In case study one of the CONSENS project, two aromatic substances were coupled by a lithiation reaction, which is a prominent example in pharmaceutical industry. The two aromatic reactants (Aniline and o-FNB) were mixed with Lithium-base (LiHMDS) in a continuous modular plant to produce the desired product (Li-NDPA) and a salt (LiF). The salt precipitates which leads to the formation of particles. The feed streams were subject to variation to drive the plant to its optimum.
The uploaded data comprises the results from four days during continuous plant operation time. Each day is denoted from day 1-4 and represents the dates 2017-09-26, 2017-09-28, 2017-10-10, 2017-10-17.
In the following the contents of the files are explained.
Die NMR-Spektroskopie stellt heutzutage eine der wichtigsten Analysenmethoden in der organischen Chemie dar. Während der Großteil aller Untersuchungen qualitativ mit dem Ziel der Stoffidentifikation und Strukturaufklärung erfolgt, erlangt die quantitative NMR-Spektroskopie (qNMR) zunehmend an Bedeutung in Forschung und Industrie. Der entscheidende Vorteil gegenüber anderen Analysenmethoden liegt in der direkten Proportionalität der Signalfläche zur Anzahl Kernspins im Messvolumen. Dies erlaubt eine kalibrationsfreie Relativquantifizierung. Zur Absolutquantifizierung reicht die Zugabe einer definierten Menge eines vom Analyten unabhängigen NMR-Standards aus.
Trotz dieser Vorteile findet sich die qNMR bislang nur vereinzelt in Normen und Standardverfahren wieder. Zahlreiche Ringversuche in Metrologie und Industrie demonstrieren die Leistungsfähigkeit moderner NMR-Spektrometer und stärken das Vertrauen in die Methode. Die Entwicklung von Validierungskonzepten, sowie die kommerzielle Verfügbarkeit geeigneter zertifizierter Referenzmaterialien erleichtern die Anwendung, insbesondere im zumeist stark regulierten industriellen Umfeld.
Neben etablierten Hochfeld-NMR Spektrometern hat sich in den letzten Jahren ein stark wachsender Markt für kompakte Benchtop-NMR Geräte auf Permanentmagnetbasis entwickelt. Die geringeren Anschaffungs- und Betriebskosten, sowie die einfache Bedienbarkeit erlaubt es auch kleineren Unternehmen in diese Analysenmethode einzusteigen. Weiterhin besteht die Möglichkeit diese mobilen Systeme näher an die Produktion zu bringen, welches von der klassischen Qualitätskontrolle bis hin zur Online-Prozesskontrolle als vollautomatisierter Analysator reicht. Die geringere Feldstärke der Systeme erfordert hier oft den Einsatz modellbasierter Ansätze zur Spektrenauswertung (z.B. Indirect Hard Modeling).
Dieser Beitrag gibt eine Übersicht über aktuelle Anwendungen und Entwicklungen der qNMR von der universellen, hochgenauen Labormethode bis hin zur robusten Anwendung als Online-Analysator im Feld.
Quantitative 1H Nuclear Magnetic Resonance (qNMR) of Aromatic Amino Acids for Protein Quantification
(2022)
qNMR is a valuable technique for metrological studies due to the uniformity of its signal response for all chemical species of an isotope of interest, which enables compound-independent calibration. However, protein quantification remained challenging as large molecules produce wide, low-intensity signals that reduce the already low sensitivity. Combining qNMR with the hydrolysis of protein samples into amino acids circumvents many of these issues and facilitates the use of NMR spectroscopy for absolute protein and peptide quantification.In this work, different conditions have been tested for quantifying aromatic amino acids and proteins. First, we examined the pH-based signal shifts in the aromatic region. The preferable pH depends on the selection of the amino acids for quantification and which internal standard substance should be used to avoid peak overlap. Several aromatic compounds, such as terephthalic acid, sulfoisophthalic acid, and benzene tricarboxylic acid, have been applied as internal standards. The quantification of amino acids from an amino acid standard, as well as from a certified reference material (bovine serum albumin), was performed. Using the first two suggested internal standards, recovery was ~ 97 % for histidine, phenylalanine, and tyrosine at a concentration of approximately 1 mM in solution. Acidic hydrolysis of a certified reference material (CRM) of bovine serum albumin (BSA) and subsequent quantification of Phe and Tyr yielded recoveries of 98 ± 2 and 88 ± 4 %, respectively, at a protein concentration of 16 g/L or 250 µM.
