In view of its intended use as a sample for proficiency testing or as a reference material the stability of the extractable trace element contents of a soil from an irrigation field was tested using the extraction with 1 mol/L ammonium nitrate solution according to DIN 19730. Therefore, changes of the extractability of sterilized and non sterilized soil samples stored at different temperatures were evaluated over a period of 18 months. Sets of bottles were kept at -20 °C, +4 °C, about +20 °C and +40 °C, respectively. The NH4NO3 extractable contents of Cd, Cr, Cu, Ni, Pb and Zn were determined immediately after bottling and then after 3, 6, 12 and 18 months with ICP-AES or ETAAS. Appropriate storage conditions are of utmost importance to prevent deterioration of soil samples prepared for the determination of NH4NO3 extractable trace element contents. Temperatures above +20 °C must be avoided. The observed changes in the extractability of the metals (especially for Cr and Cu) most likely could be related to thermal degradation of the organic matter of the soil. There is no need to sterilize dry soil samples, because microbiological activity in soils with a low moisture content appears to be negligible with regard to trace element mobilization.
Entwicklung und Herstellung von Boden-Referenzmaterialien für die anorganische Umweltanalytik
(2000)
Within the scope of a project of the 'European Association of National Metrology Institutes' (EURAMET), a European metrological dissemination System (network) providing traceable reference values assigned to matrix materials for validation purposes is described and put to the test. It enables testing laboratories (TL) to obtain comparable results for measurements under the 'EU Water Framework Directive 2000/60/EC' (WFD) and thus, to comply with a core requirement of this very directive. The dissemination system is characterized by the fact that it is available to all laboratories throughout Europe which intend to perform measurements in the context of the WFD and that it can ensure sustainable metrological traceability to the International System of Units (SI) as a reference point for the measurement results. This Dissemination system is set up in a hierarchical manner and links up the level of the national metrology institutes (NMI) with that of the TLs via an intermediate level of calibration laboratories (CL) by comparison measurements. The CLs are expert laboratories with respect to the measurement of the analytes considered here (within the project, the CLs are called potential calibration laboratories (PCL)) and are additionally involved in the organization of comparison measurements within the scope of regional quality assurance (QA) systems.
Three comparison measurements have been performed to support the approach. A total of about 130 laboratories participated in this exercise with the focus on the measurement of the priority substances Pb, Cd, Hg, and Ni defined in the WFD. The elemental concentrations in the water samples roughly corresponded to one of the established environmental quality standards (EQS), the annual average concentration (AA-EQS), which is defined in the daughter Directive 2008/105/EC of the WFD. It turned out that a significant number of TLs still need to improve their measurement methods in order to be able to fulfill the Minimum requirements of the WFD, in particular, with regard to the elements Cd and Hg probably due to their low EQS values.
Furthermore, it became obvious that the hierarchical Dissemination system suggested here actually corresponds to the measuring capabilities of the three participating Groups (NMIs, PCLs, and TLs).
With the implementation of glass casting in France in the second half of the 17th century, larger mirrors could be produced. So-called tin-mercury mirrors were made by coating glass with tin amalgam. Today, many historical mirrors are partly damaged, the image quality is compromised, and the material integrity of the mirror is threatened. The transformation from tin into an oxide starts at the surface of the amalgam and proceeds down to the glass surface. The mercury dissolves as a liquid or gaseous phase. The amalgam layer is destroyed, so that conservation is necessary. Conservators must follow strict safety precautions while handling amalgam mirrors because of possible mercury emissions and corrosion of the amalgam, during which elemental mercury accumulates near the mirror frame in droplets and is emitted into the air.
In a research project, the quantity of the mercury emissions from historical mirrors was examined and a technique to decrease these emissions was developed.
Validation of a European standard for the determination of hexavalent chromium in solid material
(2007)
A European standard for the determination of Cr(VI) in solid material has been elaborated in the framework of an international co-operation and finally validated in the course of an interlaboratory comparison. The procedure is based on the alkaline digestion prescribed by EPA method 3060A followed by ion chromatography and determines an operationally defined content of Cr(VI), including water-soluble and insoluble chromates. A preliminary robustness study was carried out in order to compare different extraction methodologies and to study the equivalency of different analytical methods for the determination of Cr(VI) in alkaline extracts of soil and waste materials. During an interlaboratory validation trial with 19 European laboratories a set of 4 samples (2 soil and 2 waste samples) was analysed to determine performance characteristics for different combinations of digestion and detection methods. With the procedures prescribed by the new European standard (EN 15192) acceptable results were obtained for both soil samples and one of the waste samples (sludge). However, for the second waste sample (fly ash) a large deviation in analytical results was observed. This indicates that particularly for waste materials a possible occurrence of strong matrix effects has to be considered and supplementary quality control data are needed in order to assess the validity of analytical results. The accuracy of the determination of Cr(VI) in solid matrices remains a challenging field in terms of maximum extraction efficiency and minimum species interconversion.
The resource-saving utilisation of recovered
waste wood is a matter of growing concern. In several
European countries, this utilisation is governed by regulations
and is dependent on the contents of certain trace
elements and organic compounds. Thus, for decisionmaking
with respect to waste wood management (recycling
or combustion), reliable analytical data are needed and, due
to their great economic and environmental impact, must be
assured by appropriate quality control. To support the
improvement in quality assurance in waste wood analysis,
for the first time, a wood reference material was certified
for its mass fractions of arsenic, cadmium, chromium,
copper, lead, mercury, and pentachlorophenol (PCP).
These analytes were selected because they represent typical
constituents of wood preservatives most widely used in the
past. Material preparation and testing of homogeneity and
stability were carried out by BAM Federal Institute for
Materials Research and Testing. The certification measurements
were performed involving selected laboratories
with documented expertise in the field of waste wood
analysis. The certified values and their corresponding
uncertainties were assigned in full compliance with the
requirements of ISO Guide 35. The certified mass fractions
and their expanded uncertainties (k/i> = 2) are as follows:
(3.1 ± 0.5) mg/kg for As, (3.02 ± 0.24) mg/kg for Cd,
(36.4 ± 2.6) mg/kg for Cr, (22.9 ± 1.7) mg/kg for Cu,
(0.60 ± 0.14) mg/kg for Hg, (39 ± 4) mg/kg for Pb, and
(7.9 ± 0.6) mg/kg for PCP. The certified material is
available as European Reference Material ERM®-CD100.
Mit der Erfindung des Glas-Guß-Walzverfahrens eröffnete sich die Möglichkeit, größere Spiegelflächen als zuvor herzustellen. Diese sogenannten Quecksilberspiegel sind mit Zinnamalgam beschichtet. Wenn Restauratoren mit Zinnamalgamspiegeln arbeiten, müssen sie strenge Sicherheitsvorkehrungen einhalten. Häufig finden sich im Spiegelrahmen Quecksilberkugeln, die beim Öffnen herausfallen und damit gesundheitsschädliches Quecksilber freisetzen können. In einem Forschungsprojekt wurden die Quecksilberemissionen aus Zinnamalgamspiegeln gemessen und Möglichkeiten entwickelt, diese zu verringern.