Efficiency and response functions were determined for a modern commercial energy-dispersive x-ray spectrometer in the photon energy range 0.1-5 keV using dispersed and undispersed synchrotron radiation. The spectrometer was equipped with an Si(Li) detector crystal, a thin-film window and a digital pulse processor. Monochromatized synchrotron radiation from the PTB VUV radiometry beamline at the electron storage ring BESSY was used for the absolute determination of the detection efficiency with a typical relative uncertainty of 1-2% in the photon energy range 0.1-1.5 keV by direct comparison with calibrated photodiodes. At higher photon energies, the efficiency was found by comparison of the measured and calculated undispersed synchrotron radiation spectrum. The absolute intensity of the synchrotron radiation was known with a relative uncertainty of less than 1%. In the overlapping region, the results from the two independent experiments are in full agreement. The energy dependence of the measured efficiency can be explained only with a detector model assuming that there is no dead layer. A simple model for the effect of incomplete charge collection (ICC) was applied to describe the measured response functions. Consequences of the ICC such as broadening and shift of low-energy peaks and redistribution of counts around 1.84 keV are explained with the model, in accordance with the experimental results.
Measurement and calculation of x-ray production efficiencies for copper, zirconium, and tungsten
(2022)
Electron probe microanalysis (EPMA) is based on physical relations between measured X-ray intensities of characteristic lines and their Xray production efficiency, which depends on the specimen composition. The quality of the analysis results relies on how realistically the physical relations describe the generation and emission of X-rays. Special experiments are necessary to measure X-ray production efficiencies. A challenge in these experiments is the determination of the detection efficiency of the spectrometer as a function of the photon energy. An energy-dispersive spectrometer was used in this work, for which the efficiency was determined at metrological synchrotron beamlines with an accuracy of ±2%. X-ray production efficiencies for the L series and the Kα series of copper and zirconium and for the M and L series of tungsten were determined at energies up to 30 keV in a scanning electron microscope. These experimental values were compared with calculated X-ray production efficiencies using physical relations and material constants applied in EPMA. The objective of the comparison is the further improvement of EPMA algorithms as well as extending the available database for X-ray production efficiencies. Experimental data for the X-ray production efficiency are also useful for the assessment of spectrum simulation software.
Energy dispersive X-ray spectrometry offers the opportunity for fast composition determination of specimens by X-ray fluorescence or electron probe microanalysis. For fundamental parameter based quantification, the knowledge of the detection efficiency of the spectrometer is essential. At low energies the efficiency is strongly influenced by X-ray absorption in the radiation entrance window. State-of-the-art windows consist of polymer foil containing C, N, and O, coated with Al and in some cases with a special B compound. The foil is supported by a Si grid to withstand the atmosphere pressure. The absorption of all these components must be known to describe the detection efficiency.
The transmittance of three types of widely used commercial windows has been measured. Transmittance curves have been fitted by analytical expressions using tabulated mass absorption coefficients. Because tabulated mass absorption coefficients do not consider near edge effects, there are strong deviations between measured and calculated transmittance below 0.6 keV. It is proposed to model the spectrometer efficiency by the measured window transmittance and calculated absorptions from front contact and possible contaminations. This reduces the number of unknown parameters drastically.
The X-ray intensities of the K-, L- and M-lines of copper, zirconium and tungsten have been measured with an energy-dispersive X-ray spectrometer of known efficiency as function of photon energy. X-ray production efficiencies were determined from the measured intensities for Kα- and L-series of Cu and Zr and for the L- and M-series of W. These data were compared to calculated X-ray production efficiencies based on the widely used matrix correction models of Pouchou and Pichoir (XPP) and Bastin (PROZA96).
Our results indicate that a replacement of the stopping power in the PROZA96 algorithm by expressions of Joy and Jablonski has only a minor influence on the calculated X-ray production efficiencies. In contrast, the modifications of the ionization cross-section show a stronger effect. We replaced the ionization cross-sections for K lines of the PROZA96 algorithm with different models.
The results for L- and M-Lines are different. For the L-lines of Cu the original XPP and PROZA96 models show the best agreement while using the Bote cross-sections result in an overestimation. For the Zr-L and W-L1, -L2, -L3 X-ray production efficiencies, the Bote cross-sections lead to a significant improvement compared to all other models. The original XPP model represents the best agreement for the M5 efficiencies but underestimates the M4 efficiencies.
There is no superior model or modification because the parameter sets in the models need to be aligned to each other. However, using the ionization cross-sections of Bote, which are based on quantum mechanical calculations, show promising results in many cases.
High-accuracy film thickness measurements in the range below 100 nm can be made by various complex methods like spectral ellipsometry (SE), scanning force microscopy (SFM), grazing incidence X-ray reflectometry (GIXR), or X-ray fluorescence analysis (XRF). The measurement results achieved with these methods are based on different interactions between the film and the probe. A key question in nanotechnology is how to achieve consistent results on a level of uncertainty below one nanometre with different techniques.
Two different types of thickness standards are realised. Metal film standards for X-ray techniques in the thickness range 10 to 50 nm are calibrated by GIXR with monochromatised synchrotron radiation of 8048 eV. The results obtained at four different facilities show excellent agreement. SiO2 on Si standards for SE and SFM in the thickness range 6 to 1000 nm are calibrated by GIXR with monochromatised synchrotron radiation of 1841 eV and with a metrological SFM. Consistent results within the combined uncertainties are obtained with the two methods. Surfaces and interfaces of both types of standards are additionally investigated by transmission electron microscopy (TEM).