We present a new high resolution X-ray imager based on a pnCCD detector and a polycapillary optics. The properties of the pnCCD like high quantum efficiency, high energy resolution and radiation hardness are maintained, while color corrected polycapillary lenses are used to direct the fluorescence photons from every spot on a sample to a corresponding pixel on the detector. The camera is sensitive to photons from 3 to 40 keV with still 30% quantum efficiency at 20 keV. The pnCCD is operated in split frame mode allowing a high frame rate of 400 Hz with an energy resolution of 152 eV for Mn Kα (5.9 keV) at 450 kcps. In single-photon counting mode (SPC), the time, energy and position of every fluorescence photon is recorded for every frame. A dedicated software enables the visualization of the elements distribution in real time without the need of post-processing the data. A description of the key components including detector, X-ray optics and camera is given. First experiments show the capability of the camera to perform fast full-field X-Ray Fluorescence (FF-XRF) for element analysis. The imaging performance with a magnifying optics (3×) has also been successfully tested.
Archaeological bone, ivory and antler, as well as objects manufactured from them, are largely studied to extract as much information as possible from these materials. Among others, one key question in archaeology is the exact identification of the material. Even if the identification of different kinds of osseous material seems to be trivial in many cases, it can be a difficult issue when small, heavily carved and more or less altered ancient objects are concerned. This study was focused on the determination of parameters allowing the distinction of archaeological ivory, bone and antler in order to identify the raw material used for the manufacture of prehistoric objects. The high performance of synchrotron radiation (SR) and laboratory-based micro X-ray computed tomography (microCT), providing highly resolved three-dimensional information on the micromorphology, permitted the establishment of such distinctive features of modern references of ivory, antler, land mammal and whale bones: ivory shows characteristic tubular pores with a diameter of about 1 to 2 µm, bone and antler show typical osteon structures. In our measured references, antler shows on average larger and more elongated shaped pores of the osteons compared to terrestrial mammal bone. This feature however depends very much on the original localization of the studied sample within the antler. Whale bones can be distinguished from the other osseous materials by a cancellous, osteoporotic-like structure with irregularly distributed rounded porosities with diameters reaching up to 500 µm. These characteristics have also been tested on determined Palaeolithic fragments, as diagenetic changes during burial have to be considered and may lead to the modification of the parameters established on the basis of modern bone references. In general, the chemical composition of bone objects can change drastically over time while micromorphological features, as evidenced by microCT, seem to be less susceptible to such alterations. In addition, microCT enables the comparison of inner and possibly less altered parts of the objects, and can be considered as completely non-destructive for small mineralised prehistoric objects. In this study, specific morphological features allowing the distinction of ivory and of whale bone from other bone and antler material were determined, even for altered materials dating back to Palaeolithic periods. Thus, we provide, in addition to archaeozoological, chemical and isotopic markers, a new non-destructive tool to identify some raw materials used for the fabrication of osseous objects ranging from recent to prehistoric periods.
With the increase in the awareness of the public in the environmental impact of oil shale utilization, it is of interest to reveal the mobility of potentially toxic trace elements in spent oil shale. Therefore, the Cr and As oxidation state in a representative Jordanian oil shale sample from the El-Lajjoun area were investigated upon different lab-scale furnace treatments. The anaerobic pyrolysis was performed in a retort flushed by nitrogen gas at temperatures in between 600 and 800 °C (pyrolytic oil shale, POS). The aerobic combustion was simply performed in porcelain cups heated in a muffle furnace for 4 h at temperatures in between 700 and 1000 °C (burned oil shale, BOS). The high loss-on-ignition in the BOS samples of up to 370 g kg–1 results from both calcium carbonate and organic carbon degradation. The LOI leads to enrichment in the Cr concentrations from 480 mg kg–1 in the original oil shale up to 675 mg kg–1 in the ≥850 °C BOS samples. Arsenic concentrations were not much elevated beyond that in the average shale standard (13 mg kg–1). Synchrotron-based X-ray absorption near-edge structure (XANES) analysis revealed that within the original oil shale the oxidation states of Cr and As were lower than after its aerobic combustion. Cr(VI) increased from 0% in the untreated or pyrolyzed oil shale up to 60% in the BOS ash combusted at 850 °C, while As(V) increased from 64% in the original oil shale up to 100% in the BOS ash at 700 °C. No Cr was released from original oil shale and POS products by the European compliance leaching test CEN/TC 292 EN 12457-1 (1:2 solid/water ratio, 24 h shaking), whereas leachates from BOS samples showed Cr release in the order of one mmol L–1. The leachable Cr content is dominated by chromate as revealed by catalytic adsorptive stripping voltammetry (CAdSV) which could cause harmful contamination of surface and groundwater in the semiarid environment of Jordan.
