Produzieren Sie schon oder kalibrieren Sie noch? – Online-NMR-Spektrometer als Smarte Feldgeräte
(2018)
Der Vortrag zeigt allgemeine Anforderungen an "smarte Feldgeräte" und deren Entwicklung in den vergangenen Jahren. Am Beispiel eines smarten Online-NMR-Sensors, der in einem EU-Projekt von der BAM entwickelt wurde, wird die Umsetzung der Anforderung aufgezeigt. Schließlich werden weitere Technologieanforderungen und Lösungsansätze vorgestellt.
Plastic pollution is the biggest environmental concern of our time. Breakdown products like micro- and nanoplastics inevitably enter the food chain and pose unprecedented health risks. In this scenario, bio-based and biodegradable plastic alternatives have been given a momentum aiming to bridge a transition towards a more sustainable future. Polyhydroxyalkanoates (PHAs) are one of the few thermoplastic polymers synthesized 100 % via biotechnological routes which fully biodegrade in common natural environments. Poly(hydroxybutyrate-cohydroxyhexanoate) [P(HB-co-HHx)] is a PHA copolymer with great potential for the commodity polymers industry, as its mechanical properties can be tailored through fine-tuning of its molar HHx content. We have recently developed a strategy that enables for reliable tailoring of the monomer content of P(HB-co-HHx).
Nevertheless, there is often a lack of comprehensive investigation of the material properties of PHAs to evaluate whether they actually mimic the functionalities of conventional plastics. We present a detailed study of P(HB-co-HHx) copolymers with low to moderate hydroxyhexanoate content to understand how the HHx monomer content influences the thermal and mechanical properties and to link those to their abiotic degradation. By increasing the HHx fractions in the range of 2 – 14 mol%, we impart an extension of the processing window and application range as the melting temperature (Tm) and glass temperature (Tg) of the copolymers decrease from Tm 165 ◦C to 126 ◦C, Tg 4 ◦C to − 5.9 ◦C, accompanied by reduced crystallinity from 54 % to 20 %.
Elongation at break was increased from 5.7 % up to 703 % at 14 mol% HHx content, confirming that the range examined was sufficiently large to obtain ductile and brittle copolymers, while tensile strength was maintained throughout the studied range. Finally, accelerated abiotic degradation was shown to be slowed down with an increasing HHx fraction decreasing from 70 % to 55 % in 12 h.
The application of compact NMR instruments to continuously flowing samples at higher temperatures or exothermically reacting mixtures is limited by the temperature sensitivity of permanent magnets. Typically, such temperature effects directly influence the achievable magnetic field homogeneity and hence measurement quality. The internal-temperature control loop of the magnet and instruments is typically not designed for compensation of such effects. Passive insulation is restricted by the small dimensions within the magnet borehole.
Here, we present a design approach for active heat shielding with the aim of variable temperature control of NMR samples for benchtop NMR instruments using a compressed airstream which is variable in flow and temperature. Based on the system identification and surface temperature measurements using an optical thermo-graphy setup, a model predictive control was set up to minimize any disturbance effect on the permanent magnet from the probe or sample temperature.
This methodology will facilitate the application of variable-temperature shielding and, therefore, potentially extend the application of compact NMR instruments to flowing samples at higher temperature than the permanent magnet.
Calibration-Free Chemical Process and Quality Control Units as Enablers for Modular Production
(2021)
Modular chemical production is a tangible translation of the digital transformation of the process industry for specialty chemicals. In particular, it enables the speeding-up of process development and thus a quicker time to market by flexibly connecting and orchestrating standardised physical modules and bringing them to life (i.e., parameterising them) with digitally accumulated process knowledge.
We focus on the specific challenges of chemical process and quality control, which in its current form is not well suited for modular production and provide possible approaches and examples of the change towards direct analytical methods, analytical model transfer or machine-supported processes.
Monitoring specific chemical properties is the key to chemical process control. Today, mainly optical online methods are applied, which require time- and cost-intensive calibration effort. NMR spectroscopy, with its advantage being a direct comparison method without need for calibration, has a high potential for closed-loop process control while exhibiting short set-up times. Compact NMR instruments make NMR spectroscopy accessible in industrial and rough environments for process monitoring and advanced process control strategies.
We present a fully automated data analysis approach which is completely based on physically motivated spectral models as first principles information (Indirect Hard Modelling – IHM) and applied it to a given pharmaceutical lithiation reaction in the framework of the European Union’s Horizon 2020 project CONSENS. Online low-field NMR (LF NMR) data was analysed by IHM with low calibration effort, compared to a multivariate PLS-R (Partial Least Squares Regression) approach, and both validated using online high-field NMR (HF NMR) spectroscopy.