Der Vortrag gibt eine Übersicht über aktuelle qNMR-Aktivitäten im Fachbereich 1.4. Dies umfasst beispielsweise die Online-Anwendung zur Prozessüberwachung, aber auch die Reinheitsuntersuchung von Reinstoffkomponenten für die Herstellung von Referenzgasgemischen. Zusätzlich wird ein Ausblick in die aktuelle Strategie hinsichtlich der Rückführung von qNMR-Ergebnissen mittels primärer Referenzmaterialien gegeben.
qNMR - The handbook
(2023)
Quantitative NMR (qNMR) has been around for a long time, but also has great potential to solve future problems in any quantitative analysis. As a primary method, it differs fundamentally from chromatographic methods: it is better described as a quantum mechanical balance. Succesful implementation of qNMR requires certain attention to detail. "qNMR - the handbook" is intended to be a guide for analysts to help understand the fundamental principles of NMR and the significant points relating to its implementation for quantitation. Regulatory considerations of qNMR adoption are explained. NMR fundamentals are explained to provide understanding. Together with many useful examples, the book is a compelling addition to the laboratory's reference library, providing all the tools that any practitioner should know to successfully implement qNMR.
The authors are qNMR pioneers and come from a variety of backgrounds including business, government and academia.
Produzieren Sie schon oder kalibrieren Sie noch? – Online-NMR-Spektrometer als Smarte Feldgeräte
(2018)
Der Vortrag zeigt allgemeine Anforderungen an "smarte Feldgeräte" und deren Entwicklung in den vergangenen Jahren. Am Beispiel eines smarten Online-NMR-Sensors, der in einem EU-Projekt von der BAM entwickelt wurde, wird die Umsetzung der Anforderung aufgezeigt. Schließlich werden weitere Technologieanforderungen und Lösungsansätze vorgestellt.
Im Zuge der Digitalisierung der Prozessindustrie werden zunehmend modellbasiere Echtzeitoptimierungsverfahren eingesetzt, sog. „Advanced Process Control“. Mithilfe der sogenannten Modifier-Adaptation ist eine iterative Betriebspunktoptimierung auch mit ungenauen Modellen möglich, sofern zuverlässige Prozessdaten zur Verfügung stehen. Am Beispiel eines smarten Online-NMR-Sensors, der in einem EU-Projekt von der BAM entwickelt wurde, konnte das Konzept in einer modularen Produktionsanlage zur Herstellung eines pharmazeutischen Wirkstoffs erfolgreich getestet werden.
Online-NMR- und -Raman-Spektroskopie – Kombination von PAT-Methoden für mehr Prozessverständnis
(2020)
Die Verbesserung des Prozessverständnisses ist eine notwendige Voraussetzung für die Anwendung von Konzepten wie „Industrial Internet of Things (IIoT)“ oder „Industrie 4.0“. Online-Methoden der Prozessanalytik (PAT) ermöglichen die direkte Verfolgung der ablaufenden Reaktionen innerhalb chemischer und pharmazeutischer Produktionsschritte und können somit einen wichtigen Beitrag für die Entwicklung von neuen, innovativen Prozessführungsstrategien liefern.
