The key comparison CCQM-K102: Polybrominated diphenyl ethers in sediment was coordinated by the JRC, Directorate F - Health, Consumers & Reference Materials, Geel (Belgium) under the auspices of the Organic Analysis Working Group (OAWG) of the Comité Consultatif pour la Quantité de Matière (CCQM). Thirteen National Metrology institutes or Designated Institutes and the JRC participated. Participants were requested to report the mass fraction (on a dry mass basis) of BDE 47, 99 and 153 in the freshwater sediment study material. The sediment originated from a river in Belgium and contained PBDEs (and other pollutants) at levels commonly found in environmental samples. The comparison was designed to demonstrate participants' capability of analysing non-polar organic molecules in abiotic dried matrices (approximate range of molecular weights: 100 to 800 g/mol, polarity corresponding to pKow < −2, range of mass fraction: 1–1000 μg/kg). All participants (except one using ultrasonic extraction) applied Pressurised Liquid Extraction or Soxhlet, while the instrumental analysis was performed with GC-MS/MS, GC-MS or GC-HRMS. Isotope Dilution Mass Spectrometry approach was used for quantification (except in one case). The assigned Key Comparison Reference Values (KCRVs) were the medians of thirteen results for BDE 47 and eleven results for BDE 99 and 153, respectively. BDE 47 was assigned a KCRV of 15.60 μg/kg with a combined standard uncertainty of 0.41 μg/kg, BDE 99 was assigned a KCRV of 33.69 μg/kg with a combined standard uncertainty of 0.81 μg/kg and BDE 153 was assigned a KCRV of 6.28 μg/kg with a combined standard uncertainty of 0.28 μg/kg. The k-factor for the estimation of the expanded uncertainty of the KCRVs was chosen as k = 2.
This key comparison aims to assess the capabilities of the participants to determine the amount-of-substance fraction oxygen in nitrogen. The GAWG has classified this as a track B comparison, due to the unexpected 50 μmol/mol argon mole fraction content of the transfer standards, which effects the achievable performance of some measurement techniques such a GC-TCD. The separation of oxygen and argon is challenging, and not all systems in use are equally well designed for it. As this analytical challenge due to a substantial fraction of argon in the transfer standards became a reality, the Gas Analysis Working Group (GAWG) decided to qualify this key comparison as a regular key comparison and not as a core comparison, which may be used to support calibration and measurement capabilities (CMCs) for oxygen in nitrogen, or for oxygen in nitrogen mixtures containing argon only (see also the section on support to CMCs).
The relevance of the COOMET.QM-S5 comparison is founded on tightening of requirements to a control of automobile emissions (realization of environmental standards of EURO 4, EURO-5). Participating laboratories: VNIIM, BAM, BelGIM, Ukrmetrteststandart, KazInMetr. This comparison was carried out in 2013-2014. This supplementary comparison supports CMC claims for: CO2 in the range (4–16) 10⁻² mol/mol; CO in the range (0.5–5) 10⁻² mol/mol; C3H8 in the range (0.01–0.3) 10⁻² mol/mol. Results:
Component CO:
All laboratories identified the values of carbon monoxide mole fraction in the gas mixture within ±0.9134%.
Component CO2:
All laboratories identified the values of carbon dioxide mole fraction in the gas mixture within ±0.3042%.
Component C3H8:
All laboratories identified the values of propane mole fraction in the gas mixture within ±0.443%.
There is a high international priority attached to activities which reduce NOx in the atmosphere. The current level of permitted emissions is typically between 50 µmol/mol and 100 µmol/mol, but lower values are expected in the future. Currently, ambient air quality monitoring regulations also require the measurement of NOx mole fractions as low as 0.2 µmol/mol. The production of accurate standards at these levels of mole fractions requires either dilution of a stable higher concentration gas standard or production by a dynamic technique, for example one based on permeation tubes.
The CCQM-K74 key comparison was designed to evaluate the level of comparability of National Metrology Institutes' measurement capabilities and standards for nitrogen dioxide (NO2) at a nominal mole fraction of 10 µmol/mol.
The measurements of this key comparison took place from June 2009 to May 2010.
Seventeen laboratories took part in this comparison coordinated by the BIPM and VSL. The key comparison reference value was based on BIPM measurement results, and the standard measurement uncertainty of the reference value was 0.042 µmol/mol.
