Scanning electron microscopes are usually equipped with energy-dispersive X-ray detectors for electron probe microanalysis. This widespread analytical method allows investigators to determine the elemental composition of specimens with a spatial resolution of about 1 µm. However, owing to the electron-specimen interaction, the emitted spectra reveal, in addition to characteristic lines, also a high level of continuous bremsstrahlung background. As a result, elements with low concentrations cannot be identified. The minimum detection limit can be diminished by two orders of magnitude if the characteristic lines are excited as fluorescence by an additional x-ray source. In this case, the emergence of bremsstrahlung is considerably reduced. Combining a high-brilliance microfocus x-ray tube with efficient polycapillary optics enables one to realize an experimental arrangement for performing local fluorescence analysis at the same point where the electron beam hits the sample. The polycapillary optics under consideration focuses the emitted x-radiation onto focal spots between 30 and 100 µm in diameter. Count rates of several thousands cps have been achieved. Elemental maps have been obtained by means of the motorized specimen stage of the microscope. Copyright © 2005 John Wiley & Sons, Ltd.
X-ray Fluorescence (XRF) with a scanning electron microscope (SEM) is a valuable completion of the analytical capabilities of SEMs. Small and compact micro-focus x-ray sources are mounted to the microscope chamber, and the x-ray spectra are monitored with conventional EDS systems. Up to now the x-ray tubes used for the micro-focus x-ray sources are equipped with beryllium windows about 100 µm thick. The poly-capillary x-ray lenses have their transmission maximum at photon energies around 10 keV. It drops down in both low- and high-energy ranges. Hence, L-radiation from an Mo or Rh target will be strongly attenuated, and the excitation of fluorescence in the soft x-ray range becomes very ineffective. A new micro-focus x-ray source was developed. It is characterised by a lower self-absorption in the tube target, thin beryllium windows and an x-ray optics having a large distance between its foci and the maximum of transmission at about 5 keV. Thus K line fluorescence of light elements becomes effectively excited by the L-radiation from Mo or Rh tube targets. The detection limit for sodium oxide in glass was found to be below 1 mass%.
The test material EDS-TM001 together with an accompanying software package, “EDX spectrometer check”, have been made available in 2009 by BAM to be employed by EDS (energy dispersive spectrometer) users to check the performance of an EDS attached to the SEM. Particularly for test laboratories operating under accreditation schemes like ISO/IEC 17025, a periodical control of the critical instrumental parameters in end-user laboratories is required. With EDS-TM001 or EDS-TM002 (second generation) test material, this periodical check is simplified to the acquisition of only one 10 kV spectrum. The software “EDX spectrometer check” is destined to evaluate automatically this spectrum and determine the performance of the EDS in terms of energy resolution and calibration as well as possible alteration of low-energy efficiency due to detector contamination. Energy resolution can be compared with the specified values according to the international ISO standard ISO 15632:2012. EDS-TM is a synthetic material consisting of a thick layer of C, Al, Mn, Cu and Zr in a well-defined composition, deposited on a steel (in case of EDS-TM001) or silicon (in case of EDS-TM002) substrate. Meanwhile, more than one hundred laboratories use the EDS-TM001 or EDS-TM002 test material for the periodical check of their EDS. A detailed description of the test material and software together with examples of application was published recently. New results and gained experiences will be presented as well.
When the FWHM of lines appearing in the EDS-TM spectrum are determined, the spectrum background must be subtracted accurately. The applied physical background subtraction procedure is robust and takes into account the transmission of the detector window. While the previous version considers only Moxtek AP windows, the new version includes selection of silicon nitride window and the case of windowless detector. Moreover, the new version allows importing of spectra in Bruker spx format and EMSA/MSA files from EDAX TEAM software.