Determination of the efficiency of an energy dispersive X-ray spectrometer up to 50 keV with a SEM
(2009)
Both electron and polychromatic photon excitations (micro-focus X-ray source) at a scanning electron microscope (SEM) are used to determine the efficiency of an energy dispersive X-ray spectrometer up to 50 keV by means of a calibrated X-ray spectrometer and reference materials (RM) specially selected for this purpose.
The accurate knowledge of the properties of polycapillary X-ray semi-lenses has a significant influence on the quantitative results in a widespread field of applications involving microfocus X-ray beams. A routine procedure for the characterisation of a polycapillary X-ray semi-lens with a scanning electron microscope (SEM) having attached an energy dispersive spectrometer (EDS) is presented in this paper. A key issue of the procedure consists of fitting the semi-lens in front of the EDS for spectra acquisition. Relevant semi-lens parameters such as focal distance, full width at half maximum (FWHM) of the acceptance area, and transmission are determined in parallelising mode of the semi-lens. Special attention has been paid to the calculation of the transmission.
Checking the performance of energy dispersive X-ray spectrometers as well as validation of the results obtained with energy dispersive X-ray spectrometry (EDX) at a scanning electron microscope (SEM) involve the use of (certified) reference and dedicated test materials. This paper gives an overview on the test materials mostly employed by SEM/EDX users and accredited laboratories as well as on those recommended in international standards. The new BAM reference material EDS-CRM, which is currently in the process of certification, is specifically designed for the characterization of EDS systems at a SEM through calibration of the spectrometer efficiency in analytical laboratories in a simple manner. The certification of the spectra by means of a reference EDS is described. The focus is on the traceability of EDS efficiency which is ensured by measurements of the absolute detection efficiency of silicon drift detectors (SDD) and Si(Li) detectors at the laboratory of the PTB using the electron storage ring BESSY II as a primary X-ray source standard. A new test material in development at BAM for testing the performance of an EDS in the energy range below 1 keV is also briefly presented.