Die NMR-Spektroskopie stellt heutzutage eine der wichtigsten Analysenmethoden in der organischen Chemie dar. Während der Großteil aller Untersuchungen qualitativ mit dem Ziel der Stoffidentifikation und Strukturaufklärung erfolgt, erlangt die quantitative NMR-Spektroskopie (qNMR) zunehmend an Bedeutung in Forschung und Industrie. Der entscheidende Vorteil gegenüber anderen Analysenmethoden liegt in der direkten Proportionalität der Signalfläche zur Anzahl Kernspins im Messvolumen. Dies erlaubt eine kalibrationsfreie Relativquantifizierung. Zur Absolutquantifizierung reicht die Zugabe einer definierten Menge eines vom Analyten unabhängigen NMR-Standards aus.
Trotz dieser Vorteile findet sich die qNMR bislang nur vereinzelt in Normen und Standardverfahren wieder. Zahlreiche Ringversuche in Metrologie und Industrie demonstrieren die Leistungsfähigkeit moderner NMR-Spektrometer und stärken das Vertrauen in die Methode. Die Entwicklung von Validierungskonzepten, sowie die kommerzielle Verfügbarkeit geeigneter zertifizierter Referenzmaterialien erleichtern die Anwendung, insbesondere im zumeist stark regulierten industriellen Umfeld.
Neben etablierten Hochfeld-NMR Spektrometern hat sich in den letzten Jahren ein stark wachsender Markt für kompakte Benchtop-NMR Geräte auf Permanentmagnetbasis entwickelt. Die geringeren Anschaffungs- und Betriebskosten, sowie die einfache Bedienbarkeit erlaubt es auch kleineren Unternehmen in diese Analysenmethode einzusteigen. Weiterhin besteht die Möglichkeit diese mobilen Systeme näher an die Produktion zu bringen, welches von der klassischen Qualitätskontrolle bis hin zur Online-Prozesskontrolle als vollautomatisierter Analysator reicht. Die geringere Feldstärke der Systeme erfordert hier oft den Einsatz modellbasierter Ansätze zur Spektrenauswertung (z.B. Indirect Hard Modeling).
Dieser Beitrag gibt eine Übersicht über aktuelle Anwendungen und Entwicklungen der qNMR von der universellen, hochgenauen Labormethode bis hin zur robusten Anwendung als Online-Analysator im Feld.
There is a need within the NMR community to progress forward in exploring new facets in which we can use analytical techniques to advance our understanding of various systems. One aspect the NMR community hasn’t fully encompassed is the validation process, which also involves setting reference standards, establishing a common language that directly relates to NMR, communication relating to validation, and much more.
This workshopcontribution starts with an overview on international metrology for qNMR spectroscopy. Since NMR is completely described by mathematical equations, the measurement unceartainty can directly be dreived from formula. Examples are presented. These are differentiated between type A and B evaluations. Finally the Expanded Unceartainty is defined. Since the user needs a risk-based unceartainty assessment, different "leagues" for routine, advanced, and high level needs are proposed to make clear, that no all sources of uncertainty have to be taken in considerention at practical levels.
Due to its advantages of being a direct comparison method, quantitative NMR spectroscopy (qNMR) becomes more and more popular in industry. While conventional high-field NMR systems are often associated with high investment and operational costs, the upcoming market of permanent-magnet based benchtop NMR systems show a considerable option for a lot of applications. The mobility of these systems allows to bring them more closely to the real production environment, e.g. for at-line quality control.
In this work we present an interlaboratory comparison study investigating the qNMR performance of state-of-the-art benchtop NMR spectrometers. Therefore, BAM prepared two samples of a mixture of NMR reference standards tetramethylbenzene (TMB) and tetrachloronitrobenzene (TCNB) at concentration levels of 200 mM and 10 mM. These “ready-to-use” samples were sent to participant laboratories, which performed analysis on their benchtop NMR equipment of different vendors and fields from 43 to 80 MHz. Raw data was reported back and further investigated by using different data analysis methods at BAM.
After this very first qNMR comparison study of benchtop NMR spectrometers show promising results, following studies are planned to cover more parts of the qNMR process, e.g. sample preparation and weighing, but also data analysis, as commonly done in similar studies for high-field NMR spectroscopy in industry and metrology.
Due to its advantages of being a direct comparison method, quantitative NMR spectroscopy (qNMR) becomes more and more popular in industry. While conventional high-field NMR systems are often associated with high investment and operational costs, the upcoming market of permanent-magnet based benchtop NMR systems show a considerable option for a lot of applications. The mobility of these systems allows to bring them more closely to the real production environment, e.g. for at-line quality control.
