For the first time, an international comparison was conducted on the determination of the purity of a high purity element. Participants were free to choose any analytical approach appropriate for their institute’s applications and services. The material tested was a high purity zinc, which had earlier been assessed for homogeneity and previously used in CCQM-K72 for the determination of six defined metallic impurities. Either a direct metal assay of the Zn mass fraction was undertaken by EDTA titrimetry, or an indirect approach was used wherein all impurities, or at least the major ones, were determined and their sum subtracted from ideal purity of 100 %, or 1 kg/kg. Impurity assessment techniques included glow discharge mass spectrometry, inductively coupled plasma mass spectrometry and carrier gas hot extraction/combustion analysis. Up to 91 elemental impurities covering metals, non-metals and semi-metals/metalloids were quantified. Due to the lack of internal experience or experimental capabilities, some participants contracted external laboratories for specific analytical tasks, mainly for the analysis of non-metals. The reported purity, expressed as zinc mass fraction in the high purity zinc material, showed excellent agreement for all participants, with a relative standard deviation of 0.011 %. The calculated reference value, w(Zn) = 0.999 873 kg/kg, was assigned an asymmetric combined uncertainty of + 0.000025 kg/kg and – 0.000028 kg/kg. Comparability amongst participating metrology institutes is thus demonstrated for the purity determination of high purity metals which have no particular difficulties with their decomposition / dissolution process when solution-based analytical methods are used, or which do not have specific difficulties when direct analysis approaches are used. Nevertheless, further development is required in terms of uncertainty assessment, quantification of non-metals and the determination of purity of less pure elements and/or for those elements suffering difficulties with the decomposition process.
A magnetically stabilized DC arc device, designed for operation with OES spectrometers was used to determine the elements Ag, Al, B, Ba, Be, Ca, Co, Cr, Cu, Fe, Ga, In, K, Li, Mg, Mn, Mo, Na, Nb, Ni, Si, Sn, Sr, Ti, V, Zr at trace levels of some μg kg−1 up to some 10 mg kg−1 in graphite powders. The coil for the generation of the homogeneous magnetic field was placed outside the closed arc chamber. The time programs of variable current strengths of the magnetic coil (up to 6 A) and of the arc (up to 17 A) which was burning in air were computer controlled. Halogenating gases (mainly CCl2F2, alternatively SF6 and NF3) were used as chemical modifiers to allow an effective release of the carbide forming trace elements. The mass flow controlled modifier gas was led through a special carrier electrode near the arc plasma. The emission radiation was guided by an optical fiber alternatively into two different ICP spectrometers in which the ICP torches were removed. The synergistic interaction of the magnetic field with the halogenating modifier gases resulted in a significant improvement in the analytical performance of the optimized analytical method. All our results for 22 trace elements were in good agreement with the means of an inter-laboratory comparison by BAM for certification of a pure graphite powder material; this holds also for our results for two other graphite materials. The optimized method showed an analytical performance suitable for comprehensive trace analysis of pure graphite. The instrumentation could be integrated into modern DC arc emission spectrometers to improve their analytical capabilities substantially.
The analytical potential of a nanosecond laser ablation inductively coupled plasma mass spectrometer (ns-LA-ICP-SFMS) system, equipped with an ultra-fast wash-out ablation chamber, is critically investigated for fast and highly spatially resolved (∼μm) qualitative elemental distribution within single cells. Initially, a low surface roughness (< 10 nm) thin In–SnO2 layer (total coating thickness ∼200 nm) deposited on glass is employed to investigate the size, morphology and overlapping of laser-induced craters obtained at different laser repetition rates, making use of Atomic Force Microscopy (AFM). Conical craters with a surface diameter of about 2 µm and depths of about 100 nm were measured after a single laser shot. Furthermore, the influence of the sampling distance (i.e. distance between the sample surface and the inner sniffer of the ablation chamber) on the LA-ICP-MS ion signal wash-out time is evaluated. A significant decrease of the transient 120Sn+ ion signal is noticed after slight variations (±200 μm) around the optimum sampling position. Ultra-fast wash-outs (< 10 ms) are achieved reducing the aerosol mixing from consecutive laser shots even when operating the laser at high repetition rates (25 – 100 Hz). Fast and highly spatially resolved images of elemental distribution within mouse embryonic fibroblast cells (NIH/3T3 fibroblast cells) and human cervical carcinoma cells (HeLa cells), incubated with gold nanoparticles (Au NPs) and Cd-based quantum dots (QDs), respectively, are determined at the optimized operating conditions. Elemental distribution of Au and Cd in single cells is achieved using a high scanning speed (50 µm/s) and high repetition rate (100 Hz). The results obtained for the distribution of fluorescent Cd-based QDs within the HeLa cells are in good agreement with those obtained by confocal microscopy. The size, morphology and overlapping of laser-induced craters in the fixed cells are also investigated using AFM, observing conical craters with a surface diameter of about 2.5 µm and depths of about 800 nm after a single laser shot.