Während die Online-Raman-Spektroskopie bereits erfolgreich im industriellen Umfeld zum Einsatz kommt, ist der Entwicklungsstand für die Online-Niederfeld-NMR-Spektroskopie bislang noch nicht ausreichend für einen routinemäßigen Einsatz innerhalb industrieller Produktionsumgebungen. Der hohe Informationsgehalt der NMR-Daten in Verbindung mit einem geringen Kalibrieraufwand macht die Methode sehr interessant für moderne Konzepte flexibler Produktionsanlagen für die Herstellung vieler unterschiedlicher Produkte in wechselnden Kampagnen. Speziell in modularen Produktionskonzepten kann so die Rüstzeit der Prozessanalytik deutlich verkürzt werden.
Am Beispiel der heterogen katalysierten Hydrogenierung von 2-Butin-1,4-diol wird der Einsatz der Kombination von Online-NMR und -Raman-Spektroskopie demonstriert. Dieser Prozessschritt aus der industriellen Synthese des Lösungsmittels Tetrahydrofuran verläuft über ein Zwischenprodukt, welches konkurrierende Reaktionspfade aufweist. Die Kombination von Informationen aus der PAT, sowie von klassischen Prozessgrößen, wie Druck, Temperatur und Durchfluss in einem hoch automatisierten Aufbau erleichtert die Entwicklung und Erprobung von neuen Konzepten für die Prozessführung.
Die NMR-Spektroskopie ist eine der zentralen nicht-invasiven Analysenmethoden in der organischen Chemie und aus dem Laboralltag nicht mehr wegzudenken. Die direkte Proportionalität zwischen der Anzahl der Atomkerne im Messvolumen und der Signalfläche im Spektrum ist vergleichbar mit einem „Zählen der Kernspins“. Der Kalibrieraufwand für die Quantifizierung ist minimal und aus dem NMR-Spektrum sind Informationen zu Struktur und Identität zugänglich.
Online-NMR-Spektroskopie wird unter Zuhilfenahme von speziellen Durchflussproben-köpfen und Messzellen bereits seit Jahrzehnten erforscht und eingesetzt. Durch die hohen Anforderungen an die Aufstellung und den Betrieb von klassischen Hochfeld-NMR-Spektrometern konnte sie allerdings nie den Sprung vom Labor in den Prozess schaffen. Dies hat sich durch die Entwicklungen im Bereich mobil einsetzbarer Benchtop-NMR-Spektrometer grundlegend verändert. Geringere Investitions- und Betriebskosten, sowie die Robustheit und einfache Bedienbarkeit dieser Systeme sind entscheidende Faktoren. Damit rückt eine Anwendung als Online-PAT-Methode technisch in greifbare Nähe. Designierte Prozess-NMR-Spektrometer sucht man heutzutage allerdings auf dem Markt meist noch vergebens.
Die Reaktionsverfolgung im Labormaßstab konnte an unterschiedlichen Systemen erfolgreich demonstriert werden. Dabei zeigten sich allerdings Limitierungen in Hinblick auf einen möglichen Prozesseinsatz. Insbesondere die Temperaturempfindlichkeit der Magnetsysteme ist hier zu nennen. Anhand eines Prototyps wurde eine aktive Temperaturisolation mittels temperierter Luftströmungen entwickelt und erprobt.
Eine Feldintegration von Laborsystemen geht sowohl technisch als auch regulatorisch mit zahlreichen Herausforderungen einher. Das raue Umfeld von Produktionsanlagen, sowie die Anforderungen an Explosionsschutz erfordern eine zugelassene Einhausung. Auf Basis von Erfahrungen mit einem ersten Prototyp erfolgte die gemeinsame Entwicklung eines möglichst flexibel einsetzbaren Analysenschranks. Neben der Hardware-Integration sind die automatisierte modellbasierte Auswertung der NMR-Spektren, sowie die Einbindung in die Prozessleittechnik essenziell für einen zuverlässigen Betrieb als PAT-Analysengerät.
Improvement in deep process understanding is a mandatory prerequisite for the application of modern concepts like Industrial Internet of Things (IIoT) or “Industrie 4.0”. The direct hyphenation with online methods of process analytical technology (PAT) allows profound insights into the actual reactions within chemical and pharmaceutical production steps and provides necessary information for associated advanced control strategies.
