The VSSA approach has the important advantage over classifying, imaging and counting techniques that it does not involve dispersion protocols. Further, the BET technique as the basis for VSSA determination it is in widespread use, generates low costs and is specified for many commercial materials. Finally, the same equipment allows for a deeper analysis by full isotherm evaluation.
The present deliverable assesses all NanoDefine powders, supplemented by further real-world materials (in total 26 powders), and quantitatively compares the relationship between the median size (by Electron Microscopy – considered as benchmark for the EC nanomaterial definition) vs. the size derived from VSSA. The VSSA method mitigates the challenges of EM to assess the thickness of platelets, but worked as well on fibbers and particles of irregular shapes. A screening strategy is proposed. If applied to the further data from real-world materials as validation set, this screening does achieve a correct classification, leaving only borderline materials for tier 2 assessment.
Three petrol reference materials were certified for their total sulfur content in support of Directive 2003/17/EC of the European Parliament and of the European Council, which stipulates that petrol with a maximum S content of 10 mg/kg must be available in all member states by 2009. Commercially available petrol was obtained and filled into borosilicate ampoules without further treatment. Homogeneity of the materials was tested and no heterogeneity was detected for two of the materials, whereas minor heterogeneity was observed for the third material. Stability of the materials was tested for 8-12 months at 60 °C and no degradation was observed. Characterisation was based on isotope-dilution mass spectrometry (IDMS) applied as primary method of measurement by three European metrology institutes and certified values were assigned using all results. The certified uncertainties include contributions of (potential) heterogeneity, potential degradation as well as characterisation. The final assigned values are:
ERM-EF211: 48.8 ± 1.7 mg/kg
ERM-EF212: 20.2 ± 1.1 mg/kg
ERM-EF213: 9.1 ± 0.8 mg/kg
The materials are available from IRMM (ERM-EF211), LGC (ERM-EF212) and BAM (ERMEF213)