A candidate material for the use as primary standard for nickel determination was characterized with respect to total purity. For element determination a pure material can serve as primary standard. However, real world materials are never absolutely pure, hence the total purity of such materials need to be determined experimentally. A reasonable target uncertainty for the purity statement is 10-4 relative. Usually, the purer the material, the easier it is to achieve this target uncertainty. There are two basic ways in order to be able to establish a total purity statement. One is to determine the main component of a high purity material by a direct method such as coulometry, gravimetry or titrimetry. However, these methods are not selective enough for one element and therefore require certain efforts to analyse the material with respect to impurities with interfering analytes. Moreover, to reach the defined target uncertainty is not easy or often impossible to achieve. The second approach is to determine the sum of all possible impurities (as mass fraction) and to subtract it from the ideal purity of 100 % (1 kg/kg). In principle all impurities (all elements not being the matrix element), metals and non-metals must be considered. In this work both approaches to determine the total purity of the nickel material were followed and compared.
The primary (solid) standards are usually used to prepare primary calibration solutions to which secondary and lower order calibration solutions are linked.
Reference materials are essential, when the accuracy and reliability of measurement results need to be guaranteed in order to generate confidence in the analysis.
These materials are frequently used for determining measurement uncertainty, for validation of methods, suitability testing and quality assurance. Especially direct solid sampling methods require reference materials for calibration.
They guarantee that measurement results can be compared to recognized reference values.
This presentation gives an overview about the use of GDMS in various certification procedures. Because it represents a fast, sensitive, multielement analyses technique without extensive sample preparation it plays a special role for the purity determination of high purity standards. Various calibration strategies and the preparation of traceable matrix matched calibration standards will be discussed.
For the certification of analyte content in matrix materials mainly techniques with solvent sample preparation are used. Here GD-MS is used to identify possible loss or contamination with analytes during the sample preparation step. Typically used acids to dissolve matrices lead to interferences in the ICP- mass spectrometric detection of various analytes and their quantification. Here GD-MS as direct method can also add an important contribution in the certification process.
Analytical glow discharges with optical (GD-OES) and mass spectrometric (GD-MS) detection are able to obtain depth resolved information about the light elements hydrogen, carbon, nitrogen and oxygen in solid samples, where most of the other analytical techniques fail. However, the interpretation or even quantification of the measured signals is still very challenging. Problems arise due to physical effects (plasma processes such as the ‘Hydrogen effect’, the Doppler effect, self absorption or diffusion of hydrogen in the sample during sputtering) chemical effects(e.g. formation of compounds with argon or the matrix, poisoning of the sample or gettering) as well as instrumental difficulties (e.g. of the sensitivity, calibration and vacuum quality)..
The GD techniques are direct solid sampling methods and require reference materials for calibration. Unfortunately the list of available certified reference materials (CRM) suited for calibration of light elements in different matrix is relatively short Therefore, sintered materials doped with the analytes H2, O2 and N2 were produced at IFW Dresden and applied as calibration standards for hydrogen, oxygen and nitrogen. Due to the high analyte concentration added, it is very likely that the real concentration agree well with the added amount of light elements in the corresponding phases. The validation of the determination of the light element concentration in the sintered samples was possible in some cases only due to the lack of suitable techniques. This fact proves the need for the development of a reliable quantification of light elements by GDS.
A systematic dependence of the sputtering rate on the composition was found and can be explained by basic principles.
Using mixtures of TiH2 and ZrH2 with Cu the ratio of hydrogen and Cu lines has a good correlation with the corresponding concentration ratio. The hydrogen emission yield however decreases over 0.3 m% hydrogen and finally the hydrogen intensity may even decrease. This behaviour can be explained by a very similar quenching of the hydrogen and copper intensity caused by the hydrogen effect. First experiments with GD-MS show no saturation of the hydrogen ion current and thus confirm the quenching of the emission yield in GD-OES.
Sintered material for the oxygen calibration (Al2O3, CuO, Cu2O and MgO mixed with Cu, Al and Mg) confirmed the blue line shift effect at O I 130.22 nm, first time reported by Köster 2009 [1]. The effect is more pronounced at Mg than in Al and Cu, which due to line interference leads to a matrix dependent EY. This effect is negligible at O I 777.19 nm and the EY is matrix independent. Using GD-MS first promising results for the calibration of oxygen with these sintered samples could be obtained, when the sputtering rate was included in the evaluation.
More recently also sintered material for nitrogen calibration (AlN mixed with Al and Si3N4 mixed with Cu) was produced and points to a matrix independent emission yield of nitrogen
Die Junganalytiker der Fachgruppe „Analytische Chemie“ der Gesellschaft Deutscher Chemiker (GDCh) veranstalteten in Zusammenarbeit mit der BAM eine Herbstschule zum Thema „Qualitätssicherung in der Analytischen Chemie“.
Schulung v.a. für Doktoranden, um den Umgang mit qualitätssichernden Aspekte bei der Entwicklung analytischer Methoden zu erlernen.
Die wichtigsten Grundbegriffe der Genauigkeitskenngrößen von Messgrößen werden in diesem Vortrag vermittelt.