While the industrial application of online Raman spectroscopy has already been successfully demonstrated, low-field NMR spectroscopy is not yet adequately developed as a robust online method for use in process industry. The high information content combined with the low calibration effort makes NMR spectroscopy a highly promising method for modern process automation with a high flexibility due to short set-up times and novel calibration concepts. This is a major advantage compared to other analytical methods, especially regarding multi-purpose plant strategies, as well as processes suffering from fluctuating quality of raw materials.
The work presented here focusses on the heterogeneous catalyzed hydrogenation of 2-butyne-1,4-diol, as a step of the synthesis of industrially important solvent tetrahydrofuran. This reaction is proceeding via an intermediate product and suffers from competitive reaction paths. In this application, online NMR and Raman spectroscopy were combined with data from classical process sensors, e.g., pressure, temperature, and flow transducers in a highly automated setup for the development of innovative control concepts.
There is a need within the NMR community to progress forward in exploring new facets in which we can use analytical techniques to advance our understanding of various systems. One aspect the NMR community hasn’t fully encompassed is the validation process, which also involves setting reference standards, establishing a common language that directly relates to NMR, communication relating to validation, and much more.
This workshopcontribution starts with an overview on international metrology for qNMR spectroscopy. Since NMR is completely described by mathematical equations, the measurement unceartainty can directly be dreived from formula. Examples are presented. These are differentiated between type A and B evaluations. Finally the Expanded Unceartainty is defined. Since the user needs a risk-based unceartainty assessment, different "leagues" for routine, advanced, and high level needs are proposed to make clear, that no all sources of uncertainty have to be taken in considerention at practical levels.
The departure from the current automation landscape to next generation automation concepts for the process industry has already begun. Smart functions of sensors simplify their use and enable plug-and-play integration, even though they may appear to be more complex at first sight. Monitoring specific information (i.e., “chemical” such as physico-chemical properties, chemical reactions, etc.) is the key to “chemical” process control.
The talk introduces a smart online NMR sensor module provided in an explosion proof housing as example. This sensor was developed for an intensified industrial process (pharmaceutical lithiation reaction step) funded by the EU’s Horizon 2020 research and innovation programme (www.consens-spire.eu). Due to NMR spectroscopy as an “absolute analytical comparison method”, independent of the matrix, it runs with extremely short set-up times in combination with “modular” spectral models. Such models can simply be built upon pure component NMR spectra within a few hours (i.e., assignment of the NMR signals to the components) instead of tedious calibrations runs.
The talk also generally covers current aspects of high-field and low-field online NMR spectroscopy for reaction monitoring and process control.
Chemical companies must find new paths to stay productive in a rapidly changing environment. One of these is the potential of digital technologies. Flexible and modular chemical plants can produce various high-quality products using multi-purpose equipment with short downtimes between campaigns and reduce time to market. Process safety is improved due to smaller amounts processed and the abilities of efficient heat-transfer allow for otherwise difficult-to-produce compounds.
To exploit these advantages, a fully automated process control along with real-time quality control is mandatory and should be based on “chemical” information. The advances of a fully automated NMR analyzer were demonstrated, using a given pharmaceutical reaction step operated within a modular pilot plant. A commercially available benchtop NMR spectrometer was integrated to the requirements of an automated chemical production environment such as explosion safety, field communication, and robust data evaluation. Obtained results were used for direct loop advanced process control and real-time optimization of the process.
NMR spectroscopy appeared as excellent online analytical tool and allowed a modular data analysis approach, which even served as reliable reference method for further PAT applications. Using the available datasets, a second data analysis approach based on artificial neural networks (ANN) was evaluated.Therefore, amount of data was augmented to be sufficient for training. The results show comparable performance, while improving the calculation time tremendously. In future, such fully integrated and interconnecting “smart” systems and processes can increase the efficiency of the production of specialty chemicals and pharmaceuticals.
