Special new techniques of atomic spectrometric methods for trace analysis of high purity metals
(2003)
Three methods for direct solid sampling of bulk material namely IR laser ablation, glow discharge and spark OES, were compared with respect to analytical figures of merit obtained for elemental analysis with atomic spectrometry. Matrices investigated were copper, pressed doped copper powder, and magnesium alloys. For the vast majority of analytes, statistical equivalence regarding precision (usually ≤ 5%) and the performance of the calibrations between the compared methods was demonstrated.
Solution-doped metal powder pellets as well as aspirated liquids were used as calibration samples to analyze pure copper and zinc certified reference materials (CRMs) by femtosecond laser ablation ICP-MS. It was demonstrated that calibration by copper pellets resulted in relative deviations up to 20%, whereas fs-LA-ICP-MS among copper-based CRMs led to inaccuracies in the same range unless nominal mass fractions were chosen to be <3 mg/kg. Calibration by zinc pellets generally provided better accuracy. Depending on the analyte considered, deviations below 10% were obtained even for mass fractions close to the limit of quantification. Our data, therefore, indicate solution-doped metal powder pellets to be suitable as calibration samples for fs-LA-ICP-MS of metals. Furthermore, the utilization of liquid standards for calibration was found to result in stronger deviations of up to 50% for both copper and zinc samples which, in addition, turned out to be dependent on the plasma conditions.
In this work, different calibration strategies for the determination of trace elements in pure copper metal by nanosecond laser ablation ICP-MS were investigated. In addition to certified reference materials (CRMs), pellets of doped copper powder were used for calibration. The micro homogeneity of the CRMs as well as the solution-doped pellets was sufficient to use them as calibration samples in combination with a laser spot size of 200 µm. In contrast, pellets doped with analytes in solid form showed a significant heterogeneity. For most of the investigated analytes and copper CRMs the measured mass fractions were within ± 20% of their certified values when other copper CRMs were used as calibration samples. When solution-doped powder pellets were used as calibration samples a systematic trend towards mass fractions below the certified values was observed for nearly all elements determined in the analysed CRMs. Thermal fractionation effects during the ablation of the solution-doped pellets were suspected as the extent of the fractionation depends on the irradiance, whereas fractionation is reduced at higher irradiance.
Two types of copper samples, compact certified copper reference materials and calibration samples prepared from liquid doped, pressed copper powders, were studied in terms of accuracy of obtained calibration functions originating from infrared spark ablation. Additionally, corresponding particle size distributions of the aerosols from infrared spark ablation were recorded. It is shown that the differences in quantification results, originating from the two sets of calibration functions, could not mainly be ascribed to different particle size distributions of the two copper sample types. Possible other causes, as different ablation rates, parts of melting and differences of the chemical constitutions of the two sample types were explored.