Filtern
Erscheinungsjahr
Dokumenttyp
- Zeitschriftenartikel (10)
- Posterpräsentation (4)
- Corrigendum (1)
Schlagworte
- GDMS (3)
- Calibration (2)
- Extractable organically bound fluorine (EOF) (2)
- Fluorination (2)
- High resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) (2)
- Adsorbable organically bound fluorine (AOF) (1)
- Aluminium (1)
- Calibration samples (1)
- Combustion ion chromatography (CIC) (1)
- Copper (1)
Organisationseinheit der BAM
Here, we describe an optimized fast and simple extraction method for the determination of per- and polyfluorinated alkyl substances (PFASs) in soils utilizing high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR–CS–GFMAS). To omit the bias of the solid phase extraction (SPE) step commonly used during the analysis of extractable organically bound fluorine (EOF) we optimized a fast and simple SPE-free extraction method. The developed extraction method consists of a liquid-solid extraction using acidified methanol without any additional SPE. Four extraction steps were representative to determine a high proportion of the EOF (>80% of eight extractions). Comparison of the optimized method with and without an additional SPE clean-up step revealed a drastic underestimation of EOF concentrations using SPE. Differences of up to 94% were observed which were not explainable by coextracted inorganic fluoride. Therefore, not only a more accurate but also a more economic as well as ecologic method (bypassing of unnecessary SPE) was developed. The procedural limit of quantification (LOQ) of the developed method was 10.30 μg/kg which was sufficient for quantifying EOF concentrations in all tested samples. For future PFAS monitoring and potential regulative decisions the herein presented optimized extraction method can offer a valuable contribution.
The article “Determination of organically bound fluorine sum parameters in river water samples - Comparison of combustion ion chromatography (CIC) and high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS)” was originally published Online First without Open Access. After publication in volume 413, issue 1, page 103–115 of the journal “Analytical and bioanalytical chemistry”, the authors decided to opt for Open Choice and to make the article an Open Access publication. Therefore, the copyright of the article has been changed to © The Author(s) 2021 and the article is forthwith distributed under the terms of the Creative Commons Attribution 4.0 International License.
The quantification of the sulphur mass fraction in pure copper and copper alloys by GDMS and LA-ICP-MS revealed a lack of traceability mainly due to a lack of suitable certified reference materials for calibrating the instruments. Within this study GDMS and LA-ICP-MS were applied as routine analytical tools to quantify sulphur in copper samples by applying reference materials as calibrators, which were characterized for their sulphur mass fraction by IDMS beforehand. Different external calibration strategies were applied including a matrix cross type calibration. Both techniques with all calibration strategies were validated by using certified reference materials (others than those used for calibration) and good agreement with the reference values was achieved except for the matrix cross type calibration, for which the agreement was slightly worse. All measurement results were accompanied by an uncertainty statement. For GDMS, the relative expanded (k = 2) measurement uncertainty ranged from 3% to 7%, while for LA-ICP-MS it ranged from 11% to 33% when applying matrix-matched calibration in the sulphur mass fraction range between 25 mg kg-1 and 1300 mg kg-1. For cross-type calibration the relative expanded (k = 2) measurement uncertainty need to be increased to at least 12% for GDMS and to at least 54% for LA-ICP-MS to yield metrological compatibility with the reference values. The so obtained measurement results are traceable to the international system of units (SI) via IDMS reference values, which is clearly illustrated by the unbroken chain of calibrations in the metrological traceability scheme.
Fast flow glow discharge mass spectrometry with a Grimm-type ion source providing a high sputter rate was used for the determination of major nonmetallic impurities in magnesium. The analytical signal was found to be strongly influenced by the electrical discharge parameters. For calibration by standard addition, synthetic standard samples were produced in two different ways—namely, by pressing and by sintering doped metal powders. The observed sensitivity of the calibration curves was shown to depend on the particle size of the powder. For the magnesium powders, the mass fractions of oxygen, nitrogen, boron, and silicon were determined to be about 0.01 kg·kg-1 (relative standard deviation approximately 10–20 %), 2,700 mg·kg-1, 150 mg·kg-1, and 300 mg·kg-1, respectively.