Global demand for pigments is expected to increase to around 13.4 million tons per year by 2030, representing approximately $22 billion – 5% of which will be azo pigments as the main organic pigments. Azo pigments, i.e., especially the azo bridge required for them, are almost exclusively produced in technology by the so-called azo coupling reaction in large-scale approaches on the scale of 50–100 cubic meters. Among other things, one of the reasons why large-scale production itself is still essentially discontinuous is the lack of analytical methods for a continuous azo pigment synthesis.
In an azo coupling process, an aromatic amine is diazotized to form the diazonium compound (diazo compound), which then reacts with a coupling component, i.e., "couples". A slight excess of the coupler is desired for the end of the reaction, which stoichiometrically consumes the remaining (harmful) diazo compound, i.e., below 500 ppm. This is carried out with a manual spotting reaction of a fast-binding coupler.
To carry out continuous azo coupling, care must be taken to ensure that the actual coupling reaction proceeds as rapidly as possible, i.e., preferentially, compared with possible side reactions. Reliable measuring methods are thus a decisive criterion for the useful performance of a continuous azo pigment synthesis. An analytical online procedure is difficult due to several limitations at once: solid dispersions extremely limit the range of reproducible analytical procedures. In addition, the appearance of a signal (e.g., excess of a starting material) in the ppm range must be reliably detected against the matrix background.
The back-titration method and benchtop NMR spectroscopy were used to overcome these problems. We present a feasibility study for a suitable process analysis with benchtop NMR spectroscopy on disperse systems for equimolar dosing of the "diazotization" or "coupling" sub-step in azo coupling, which was carried out in cooperation with a chemical company. This could replace tedious manual spotting
ISO/IEC 17025 is the worldwide quality standard for testing and calibration laboratories. It is the basis for accreditation by an accreditation body. The current version was published in 2018.
Implementing ISO/IEC 17025 as part of laboratory quality initiatives offers both laboratory and business benefits, such as expanding the potential customer base for testing and/or calibration, increasing the reputation and image of the laboratory at national and international level, continuous improvement of the data quality and the effectiveness of the laboratory or creation of a good basis for most other quality systems in the laboratory sector, such as GxP. The main difference between a proper approach to analysis and a formal accreditation is shown in a targeted documentation, especially on the qualification of the personnel, the test equipment and the validation of the analytical methods.
Using quantitative NMR spectroscopy as an example, it is shown how accreditation can be carried out and what documentation is required. In our case, we have described the procedure in an SOP ("Determination of the quantitative composition of simple mixtures of structurally known compounds with 1H-NMR spectroscopy") and supported it with a modular system of organizational and equipment SOPs. The special feature is that the accredited method is independent for the choice of the analyte and the matrix and therefore it is possible to operate with a single validated method. In our case, we have proposed three quality levels ("leagues") with different levels of analytical effort, which differ in their measurement uncertainty, in order to simplify the workflow and analysis design.
Due to its advantages of being a direct comparison method, quantitative NMR spectroscopy (qNMR) becomes more and more popular in industry. While conventional high-field NMR systems are often associated with high investment and operational costs, the upcoming market of permanent-magnet based benchtop NMR systems show a considerable option for a lot of applications. The mobility of these systems allows to bring them more closely to the real production environment, e.g. for at-line quality control.
In this work we present an interlaboratory comparison study investigating the qNMR performance of state-of-the-art benchtop NMR spectrometers. Therefore, BAM prepared two samples of a mixture of NMR reference standards tetramethylbenzene (TMB) and tetrachloronitrobenzene (TCNB) at concentration levels of 200 mM and 10 mM. These “ready-to-use” samples were sent to participant laboratories, which performed analysis on their benchtop NMR equipment of different vendors and fields from 43 to 80 MHz. Raw data was reported back and further investigated by using different data analysis methods at BAM.
