Determination of Radon Exhalation Rates from Construction Materials using Emission Test Chambers
(2015)
Determination of Radon Exhalation Rates from Construction Materials using Emission Test Chambers
(2015)
The inhalation of radon (222Rn) decay products is the leading cause of lung cancer apart from tobacco smoking. Besides the permeation of radon from the subsoil through the basement as main source of radon in indoor air, also building materials have to be taken into consideration, especially at low air change rates in buildings. The Construction Products Regulation (EC, 2010) gives essential requirements for construction works regarding the release of dangerous substances such as toxic gases and radiation to which radon can be assigned. The recently adopted Basic Safety Standards Directive (EC, 2013), which has to be ratified by each European member state in between the next three years sets reference levels for indoor radon concentrations for the first time. In research project financed by the German Ministry for the Environment, Nature Conservation, Building and Nuclear Safety a practical, reliable and easily applicable test procedure for the determination of radon exhalation from building materials – based on ISO 16000-9 (ISO, 2006) and CEN/TS 16516 (CEN, 2013) – should be developed. In contrast to the static test procedure published by Richter et al. (2013), dynamically operated test chambers shall be foregrounded. First results of this study are presented, focused on the reliable measurement of radon background concentration and the selection of suitable radon test devices, representing the basic elements of the subsequent work.
Compared to the designated European upper reference level of 300 Bq/m³ for indoor radon concentrations, the contribution of building materials appears to be low. Taking into account the recommended limit of 100 Bq/m³ by WHO as well as the increased dose conversion factor, the contribution of building products is quite relevant, especially at low air change rates. Therefore in Germany a maximum contribution of building materials to the indoor radon concentration of 20 Bq/m³ is discussed.
Because measurements of low concentrations are associated with high uncertainties, radon exhalation rates are often determined in chambers without air exchange and indoor conditions are extrapolated.
This study investigated the suitability of direct low level radon measurement under simulated indoor conditions with commercially available instruments, based on the preliminary EU standard for the determination of the emission of volatile organic compounds (VOC) from building materials in test chambers.
Measurements are possible with a relative uncertainty ranging from 7 to 18%. Four major problems were identified:
1. Detector sensitivity: Two of four instruments were identified to determine low radon concentrations in a sufficiently short measuring time.
2. Thoron discrimination: Thoron was discriminated by aging the sampled air in a hose before the detector.
3. Sampling setup: An external chamber measurement was feasible as long as all main system parts were under slight overpressure.
4. Blank value: For VOC-emission testing ambient air, conditioned by filtration and sorption, is used to generate an air exchange. Regarding radon, this treatment is non-effective. If all instrument background levels have been characterized, the blank value must be measured with a second detector in parallel.
Compared to the designated European upper reference level of 300 Bq m-3 for indoor radon concentrations, the contribution of building materials appears to be low. Taking into account the recommended limit of 100 Bq m-3 by WHO as well as the increased dose conversion factor, the contribution of building products is quite relevant, especially at low air change rates. Therefore in Germany a maximum contribution of building materials to the indoor radon concentration of 20 Bq m-3 is discussed.
Because measurements of low concentrations are associated with high uncertainties, radon exhalation rates are often determined in chambers without air exchange and indoor conditions are extrapolated.
This study investigated the suitability of direct low level radon measurement under simulated indoor conditions with commercially available instruments, based on the preliminary EU standard for the determination of the emission of volatile organic compounds (VOC) from building materials in test chambers.
Measurements are possible with a relative uncertainty ranging from 7 to 18 %. Four major problems were identified:
1. Detector sensitivity: Two of four instruments were identified to determine low radon concentrations in a sufficiently short measuring time.
2. Thoron discrimination: Thoron was discriminated by aging the sampled air in a hose before the detector.
3. Sampling setup: An external chamber measurement was feasible as long as all main system parts were under slight overpressure.
4. Blank value: For VOC-emission testing ambient air, conditioned by filtration and sorption, is used to generate an air exchange. Regarding radon, this treatment is non-effective. If all instrument background levels have been characterized, the blank value must be measured with a second detector in parallel.
Prüfverfahren zur Bestimmung der Radon-Exhalationsrate aus Bauprodukten in VOC-Emissionskammern
(2017)
Es wurden die wesentlichen Ergebnisse des abgeschlossenen UFOPLAN-Vorschungsvorhabens "Methodische Untersuchungen zur Qualifizierung der VOC-Kammern zur Bestimmung der Radonexhalationsrate von Bauprodukten, Los 1" vorgestellt.
Das entwickelte Prüfverfahren eignet sich zur Bestimmung der Radonexhalationsrate von Bauprodukten mittels kommerziell erhältlicher Radon-Messtechnik in dynamisch betriebenen VOC-Emissionsprüfkammern. Es ist robust gegenüber einem weiten Bereich von Parametereinstellungen hinsichtlich der Prüfparameter Prüfkammervolumen, Produktbeladung, Luftwechselrate und flächenspezifische Luftdurchflussrate. Entscheidend für die Messung im Bereich geringer Radonaktivitätskonzentration ist es, die Zuluftkonzentration (Blindwert) der Emissionsprüfkammern auf einem konstant geringen Niveau zu halten oder durch eine – zur eigentlichen Baustoffmessung – parallele Messung der Zuluft zu kompensieren. Die gängige Aufbereitung der Zuluft der VOC-Prüfkammern mittels mehrfacher Filtration und Sorption über Silica-Gel- sowie verhältnismäßig kleinvolumigen Aktivkohle-Filtern ist bezüglich Radon unwirksam.
Compared to the intended EU reference level of 300 Bq m−3 for indoor radon concentrations, the contribution of building materials appears to be low. Considering the recommended limit of 100 Bq m−3 by WHO, their contribution is supposed to be relevant, especially at low air exchange rates. This study as part of a two-part research project investigated the suitability of direct low level 222Rn measurement under simulated indoor conditions with commercial radon monitors and dynamically operated emission test chambers. Active measuring devices based on ionisation or scintillation chambers with 1-σ uncertainties below 8.6% at 20 Bq m−3 were found to be best suitable for a practical test procedure for the determination of radon exhalation rates of building materials. For the measurement of such low concentrations, the knowledge of the accurate device background level is essential.
The presentation gives a brief overview of the main goals, the used methods and the first results of the PMARS OdourSense project. In cooperation with the Moulay Ismaïl University in Meknès (Morocco), a multi-gas sensor system for odour nuisance monitoring in outdoor air and a method for testing/calibration of such devices are being developed in this project.
Der Beitrag präsentiert die erarbeiteten technischen Richtlinien zur Anwendung und Prüfung von Multigassensoren innerhalb der Richtlinienreihe VDI/VDE 3518 sowie in diesem Zusammenhang die Möglichkeiten der Sensorprüfung im akkreditierten Prüflabor des Fachbereichs 8.1 der BAM. Im Speziellen wird das in Kürze veröffentlichte Blatt 3 der Richtlinienreihe vorgestellt, das Bezug auf Multigassensoren für geruchsbezogene Messungen mit elektronischen Nasen nimmt.
Der Beitrag präsentiert die erarbeiteten technischen Richtlinien zur Anwendung und Prüfung von Multigassensoren innerhalb der Richtlinienreihe VDI/VDE 3518 sowie in diesem Zusammenhang die Möglichkeiten der Sensorprüfung im akkreditierten Prüflabor des Fachbereichs 8.1 der BAM. Im Speziellen wird das in Kürze veröffentlichte Blatt 3 der Richtlinienreihe vorgestellt, das Bezug auf Multigassensoren für geruchsbezogene Messungen mit elektronischen Nasen nimmt.