Online monitoring and process control requires fast and noninvasive analytical methods, which are able to monitor the concentration of reactants in multicomponent mixtures with parts-per-million resolution. Online NMR spectros-copy can meet these demands when flow probes are directly coupled to reactors, since this method features a high linearity between absolute signal area and sample concentration, which makes it an absolute analytical com-parison method being independent on the matrix. Due to improved magnet design and field shimming strategies portable and robust instruments have been introduced to the market by several manufacturers during the last few years. First studies with this technology showed promising results to monitor chemical reaction in the laboratory.
Ugly Spectra and Lousy interfaces – Challenges for Compact NMR Spectroscopy in Process Control
(2016)
With the introduction of advanced process analytical technology, the closeness of key process variables to their limits can be directly controlled and the product can be classified or even released in real time.
Compact NMR instruments can make NMR spectroscopy accessible in industrial and harsh environments for process control.
Driven mostly by the search for chemical syntheses under biocompatible conditions, so called "click" chemistry rapidly became a growing field of research. The resulting simple one-pot reactions are so far only scarcely accompanied by an adequate optimization via comparably straightforward and robust analysis techniques possessing short set-up times. Here, we report on a fast and reliable calibration-free online NMR monitoring approach for technical mixtures. It combines a versatile fluidic system, continuous-flow measurement of 1H spectra with a time interval of 20 s per spectrum, and a robust, fully automated algorithm to interpret the obtained data. As a proof-of-concept, the thiol-ene coupling between N-boc cysteine methyl ester and allyl alcohol was conducted in a variety of non-deuterated solvents while its time-resolved behaviour was characterized with step tracer experiments. Overlapping signals in online spectra during thiol-ene coupling could be deconvoluted with a spectral model using indirect hard modeling and were subsequently converted to either molar ratios (using a calibrationfree approach) or absolute concentrations (using 1-point calibration). For various solvents the kinetic constant k for pseudo-first order reaction was estimated to be 3.9 h-1 at 25 °C. The obtained results were compared with direct integration of non-overlapping signals and showed good agreement with the implemented mass balance.
Compact nuclear magnetic resonance (NMR) instruments make NMR spectroscopy and relaxometry accessible in industrial and harsh environments for reaction and process control. An increasing number of applications are reported. Robust field integration of NMR systems have to face explosion protection or integration into process control systems with short set-up times. Tis paves the way for industrial automation in real process environments.
Automated data preparation and analysis are cornerstones for a breakthrough of NMR techniques for process control. Particularly, robust chemometrics as well as automated signal processing methods have to be (further) developed especially for NMR spectroscopy in process control. This becomes even more important for so called “smart sensors” providing the basis for the future project “Industrie 4.0”, and Industrial Internet of Things (IIoT), along with current requirements to process control, model based control, or soft sensing. Data analysis techniques are available but currently mostly used for off-line data analysis to detect the causes of variations in the product quality.
The talk presents current research activities towards process control with compact NMR and reflects “cultural differences” between the interdisciplinary parties involved.
The Royal Society of Chemistry NMR Discussion Group and Molecular Spectroscopy Group would like to invite you to the 2017 Spring Meeting, which will be held at GlaxoSmithKline (GSK), Stevenage. The theme for the meeting is “Low level detection and quantification by NMR” and different NMR technologies, including solution state NMR, solid state NMR and benchtop/low field NMR will be discussed. The presentations will cover a range of NMR related disciplines, including conventional low level detection and quantification, the use of cryoprobes, quantification of polymorphism using ssNMR and also methods for spectral simplification. Recent developments and applications of hyperpolarisation techniques, within both solution state and solid state NMR, will be presented in conjunction with the effect these sensitivity enhancements have with respect to quantification and limits of detection.
Smart functions of sensors simplify their use and enable plug-and-play, even though they are more complex. This is particularly important for, self-diagnostics, self-calibration and self-configuration/parameterization. Intelligent field devices, digital field networks, Internet Protocol (IP)-enabled connectivity and web services, historians, and advanced data analysis software are providing the basis for the future project “Industrie 4.0” and Industrial Internet of Things (IIoT).
Important smart features include connectivity and communication ability according to a unified protocol (OPC-UA currently most widely discussed), maintenance and operating functions, traceability and compliance, virtual description to support a continuous engineering, and well as interaction capabilities between sensors. This is a prerequisite for the realization of Cyber Physical Systems (CPS) within these future automation concepts for the process industry. Therefore, smart process sensors enable new business models for users, device manufacturers, and service providers.
The departure from current automation to smart sensor has already begun. Further development is based on the actual situation over several steps. Possible perspectives will be via additional communication channels to mobile devices, bidirectional communication, integration of the cloud and virtualization. The integration of virtual runtime environments can provide a more flexible topology for process control environments.
The talk summarizes the currently discussed requirements to process sensors 4.0 and introduces an online NMR sensor as an example, which was developed in the EU project CONSENS.
The talk reflects how PAT could be applied in future developments ofpharma manufacturing. It shows the benefits, increase quality, size and increasing speed of production with significant reduction of quality costs, which are possible. Using Smart Sensors and model based data evaluation methods are the key to reduce set-up times and costs. Industry 4.0 will help shape the Pharmaceutical industry of tomorrow. This is demonstrated by an example using modular production units for Continuous Manufacturing. The development of a smart online NMR analyser is shown.
The departure from the current automation landscape to next generation automation concepts for the process industry has already begun. Smart functions of sensors simplify their use and enable plug-and-play integration, even though they may appear to be more complex at first sight. Smart sensors enable concepts like self-diagnostics, self-calibration, and self-configuration/parameterization whenever our current automation landscape allows it. Here we summarize the currently discussed general requirements for process sensors 4.0 and introduce a smart online NMR sensor module as example, which was developed for an intensified industrial process funded by the EU’s Horizon 2020 research and innovation programme (www.consens-spire.eu).
Currently, research in chemical manufacturing moves towards flexible plug-and-play approaches focusing on modular plants, capable of producing small scales on-demand with short down-times between individual campaigns. This approach allows for efficient use of hardware, a faster optimization of the process conditions, and thus, an accelerated introduction of new products to the market. Driven mostly by the search for chemical syntheses under biocompatible conditions, so-called “click” chemistry rapidly became a growing field of research. The resulting simple one-pot reactions are so far only scarcely accompanied by an adequate optimization via comparably straightforward and robust analysis techniques.
Here we report on a fast and reliable calibration-free online high field NMR monitoring approach for technical mixtures. It combines a versatile fluidic system, continuous-flow measurement with a time interval of 20 s per spectrum, and a robust, automated algorithm to interpret the obtained data. All spectra were acquired using a 500 MHz NMR spectrometer (Varian) with a dual band flow probe having a 1/16-inch polymer tubing working as a flow cell. Single scan 1H NMR spectra were recorded with an acquisition time of 5 s, relaxation delay of 15 s. As a proof-of-concept, the thiol-ene coupling between N-boc cysteine methyl ester and allyl alcohol was conducted in non-deuterated solvents while its time-resolved behaviour was characterised with step tracer experiments.
Through the application of spectral modeling the signal area for each reactant can be deconvoluted in the online spectra and thus converted to the respective concentrations or molar ratios. The signals which were suitable for direct integration were used herein for comparison purposes of both methods.