Larger spectral bandwidth and higher photon flux density are the major advantages of multilayer monochromators over crystal-based devices. Especially for synchrotron-based hard X-ray microimaging applications the increased photon flux density is important in order to achieve high contrast and resolution in space and/or time. However, the modifications on the beam profile induced by reflection on a multilayer are a drawback which can seriously harm the performance of such a monochromator. A recent study [A. Rack, T. Weitkamp, M. Riotte, D. Grigoriev, T. Rack, L. Helfen, T. Baumbach, R. Dietsch, T. Holz, M. Krämer, F. Siewert, M. Meduna, P. Cloetens, E. Ziegler, J. Synchrotron Radiat. 17 (2010) 496510] has shown that the modifications in terms of beam flatness and coherence preservation can be influenced via the material composition of the multilayer coating. The present article extends this knowledge by studying further material compositions used on a daily basis for hard X-ray monochromatization at the beamline ID19 of the European Synchrotron Radiation Facility.
Synchrotron X-ray absorption edge imaging was used to investigate the ruthenium distribution in both fresh and aged Pt/Ru-based membrane electrode assemblies (MEA) of direct methanol fuel cells. MEAs aged in different ways were analyzed: artificially aged by MeOH depletion and aged for 1700 h in an operating fuel cell stack. An element sensitive tomographic technique – differential X-ray absorption edge tomography – was applied allowing for a 3D-visualization of the ruthenium distribution within the MEA. We found a markedly changed Ru distribution after aging which is correlated to the GDL structure, the flow field geometry, and CO2 transport in the methanol solution.
The three-dimensional water distribution and water transport paths in the gas diffusion layer (GDL) and the adjacent micro-porous layer (MPL) of a polymer electrolyte membrane fuel cell (PEMFC) were analyzed during cell operation. The technique of quasi in-situ X-ray tomography was used for a 3D visualization of the water distribution and the structure of the GDL at different operating conditions. Based on findings from in-situ radiographic measurements water transport paths were detected and subsequently examined by tomography. The combination of these 2D and 3D techniques allows for a fully three-dimensionally resolved visualization of transport paths through the GDL.
A new setup for fast spatially resolved measurements of elemental trace amounts under total reflection conditions using a new colour X-ray camera is described. Samples prepared on conventional total reflection X-ray fluorescence (TXRF) reflectors were measured at BESSY II synchrotron. A spatial resolution of 50 × 50 µm² was obtained, while the required time for the investigation of a 10 × 10 mm² sample is 30 seconds. The set-up is up to 350 times faster than conventional X-ray fluorescence systems for elemental traces. The major components of the X-ray camera are polycapillary optics and a pn-CCD chip with an active area of 13 × 13 mm². This area is divided into 264 × 264 pixels of 48 × 48 µm². A full X-ray spectrum with a resolution of 152 eV @ 5.9 keV and a chip temperature of 246 K is recorded for each pixel. The chip has a read-out rate of 400 Hz.
Biocompatible materials such as titanium are regularly applied in oral surgery. Titanium-based implants for the replacement of missing teeth demand a high mechanical precision in order to minimize micro-bacterial leakage, especially when two-piece concepts are used. Synchrotron-based hard x-ray radiography, unlike conventional laboratory radiography, allows high spatial resolution in combination with high contrast even when micro-sized features in such highly attenuating objects are visualized. Therefore, micro-gap formation at interfaces in two-piece dental implants with the sample under different mechanical loads can be studied. We show the existence of micro-gaps in implants with conical connections and study the mechanical behavior of the mating zone of conical implants during loading. The micro-gap is a potential source of implant failure, i.e., bacterial leakage, which can be a stimulus for an inflammatory process.