X-ray scattering datasets for samples described in the 2022 publication "Molecular Mobility of Polynorbornenes with Trimethylsiloxysilyl side groups: Influence of the Polymerization Mechanism". This dataset includes both raw and processed X-ray scattering data for samples APTCN and MPTCN, alongside background measurements files (BKG).
X-ray scattering datasets for samples described in the 2022 publication "Side chain length dependent dynamics and conductivity in self assembled ion channels". This dataset includes both raw and processed X-ray scattering data for samples ILC8, ILC10, ILC12, ILC14 and ILC16 alongside background measurement files (BKG).
McSAS3
(2023)
McSAS3 is a refactored version of the original McSAS (see DOI 10.1107/S1600576715007347). This software fits scattering patterns to obtain size distributions without assumptions on the size distribution form. The refactored version has some neat features:
- Multiprocessing is included, spread out over as many cores as number of repetitions!
- Full state of the optimization is stored in an organized HDF5 state file.
- Histogramming is separate from optimization and a result can be re-histogrammed as many times as desired.
- SasModels allow a wide range of models to be used
- If SasModels does not work (e.g. because of gcc compiler issues on Windows or Mac), an internal sphere model is supplied
- Simulated data of the scattering of a special shape can also be used as a McSAS fitting model. Your models are infinite!
- 2D fitting also works.
These are four datasets that were made available to the participants of the Small-angle Scattering data analysis round robin. The intent was to find out how comparable results from different researchers are, who analyse exactly the same processed, corrected dataset.
In this repository, there are:
1) a PDF document with more details for the study,
2) the datasets for people to try and fit
3) an Excel spreadsheet to document the results.
Datasets 1 and 2 were modified from: Deumer, Jerome, & Gollwitzer, Christian. (2022). npSize_SAXS_data_PTB (Version 5) [Data set]. Zenodo. https://doi.org/10.5281/zenodo.5886834
Datasets 3 and 4 were collected in-house on the MOUSE instrument.
Small-angle Scattering Data Analysis Round Robin: anonymized results, figures and Jupyter notebook
(2023)
The intent of this round robin was to find out how comparable results from different researchers are, who analyse exactly the same processed, corrected dataset.
This zip file contains the anonymized results and the jupyter notebook used to do the data processing, analysis and visualisation. Additionally, TEM images of the samples are included
## Summary:
This notebook and associated datasets (including VASP details) accompany a manuscript available on the ArXiv (https://doi.org/10.48550/arXiv.2303.13435) and hopefully soon in a journal as short communication as well. Most of the details needed to understand this notebook are explained in that paper with the same title as above. For convenience, the abstract is repeated here:
## Paper abstract:
We demonstrate a strategy for simulating wide-range X-ray scattering patterns, which spans the small- and wide scattering angles as well as the scattering angles typically used for Pair Distribution Function (PDF) analysis. Such simulated patterns can be used to test holistic analysis models, and, since the diffraction intensity is presented coupled to the scattering intensity, may offer a novel pathway for determining the degree of crystallinity.
The ``Ultima Ratio'' strategy is demonstrated on a 64-nm Metal Organic Framework (MOF) particle, calculated from $Q<0.01$\,$\mathrm{nm}^{-1}$ up to $Q\approx150$\,$\mathrm{nm}^{-1}$, with a resolution of 0.16\,\AA. The computations exploit a modified 3D Fast Fourier Transform (3D-FFT), whose modifications enable the transformations of matrices at least up to $8000^3$ voxels in size. Multiple of these modified 3D-FFTs are combined to improve the low-$Q$ behaviour.
The resulting curve is compared to a wide-range scattering pattern measured on a polydisperse MOF powder.
While computationally intensive, the approach is expected to be useful for simulating scattering from a wide range of realistic, complex structures, from (poly-)crystalline particles to hierarchical, multicomponent structures such as viruses and catalysts.
Electric Safety Interlock
(2023)
This interlock is designed to prevent electrical shock from high voltage (>60V) equipment.
While the general safety interlock can be generically applied, this particular example employs an external vacuum-activated switch. It is for safeguarding human operations inside a vacuum sample chamber while the chamber doors are open. The circuit is closed (output is active) when a sufficient level of vacuum is reached, i.e. when all accessible openings are necessarily closed.
The initial application is to interrupt power to a 220V, 250W heating cartridge (itself mounted inside a small sample holder with potentially exposed contacts) when the sample chamber is open.
The external circuit can be modified to use different interlock mechanisms as needed. Note that the external interlock circuit is only a single circuit (with two signal lines) and thus is not protected against external shorts.
To accomodate a range of safety interlocks, the 4-pin M12 connector is wired as follows: Pin 1 (Brown): +24V for power supply, max current 0.6A Pin 2 (White): Safety interlock system signal 1 (0 or 24V) Pin 3 (Blue) : Safety interlock system signal 2 (0 or 24V) Pin 4 (Black): 0V for power supply
The safety is interlocked (output active) when both signal pins are set high (24V), with sufficient current to activate the two relays. Pin 1 and 4 can be used to power safety hardware (such as light curtains or proximity detectors) with 24VDC up to a current of 0.6A. A larger power supply can be installed when higher currents are needed, while staying within the current limits imposed by the wiring cross-section.