In this work we present an interlaboratory comparison study investigating the qNMR performance of state-of-the-art benchtop NMR spectrometers. Therefore, BAM prepared two samples of a mixture of NMR reference standards tetramethylbenzene (TMB) and tetrachloronitrobenzene (TCNB) at concentration levels of 200 mM and 10 mM. These “ready-to-use” samples were sent to participant laboratories, which performed analysis on their benchtop NMR equipment of different vendors and fields from 43 to 80 MHz. Raw data was reported back and further investigated by using different data analysis methods at BAM.
After this very first qNMR comparison study of benchtop NMR spectrometers show promising results, following studies are planned to cover more parts of the qNMR process, e.g. sample preparation and weighing, but also data analysis, as commonly done in similar studies for high-field NMR spectroscopy in industry and metrology.
Chemische Prozesskontrolle für modulare Produktionskonzepte mit direkten analytischen Methoden
(2020)
Flexible Produktionskonzepte werden oft mit modularen, schnell rekonfigurierbaren Einheiten verbunden. Um die Vorteile dieser flexiblen Produktion tatsächlich nutzen zu können ist neben einer geeigneten Infrastruktur auch eine ebenso anpassungsfähige Qualitätskontrolle unverzichtbar. Analysengeräte und Sensoren werden heutzutage immer kompakter und kostengünstiger in ihrer Anschaffung. Dadurch wird es möglich eine große Anzahl von Messstellen innerhalb einer Anlage zu realisieren. Die erfassten Datenmengen erfordern zuverlässige und robuste Auswerteverfahren, die sich in das Gesamtkonzept der Automatisierung einer derart flexiblen Produktion nahtlos integrieren. Referenzdaten aus direkten Methoden wie der NMR-Spektroskopie können hier einen wertvollen Beitrag zur Generierung von Prozesswissen für die Entwicklung neuer Modelle für kostengünstigere Analysenverfahren leisten.
Modular plants using intensified continuous processes represent an appealing concept to produce pharmaceuticals. It can improve quality, safety, sustainability, and profitability compared to batch processes, and it enables plug-and-produce reconfiguration for fast product changes. To facilitate this flexibility by real-time quality control, we developed a solution that can be adapted quickly to new processes and includes a compact Nuclear Magnetic Resonance (NMR) spectrometer for online quality monitoring as well as a new model-based control approach. The NMR sensor is a benchtop device enhanced to the requirements of automated chemical production including ro-bust evaluation of sensor data.
Here, we present alternatives for the quantitative determination of the analytes using modular, physically motivated models. These models can be adapted to new substances solely by the use of their corresponding pure component spectra, which can either be derived from experimental spectra as well as from quantum mechanical models or NMR predictors. Modular means that spec-tral models can simply be exchanged together with alternate reagents and products. Beyond that, we comprehensively calibrated an NIR spectrometer based on online NMR process data for the first time within an industrial plant. The integrated solution was developed for a metal organic reac-tion running on a commercial-scale modular pilot plant and it was tested under industrial conditions.
A big problem with the chemistry literature is that it is not standardized with respect to precise operational parameters, and real time corrections are hard to make without expert knowledge. This lack of context means difficult reproducibility because many steps are ambiguous, and hence depend on tacit knowledge. Here we present the integration of online NMR into an automated chemical synthesis machine (CSM aka. “Chemputer” which is capable of small-molecule synthesis using a universal programming language) to allow automated analysis and adjustment of reactions on the fly. The system was validated and benchmarked by using Grignard reactions which were chosen due to their importance in synthesis. The system was monitored in real time using online-NMR, and spectra were measured continuously during the reactions. This shows that the synthesis being done in the Chemputer can be dynamically controlled in response to feedback optimizing the reaction conditions according to the user requirements.
Im Zuge der Digitalisierung der Prozessindustrie werden zunehmend modellbasiere Echtzeitoptimierungsverfahren eingesetzt, sog. „Advanced Process Control“. Mithilfe der sogenannten Modifier-Adaptation ist eine iterative Betriebspunktoptimierung auch mit ungenauen Modellen möglich, sofern zuverlässige Prozessdaten zur Verfügung stehen. Am Beispiel eines smarten Online-NMR-Sensors, der in einem EU-Projekt von der BAM entwickelt wurde, konnte das Konzept in einer modularen Produktionsanlage zur Herstellung eines pharmazeutischen Wirkstoffs erfolgreich getestet werden.
