ratio measurements of the major abundant isotopes. Thus it seems that a direct measurement of isotopic gas mixtures (e.g. of natural isotopic composition) is now possible for sulfurand probably also for other gaseous isotopeswithout necessarily having to rely on calibration by means of values provided by measurements of gravimetrically prepared isotope mixtures. However, synthetic mixtures may be needed for validation and verification purposes, in particular for quality assurance.
Isotope amount ratios of krypton were measured on subsamples from one large batch of high purity krypton separated from the atmosphere. Synthetic mixtures of enriched krypton isotopes were used to calibrate the measurements with small uncertainties. The result is a primary isotopic gas standard (PIGS) IRMM-2030 with certified values for isotope ratios, isotopic composition, and molar mass of krypton with small combined uncertainties uc, evaluated according to the ISO/BIPM Guide (GUM). It is commercially available from IRMM-Geel or from MESSER (Duisburg, D). The certified krypton isotope amount ratios in the PIGS IRMM-2030 are as follows: n(78Kr)/n(84Kr)=0.006 232 5(55), n(80Kr)/n(84Kr)=0.040 107(17), n(82Kr)/n(84Kr)=0.203 43(12), n(83Kr)/n(84Kr)=0.201 79(11), and n(86Kr)/n(84Kr)=0.303 205(59) with expanded uncertainty U=kuc and coverage factor k=2. The molar mass of Kr in this sample is M(Kr)=83.798 02(16) g/mol. These values are in good agreement with published measurements of atmospheric krypton but have smaller combined uncertainties and are calibrated by means of synthetic isotope mixtures. The values of the PIGS are traceable to the SI. Measurements of isotope amount ratios of krypton in other samples can be linked to SI using this PIGS.
Calibrated values have been obtained for sulfur isotope abundance ratios of sulfur isotope reference materials distributed by the IAEA (Vienna). For the calibration of the measurements, a set of synthetic isotope mixtures were prepared gravimetrically from high purity Ag2S materials enriched in32S, 33S, and 34S. All materials were converted into SF6 gas and subsequently, their sulfur isotope ratios were measured on the SF5+ species using a special gas source mass spectrometer equipped with a molecular flow inlet system (IRMMs Avogadro II amount comparator). Values for the 32S/34S abundance ratios are 22.650 4(20), 22.142 4(20), and 23.393 3(17) for IAEA-S-1, IAEA-S-2, and IAEA-S-3, respectively. The calculated 32S/34S abundance ratio for V-CDT is 22.643 6(20), which is very close to the calibrated ratio obtained by Ding et al. (1999). In this way, the zero point of the VCDT scale is anchored firmly to the international system of units SI. The 32S/33S abundance ratios are 126.942(47), 125.473(55), 129.072(32), and 126.948(47) for IAEA-S-1, IAEA-S-2, IAEA-S-3, and V-CDT, respectively. In this way, the linearity of the V-CDT scale is improved over this range. The values of the sulfur molar mass for IAEA-S-1 and V-CDT were calculated to be 32.063 877(56) and 32.063 911(56), respectively, the values with the smallest combined uncertainty ever reported for the sulfur molar masses (atomic weights).