After this very first qNMR comparison study of benchtop NMR spectrometers show promising results, following studies are planned to cover more parts of the qNMR process, e.g. sample preparation and weighing, but also data analysis, as commonly done in similar studies for high-field NMR spectroscopy in industry and metrology.
Due to its advantages of being a direct comparison method, quantitative NMR spectroscopy (qNMR) becomes more and more popular in industry. While conventional high-field NMR systems are often associated with high investment and operational costs, the upcoming market of permanent-magnet based benchtop NMR systems show a considerable option for a lot of applications. The mobility of these systems allows to bring them more closely to the real production environment, e.g. for at-line quality control.
In this work we present an interlaboratory comparison study investigating the qNMR performance of state-of-the-art benchtop NMR spectrometers. Therefore, BAM prepared two samples of a mixture of NMR reference standards tetramethylbenzene (TMB) and tetrachloronitrobenzene (TCNB) at concentration levels of 200 mM and 10 mM. These “ready-to-use” samples were sent to participant laboratories, which performed analysis on their benchtop NMR equipment of different vendors and fields from 43 to 80 MHz. Raw data was reported back and further investigated by using different data analysis methods at BAM.
After this very first qNMR comparison study of benchtop NMR spectrometers show promising results, following studies are planned to cover more parts of the qNMR process, e.g. sample preparation and weighing, but also data analysis, as commonly done in similar studies for high-field NMR spectroscopy in industry and metrology.
Due to its advantages of being a direct comparison method, quantitative NMR spectroscopy (qNMR) becomes more and more popular in industry. While conventional high-field NMR systems are often associated with high investment and operational costs, the upcoming market of permanent-magnet based benchtop NMR systems show a considerable option for a lot of applications. The mobility of these systems allows to bring them more closely to the real production environment, e.g. for at-line quality control.
In this work we present an interlaboratory comparison study investigating the qNMR performance of state-of-the-art benchtop NMR spectrometers. Therefore, BAM prepared two samples of a mixture of NMR reference standards tetramethylbenzene (TMB) and tetrachloronitrobenzene (TCNB) at concentration levels of 200 mM and 10 mM. These “ready-to-use” samples were sent to participant laboratories, which performed analysis on their benchtop NMR equipment of different vendors and fields from 43 to 80 MHz. Raw data was reported back and further investigated by using different data analysis methods at BAM.
After this very first qNMR comparison study of benchtop NMR spectrometers show promising results, following studies are planned to cover more parts of the qNMR process, e.g. sample preparation and weighing, but also data analysis, as commonly done in similar studies for high-field NMR spectroscopy in industry and metrology.
Industrielle Anwendungen der Niederfeld-NMR-Spektroskopie für die Qualitätskontrolle von Silanen
(2021)
Aus der Kombination verschiedener Silane als Ausgangsstoffe und als Produkt der Hydrolyse durch eine Reihe von Alkoholen oder Wasser ergibt sich eine Palette von hunderten technischen Produkten für verschiedenste Anwendungen. Funktionelle Trialkoxysilane haben sich in den vergangenen Jahren als vielseitig einsetzbare Organosilane erwiesen. Die Einsatzgebiete erstrecken sich vom Witterungsschutz von Bauwerken über haftvermittelnde Eigenschaften in der Glasfaserindustrie, bei Dicht- und Klebstoffen, in Farben und Lacken bis hin zur Modifizierung von polymeren Werkstoffen.
Kommerzielle Benchtop-NMR-Spektrometer haben das Potential auch im Bereich der Silanchemie als Online-Methode zur Reaktionsüberwachung und für die Qualitätskontrolle eingesetzt zu werden. Interessante NMR-Kerne für die oben genanntem Produkte sind 1H und 29Si. In einer gemeinsamen Forschungskooperation zwischen EVONIK und BAM wurde anhand verschiedener Fallstudien die Anwendbarkeit der Niederfeld-NMR-Spektroskopie zur chemischen Analyse von Silanen evaluiert. Die Möglichkeiten der Konzentrationsbestimmung erweitern die Anwendungsgebiete, in denen bestehende Technologien wie z. B. NIR, Raman, UV/VIS, etc. mangels Referenzdaten nicht quantitativ eingesetzt werden können.