Die NMR-Spektroskopie ist eine der zentralen nicht-invasiven Analysenmethoden in der organischen Chemie und aus dem Laboralltag nicht mehr wegzudenken. Die direkte Proportionalität zwischen der Anzahl der Atomkerne im Messvolumen und der Signalfläche im Spektrum ist vergleichbar mit einem „Zählen der Kernspins“. Der Kalibrieraufwand für die Quantifizierung ist minimal und aus dem NMR-Spektrum sind Informationen zu Struktur und Identität zugänglich.
Online-NMR-Spektroskopie wird unter Zuhilfenahme von speziellen Durchflussproben-köpfen und Messzellen bereits seit Jahrzehnten erforscht und eingesetzt. Durch die hohen Anforderungen an die Aufstellung und den Betrieb von klassischen Hochfeld-NMR-Spektrometern konnte sie allerdings nie den Sprung vom Labor in den Prozess schaffen. Dies hat sich durch die Entwicklungen im Bereich mobil einsetzbarer Benchtop-NMR-Spektrometer grundlegend verändert. Geringere Investitions- und Betriebskosten, sowie die Robustheit und einfache Bedienbarkeit dieser Systeme sind entscheidende Faktoren. Damit rückt eine Anwendung als Online-PAT-Methode technisch in greifbare Nähe. Designierte Prozess-NMR-Spektrometer sucht man heutzutage allerdings auf dem Markt meist noch vergebens.
Die Reaktionsverfolgung im Labormaßstab konnte an unterschiedlichen Systemen erfolgreich demonstriert werden. Dabei zeigten sich allerdings Limitierungen in Hinblick auf einen möglichen Prozesseinsatz. Insbesondere die Temperaturempfindlichkeit der Magnetsysteme ist hier zu nennen. Anhand eines Prototyps wurde eine aktive Temperaturisolation mittels temperierter Luftströmungen entwickelt und erprobt.
Eine Feldintegration von Laborsystemen geht sowohl technisch als auch regulatorisch mit zahlreichen Herausforderungen einher. Das raue Umfeld von Produktionsanlagen, sowie die Anforderungen an Explosionsschutz erfordern eine zugelassene Einhausung. Auf Basis von Erfahrungen mit einem ersten Prototyp erfolgte die gemeinsame Entwicklung eines möglichst flexibel einsetzbaren Analysenschranks. Neben der Hardware-Integration sind die automatisierte modellbasierte Auswertung der NMR-Spektren, sowie die Einbindung in die Prozessleittechnik essenziell für einen zuverlässigen Betrieb als PAT-Analysengerät.
qNMR - The handbook
(2023)
Quantitative NMR (qNMR) has been around for a long time, but also has great potential to solve future problems in any quantitative analysis. As a primary method, it differs fundamentally from chromatographic methods: it is better described as a quantum mechanical balance. Succesful implementation of qNMR requires certain attention to detail. "qNMR - the handbook" is intended to be a guide for analysts to help understand the fundamental principles of NMR and the significant points relating to its implementation for quantitation. Regulatory considerations of qNMR adoption are explained. NMR fundamentals are explained to provide understanding. Together with many useful examples, the book is a compelling addition to the laboratory's reference library, providing all the tools that any practitioner should know to successfully implement qNMR.
The authors are qNMR pioneers and come from a variety of backgrounds including business, government and academia.
Compact NMR Spectroscopy in the field: A Versatile Tool for Automated Continuous-Flow Production
(2022)
Chemical companies must find new paths to successfully survive in a changing environment. The potential of digital technologies belongs to these. Flexible and modular chemical plants can produce various high-quality products using multi-purpose equipment with short downtimes between campaigns and reduce time to market for new products. Intensified continuous production plants allow for difficult to produce compounds.
Therefore, fully automated “chemical” process control along with real-time quality control are prerequisites to such concepts and thus should be based on “chemical” information. A commercially available benchtop NMR spectrometer was integrated to the full requirements of an automated chemical production environment such as, e.g., explosion safety, field communication, and robust evaluation of sensor data. It was thereof used for direct loop advanced process control and real-time optimization of the process. Field studies in modular and conventional production plant setups show promising results gaining process knowledge for further optimization. NMR appeared as preeminent online analytical method and allow using a modular data analysis tool, which even served as reliable reference method for further PAT applications (e.g. NIR spectroscopy). In the future, such fully integrated and intelligently interconnecting “smart” systems and processes can speed up the high-quality production of specialty chemicals and pharmaceuticals.