This key comparison demonstrated that the results of the majority of the participants agreed within limits of ±3% relative to the reference value. The results of only one laboratory lay significantly outside these limits. Likewise this comparison made clear that a full interpretation of the results of the comparison needed to take into account the presence of nitric acid (in the range 100 nmol/mol to 350 nmol/mol) in the cylinders circulated as part of the comparison, as well as the possible presence of nitric acid in the primary standards used by participating laboratories.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
DatesIssue 1A (Technical Supplement 2012)
This pilot study compares the performance of participants in analyzing gas mixtures of nitrogen dioxide in nitrogen by comparison with in-house gravimetric standards using Fourier transformed infrared spectroscopy (FT-IR). In this study the same gas mixtures were used as in the key comparison CCQM-K74, which was designed to evaluate the level of comparability of National Metrology Institutes' measurement capabilities for nitrogen dioxide (NO2) at a nominal mole fraction of 10 µmol/mol. In the comparison CCQM-K74 most of the participants used chemiluminescence, with a small number using UV absorption or FT-IR spectroscopy, and thus it is of interest to improve understanding of the comparative performance of these techniques because they do not exhibit any cross-sensitivity to nitric acid (HNO3), which was known to be present in the mixtures used for the comparison.
The results of this pilot study indicate good consistency and a level of agreement similar to that reported in the comparison CCQM-K74, demonstrating that FT-IR can be operated as a comparison method when calibrated with appropriate gas standards and can achieve similar measurement uncertainties to chemiluminescence and UV absorption techniques.
An additional pilot study, CCQM-P110-B2, was conducted on the same gas mixtures in parallel with this pilot study. The second study addressed FT-IR spectroscopy when used to measure the gas mixtures with respect to reference spectra. The results of this second study will be reported elsewhere.
Main text. To reach the main text of this paper, click on Final Report.
The final report has been peer-reviewed and approved for publication by the CCQM-GAWG.
DatesIssue 1A (Technical Supplement 2012)
This key comparison was performed to demonstrate the capability of NMIs to analyse the purity of methane for use as a source gas in the preparation of standard gas mixtures. This capability is an essential requirement for the preparation of accurate standards of natural gas and some other fuels.
Since it is difficult to carry out a comparison with individual samples of pure gas, the sample for this comparison was a synthetic mixture of high purity methane with selected added impurities of nitrogen, argon, carbon dioxide and ethane. These mixtures were prepared by a gas company as a batch of 10 cylinders and their homogeneity and stability were evaluated by NMIJ.
The KCRVs for the four different analytes in this key comparison are based on a consensus of values reported by participants. The uncertainties in the degrees of equivalence were calculated by combining the reported uncertainties with the homogeneity of the samples and the uncertainty of the KCRV. The results submitted are generally consistent with the KCRV within the estimated uncertainties.
Finally, this comparison demonstrates that the analysis of nitrogen, argon, carbon dioxide and ethane in methane at amount fractions of 1 µmol/mol to 5 µmol/mol is generally possible with an uncertainty of 5% to 10%.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
DatesIssue 1A (Technical Supplement 2012)
Refinery gas is a complex mixture of hydrocarbons and non-combustible gases (e.g., carbon monoxide, carbon dioxide, nitrogen, helium). It is obtained as part of the refining and conversion of crude oil. This key comparison aims to evaluate the measurement capabilities for these types of mixtures. The results of the key comparison indicate that the analysis of a refinery-type gas mixture is for some laboratories a challenge. Overall, four laboratories (VSL, NIM, NPL and VNIIM) have satisfactory results. The results of some participants highlight some non-trivial issues, such as appropriate separation between saturated and unsaturated hydrocarbons, and issues with the measurement of nitrogen, hydrogen and helium.
Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.
The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).
DatesIssue 1A (Technical Supplement 2012)
Within the National Metrology Institutes (NMIs) and designated laboratories, an interlaboratory comparison, CCQM-P107, was conducted to verify the degree of international comparability concerning the results of purity analysis. The mass fractions of Ag, Bi, Cd, Cr, Ni, Tl at the lower mg/kg-level in a high purity zinc material were determined, but the real measurand in metrological sense was the sum of the six mass fractions. Homogeneity was investigated by glow discharge mass spectrometry, reference values were obtained using isotope dilution mass spectrometry. Six NMIs participated, contributing eight independent data sets. The agreement amongst the results of the participants, their median and the agreement with the reference values were usually excellent and in almost all cases below the target uncertainty of 30% relative. In this manner, the accuracy of results and the comparability between the participants was demonstrated to be established.