Direct current (DC) fast flow glow discharge mass spectrometry is an important and versatile tool for multielemental trace analysis of conductive solid materials down to the µg/kg level. Special benefits are minimum demands on sample preparation, avoiding losses and contamination, and only short analysis time.
For fast flow GD sources, the quantification strategy based on relative standard sensitivity factors (Standard RSFs) which are independent of matrix is not sufficiently satisfying regarding accuracy for each matrix and element [1]. Therefore matrix-matched calibration samples (MMCS) are required to achieve reliable quantification. In fact, the list of existing certified reference materials (CRM) appropri-ate for calibration in trace analysis is rather short.
Convenient synthesis of homogeneous MMCS, as e.g. easily obtained in liquid sampling spectrometry, can facilitate the application of fast flow GD-MS for quantification of impurities in a variety of matrices.
Pressing of metal powders for the preparation of MMCS for GD-MS was suggested earlier [2], the method was further modified by use of analyte solutions for doping of rather pure metal powders such as Cu and Zn [3], but has not yet been applied Ni matrix.
In the present work we describe the determination of matrix-dependent relative sensitivity factors (RSFs) for Mg, Al, Cr, Mn, Fe, Co, Cu, Zn, Ag, Cd, Tl, Pb and Bi in pure Ni by using the liquid doping approach for the preparation of synthetic pressed Ni-powder samples. A four point-calibration was used applying the fast flow glow discharge mass spectrometer, Element GD (Thermo Fisher). The determined RSF were verified against suitable CRM and compared with the standard RSF given by the supplier of the instrument.
The obtained results demonstrate a satisfying Agreement with the certified values of the CRM and a significant improvement for the quantification of most of the determined elements in comparison with standard RSF.
The formation of diffusion layers on Zn layers deposited on Al substrates is mainly used to prevent corrosion effects. Evaluation of the influence exerted by different coating methodologies and heat treatments on the formation of these diffusion layers is of great interest for the aluminium industry. Particularly, multi-elemental in-depth distributions of major, minor and trace elements in Zn-coatings is highly demanded before and after heat treatments. A fast characterization of these materials require a direct solid analytical technique able to provide high sensitivity and high depth resolution. For this purpose, an improved analytical method based on the use of pulsed glow discharge sector field mass spectrometry (pulsed-GD-SFMS) is investigated. Glow discharge operating parameters (e.g. pulse duration, pulse frequency), glow discharge source design (e.g. flow tube lengths), and SFMS mass spectra acquisition conditions (e.g. integration time) are evaluated to achieve low sputtering rates, high mass spectra acquisition rates and improved depth resolution. At the optimize conditions Zn coatings deposited by arc-spray and electrodeposition are analysed before and after heat treatments to evaluate the diffusion of different key elements. Moreover, results are validated using femtosecond laser ablation (fs-LA)-ICP-MS, which provides additional information about the heterogeneous distribution of some elements in the Zn coatings.
In this study, we compare combustion ion chromatography (CIC) and high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) with respect to their applicability for determining organically Bound fluorine sum parameters. Extractable (EOF) and adsorbable (AOF) organically bound fluorine as well as total fluorine (TF) were measured in samples fromriver Spree in Berlin, Germany, to reveal the advantages and disadvantages of the two techniques used as well as the two established fluorine sum Parameters AOF and EOF. TF concentrations determined via HR-CS-GFMAS and CIC were comparable between 148 and 270 μg/L. On average, AOF concentrations were higher than EOF concentrations, with AOF making up 0.14–0.81% of TF (determined using CIC) and EOF 0.04–0.28% of TF (determined using HR-CSGFMAS).
The results obtained by the two independent methods were in good agreement. It turned out that HR-CS-GFMAS is a more sensitive and precise method for fluorine analysis compared to CIC. EOF and AOF are comparable tools in Risk evaluation for the emerging pollutants per- and polyfluorinated alkyl substances; however, EOF is much faster to conduct.