The application of compact NMR instruments to hot flowing samples or exothermically reacting mixtures is limited by the temperature sensitivity of permanent magnets. Typically, such temperature effects directly influence the achievable magnetic field homogeneity and hence measurement quality. The internal-temperature control loop of the magnet and instruments is not designed for such temperature compensation. Passive insulation is restricted by the small dimensions within the magnet borehole. Here, we present a design approach for active heat shielding with the aim of variable temperature control of NMR samples for benchtop NMR instruments using a compressed airstream which is variable in flow and temperature. Based on the system identification and surface temperature measurements through thermography, a model predictive control was set up to minimise any disturbance effect on the permanent magnet from the probe or sample temperature. This methodology will facilitate the application of variable-temperature shielding and, therefore, extend the application of compact NMR instruments to flowing sample temperatures that differ from the magnet temperature.
Compact NMR Spectroscopy in the field: A Versatile PAT Tool for Production of Specialty Chemicals
(2023)
Companies of chemical industry find themselves more often in a rapidly changing environment, e.g., due to variability of raw material quality or energy costs and efficiency. Process optimization and new process concepts become more and more important. For example, flexible and modular chemical plants can produce various high-quality products using multi-purpose equipment with short downtimes between campaigns and reduce time to market for new products. Intensified continuous production plants allow for difficult to produce compounds like exothermic reactions with high heat dissipation.
Highly automated “chemical” process control along with real-time quality control are prerequisites to such concepts and thus should be based on “chemical” information. A commercially available benchtop NMR spectrometer was integrated to the full requirements of an automated chemical production environment such as explosion safety, field communication, and robust evaluation of sensor data. Field studies in modular and conventional production plant setups show promising results gaining process knowledge for further optimization. NMR appeared as preeminent online analytical method and allow using a modular data analysis approach, which can even serve as reliable reference method for further calibration-dependent PAT applications (e.g., NIR or Raman spectroscopy).
Based on experiences from earlier field studies an improved NMR analyzer enclosure setup was developed and built, including the option of a secondary method (e.g., optical spectroscopy). Integrated control systems allow for a flexible implementation based on the available automation infrastructure at the chemical plant or pilot plant setup. In the future, fully integrated and intelligently interconnecting “smart” PAT systems and processes have the potential speed up the setup of production equipment for chemicals and pharmaceuticals and therefore help to reduce the time-to-market.
Global demand for pigments is expected to increase to around 13.4 million tons per year by 2030, representing approximately $22 billion – 5% of which will be azo pigments as the main organic pigments. Azo pigments, i.e., especially the azo bridge required for them, are almost exclusively produced in technology by the so-called azo coupling reaction in large-scale approaches on the scale of 50–100 cubic meters. Among other things, one of the reasons why large-scale production itself is still essentially discontinuous is the lack of analytical methods for a continuous azo pigment synthesis.
In an azo coupling process, an aromatic amine is diazotized to form the diazonium compound (diazo compound), which then reacts with a coupling component, i.e., "couples". A slight excess of the coupler is desired for the end of the reaction, which stoichiometrically consumes the remaining (harmful) diazo compound, i.e., below 500 ppm. This is carried out with a manual spotting reaction of a fast-binding coupler.
To carry out continuous azo coupling, care must be taken to ensure that the actual coupling reaction proceeds as rapidly as possible, i.e., preferentially, compared with possible side reactions. Reliable measuring methods are thus a decisive criterion for the useful performance of a continuous azo pigment synthesis. An analytical online procedure is difficult due to several limitations at once: solid dispersions extremely limit the range of reproducible analytical procedures. In addition, the appearance of a signal (e.g., excess of a starting material) in the ppm range must be reliably detected against the matrix background.
The back-titration method and benchtop NMR spectroscopy were used to overcome these problems. We present a feasibility study for a suitable process analysis with benchtop NMR spectroscopy on disperse systems for equimolar dosing of the "diazotization" or "coupling" sub-step in azo coupling, which was carried out in cooperation with a chemical company. This could replace tedious manual spotting
The application of compact NMR instruments to hot flowing samples or exothermically reacting mixtures is limited by the temperature sensitivity of permanent magnets. Typically, such temperature effects directly influence the achievable magnetic field homogeneity and hence measurement quality. The internal-temperature control loop of the magnet and instruments is not designed for such temperature compensation. Passive insulation is restricted by the small dimensions within the magnet borehole.
Here, we present a design approach for active heat shielding with the aim of variable temperature control of NMR samples for benchtop NMR instruments using a compressed airstream which is variable in flow and temperature. Based on the system identification and surface temperature measurements through thermography, a model predictive control was set up to minimise any disturbance effect on the permanent magnet from the probe or sample temperature.
This methodology will facilitate the application of variable-temperature shielding and, therefore, extend the application of compact NMR instruments to flowing sample temperatures that differ from the magnet temperature.