- Copper (3)
- Calibration (2)
- Chemical analysis (2)
- ERM (2)
- Traceability (2)
- Zinc (2)
- Arrhenius model (1)
- Boron (1)
- CRM (1)
- Certified reference materials (1)
Calibration of two commercially available glow discharge double focusing mass spectrometers, the VG 9000 and Element GD, is described using synthetic pin standards pressed from solution doped copper and zinc matrices. A special pressing die was developed for this purpose and optimal results were obtained with the highest possible pressures, i.e., 95 kN·cm- 2. This calibration approach permits the determination of trace element mass fractions down to µg·kg- 1 with small uncertainties and additionally provides traceability of the GD-MS results in the most direct manner to the SI (International System of Units). Results were validated by concurrent measurements of a number of compact copper and zinc certified reference materials. The impact of the sample pin cross-section (circular or square) was investigated with the use of a new pin-sample holder system for the Element GD. The pin-sample holder was designed by the manufacturer for pin-samples having circular cross-section; however, samples with square pin cross-section were also shown to provide acceptable results. Relative Sensitivity Factors for some 50 analytes in copper (VG 9000, Element GD) and zinc matrices (VG 9000) are presented. The field of applicability of GD-MS may be considerably extended via analysis of pin geometry samples based on their ease of preparation, especially with respect to the accuracy and traceability of the results and the enhanced number of analytes which can be reliably calibrated using such samples.
The application of metrology in chemical analysis is today hampered by psychological and practical barriers. The argumentation is based on the differences in physical and chemical metrology, which arises mainly from the objects of the measurements. The term measurement is discussed in the realm of analytical chemistry with a special focus on the inherent relation between chemical identification and determination as well as the consequent use of the concept of traceability and uncertainty.
ISO Guide 35 deals with RM stability issues and scrutinizes the evaluation of stability testing results under the assumption that either there is no trend at all (a rather rare situation), or any observed deterministic change is insignificant and thus can be neglected. However, market demands for reliable reference materials are obviously not limited to stable or at least seemingly stable materials. In many analytical applications, analytes and measurands under consideration are known, or at least suspected, to be unstable on time scales that may vary widely from measurand to measurand. The Federal Institute for Materials Research and Testing (BAM) has developed (and successfully uses) an integrated approach in its certification practice. The approach is based on an initial stability study and subsequent post-certification monitoring. Data evaluation is model-based and takes advantage of all information collected in the stability testing scheme(s). It thus allows one to deal with any kind of instability observed, to assess limiting time intervals at any stress condition in the range tested, to estimate a final expiry date for materials with detected instabilities or the maximum admissible re-testing interval for seemingly stable materials, and to assess maximum admissible stress loads during delivery of the material to the customer. The article describes (and exemplifies) typical study layout, the model selection, and the integrated data assessment.
Mass spectrometric detection can play a prominent role in the assessment of different reactions occurring at varied temperatures in a milled ZnO/Al mixture. This is the first time that online mass spectrometric information was used as a tool for monitoring and understanding the chemical reduction process mechanism. We have observed four different types of reaction taking place: (1) distillation of Zn metal, (2) reduction of ZnO by activated Al, (3) melting of Al and finally (4) reduction of ZnO by inactivated Al. The experimental conditions and results observed by QMS were supported with literature data and physical measurement data from X-Ray Diffraction (XRD) which gave us an idea about the complex reaction cascade which occurred during the formation of the zinc metal.
Synthetic pressed metal powder standards doped with standard solutions were used for the calibration of both commercially available high resolution GD-MS instruments Element GD and VG 9000 for zinc matrix. Different quantification procedures (IBR, Standard RSF, matrix matched RSF from the calibration with CRMs and use of doped synthetic standards) are compared using zinc matrix as an example, whereas the calibration with doped pellets turned out to be the best quantification technique for high-purity materials. The applicability of the Standard RSF concept is scrutinised. In this context, RSF values for several Matrices (Co, Cu, Fe, In and Zn) are reported additionally.
In Germany a system of primary amount of substance standards for elemental analysis is being developed, which forms the material´s basis for the National Standards for element determination. The metrological concept and technical approach is briefly described. The objective of traceability in chemistry, in this case for elemental determination, is illustrated on the example of copper.
In this work, different calibration strategies for the determination of trace elements in pure copper metal by nanosecond laser ablation ICP-MS were investigated. In addition to certified reference materials (CRMs), pellets of doped copper powder were used for calibration. The micro homogeneity of the CRMs as well as the solution-doped pellets was sufficient to use them as calibration samples in combination with a laser spot size of 200 µm. In contrast, pellets doped with analytes in solid form showed a significant heterogeneity. For most of the investigated analytes and copper CRMs the measured mass fractions were within ± 20% of their certified values when other copper CRMs were used as calibration samples. When solution-doped powder pellets were used as calibration samples a systematic trend towards mass fractions below the certified values was observed for nearly all elements determined in the analysed CRMs. Thermal fractionation effects during the ablation of the solution-doped pellets were suspected as the extent of the fractionation depends on the irradiance, whereas fractionation is reduced at higher irradiance.