Eine Fallstudie setzte dazu an, den Hydrolyse- und Kondensationsverlauf mit geeigneter Online-NMR-Analytik zu beobachten, den Reaktionsfortschritt der Hydrolyse und Kondensation auf dieser Basis besser zu verstehen und zu optimieren. Zu diesem Zweck werden durch Zugabe von Wasser zunächst die Alkoxysubstituenten eines Trialkoxysilans hydrolysiert und entsprechende Silanole gebildet. Diese können dann über eine SiOH-Funktion an den zu modifizierenden Werkstoff anbinden und über weitere Silanolgruppen unter Ausbildung von Siloxaneinheiten vernetzen. In einer weiteren Fallstudie wurde die Kinetik der Aufspaltungen einer cyclischen Silanverbindung untersucht. Die Online-NMR-Analytik kam hierbei sowohl im Labor als auch in der industriellen Produktionsanlage zum Einsatz. Hierfür wurde eine automatisierte Einhausung verwendet, welche den Einsatz eines kommerziellen NMR-Spektrometers in explosionsgeschützten Bereichen ermöglicht.
Aus der Kombination verschiedener Silane als Ausgangsstoffe und als Produkt der Hydrolyse durch eine Reihe von Alkoholen oder Wasser ergibt sich eine Palette von hunderten technischen Produkten für verschiedenste Anwendungen. Funktionelle Trialkoxysilane haben sich in den vergangenen Jahren als vielseitig einsetzbare Organosilane erwiesen. Die Einsatzgebiete erstrecken sich vom Witterungsschutz von Bauwerken über haftvermittelnde Eigenschaften in der Glasfaserindustrie, bei Dicht- und Klebstoffen, in Farben und Lacken bis hin zur Modifizierung von polymeren Werkstoffen.
Kommerzielle Benchtop-NMR-Spektrometer haben das Potential auch im Bereich der Silanchemie als Online-Methode zur Reaktionsüberwachung und für die Qualitätskontrolle eingesetzt zu werden. Interessante NMR-Kerne für die oben genanntem Produkte sind 1H und 29Si. In einer gemeinsamen Forschungs-kooperation zwischen EVONIK und BAM wurde anhand verschiedener Fallstudien die Anwendbarkeit der Niederfeld-NMR-Spektroskopie zur chemischen Analyse von Silanen evaluiert. Die Möglichkeiten der Konzentrationsbestimmung erweitern die Anwendungsgebiete, in denen bestehende Technologien wie z. B. NIR, Raman, UV/VIS, etc. mangels Referenzdaten nicht quantitativ eingesetzt werden können.
Eine Fallstudie setzte dazu an, den Hydrolyse- und Kondensationsverlauf mit geeigneter Online-NMR-Analytik zu beobachten, den Reaktionsfortschritt der Hydrolyse und Kondensation auf dieser Basis besser zu verstehen und zu optimieren. Zu diesem Zweck werden durch Zugabe von Wasser zunächst die Alkoxysubstituenten eines Trialkoxysilans hydrolysiert und entsprechende Silanole gebildet. Diese können dann über eine SiOH-Funktion an den zu modifizierenden Werkstoff anbinden und über weitere Silanolgruppen unter Ausbildung von Siloxaneinheiten vernetzen. In einer weiteren Fallstudie wurde die Kinetik der Aufspaltungen einer cyclischen Silanverbindung untersucht. Die Online-NMR-Analytik kam hierbei sowohl im Labor als auch in der industriellen Produktionsanlage zum Einsatz. Hierfür wurde eine automatisierte Einhausung verwendet, welche den Einsatz eines kommerziellen NMR-Spektrometers in explosionsgeschützten Bereichen ermöglicht.