In all fluorescence-based techniques, the measured signals contain not only sample-related but also instrument-specific contributions, which limit the direct comparison of fluorescence data obtained e.g. on different devices or at different times and often hamper quantification. To rule out instrumentation as major source of variability of emission data, accepted fluorescence standards and procedures for the control of instrument specifications and long-term performance are required. For flow cytometry (FCM), a broad variety of fluorophore-stained polymer beads differing in emission wavelength and intensity is available for the testing of the alignment, sensitivity, and other parameters of FCM. These calibration tools are intended to facilitate the assessment of instrument performance to ensure reliable measurements and to improve the comparability of FCM experiments.
As a step towards an improved comparability of fluorescence data, with special emphasis on spectroscopic methods measuring nano- and micrometer-sized fluorescent objects, we are currently developing a set of fluorescent polystyrene (PS) beads loaded with luminophores from the certified BAM-Kit “Spectral fluorescent standards”, initially developed for the calibration of fluorescence spectrometers. Here, we present first results from studies of these fluorophore-loaded polymer beads. Moreover, new beads are made to supplement this kit by encapsulating near-infrared (NIR)-emissive luminophores in PS beads to cover the UV/VIS, and NIR wavelength range.
These beads are designed for calibration of flow cytometers and other fluorescence imaging systems to meet the increasing demand for reliable and comparable fluorescence data especially in strongly regulated areas like e.g. medical diagnostics.
There is an increasing interest in bridging the gap between the photoluminescence (PL) properties of nanomaterials like semiconductor nanocrystals (QDs) commonly assessed in ensemble studies and the PL features of single QDs for life sciences applications such as bioimaging studies or use in microfluidic assays. The fluorescence quantum yield (ΦF) is a key performance parameter for all molecular and nanoscale emitters, increasingly employed in nanoscience, nanotechnology, and medical diagnostics.
ΦF determines not only the signal size together with the reporter´s molar extinction coefficient, yet it is particularly relevant for nanocrystals like QDs with coordinatively bound surface ligands and size- and surface chemistry-dependent PL characteristics.
The importance of ΦF measurements at ultralow concentration encouraged us to explore the potential of fluorescence correlation spectroscopy (FCS) for the relative determination of ΦF of ligand-stabilized CdTe nanocrystals in comparison to molecular dyes with closely matching spectral properties and known ΦF.
We describe a FCS-based method for the relative determination of ΦF of dispersed QDs at ultralow concentrations, and procedures to overcome QD-inherent challenges like complex and power-dependent blinking behavior as well as ligand- and QD-specific aggregation. We could demonstrate the potential of this approach by comparison with steady state ensemble measurements.
Time-resolved flow cytometry
(2019)
The fast identification of a large number of analytes or events is increasingly required in bioanalytical, diagnostic, and security applications. The versatility and straightforward use make multiparametric fluorescence techniques particularly interesting as detection techniques. An established method for high-throughput single-cell and single-particle measurements is flow cytometry (FCM). Using only spectral encoding without further intensity information, state-of-the-art instruments equipped with several light sources and detectors can resolve almost 20 different color codes. However, this is not sufficient to answer complex research questions, e.g. in cell biology and immunology. In contrast, routine applications demand low-cost and sometimes even portable instruments and thus a minimum number of instrument components. Thus, there are currently two main research directions in FCM: the development of methods that can either address increasingly complex analytical challenges or provide low-cost and robust approaches for routine multiplex analyses. Common spectral multiplexing approaches face limitations in both directions. On the one hand, spectral overlap of labels restricts the number of codes and makes elaborate correction schemes necessary. On the other hand, even for lower degrees of multiplexing often a sophisticated optical setup is needed. An alternative to spectral multiplexing and intensity encoding is to exploit the luminescence lifetime (LT) as an encoding parameter. This can allow for extending the parameter space in combination with spectral encoding or result in more simple and compact devices due to fewer optical components. The availability of fast electronics enables miniaturized and portable lifetime measurement setups at relatively low cost. LT-FCM requires to master LT determination with a limited number of detected photons due to the short interaction time of the encoded objects with the laser spot. In this study, we address this issue for time-domain cytometry and present a novel lifetime flow cytometry (LT-FCM) platform based on a compact setup and straightforward time-domain measurements utilizing LT-encoded luminescent beads. Moreover, we present the realization of a first bioanalytical assay with LT-encoded beads.
The overall interest in nanotoxicity, triggered by the increasing use of nanomaterials in the material and life sciences, and the synthesis of an ever increasing number of new functional nanoparticles calls for standardized test procedures and for efficient approaches to screen the potential genotoxicity of these materials. Aiming at the development of fast and easy to use, automated microscopic methods for the determination of the genotoxicity of different types of nanoparticles, we assess the potential of the fluorometric γH2AX assay for this purpose. This assay, which can be run on an automated microscopic detection system, relies on the detection of DNA double strand breaks (DSB) as a sign for genotoxicity. Here, we provide first results obtained with broadly used nanomaterials like CdSe/CdS and InP/ZnS quantum dots as well as iron oxide, gold, and polymer particles of different surface chemistry with previously tested colloidal stability and different cell lines like Hep-2 and 8E11 cells, which reveal a dependence of the genotoxicity on the chemical composition as well as the surface chemistry of these nanomaterials. These studies will be also used to establish nanomaterials as positive and negative genotoxicity controls or standards for assay performance validation for users of this fluorometric genotoxicity assay. In the future, after proper validation, this microscopic platform technology will be expanded to other typical toxicity assays.
Luminescence techniques are amongst the most commonly used analytical methods in life and material sciences due to their high sensitivity, nondestructive character, and easy instrumentation suitable for miniaturization. Photoluminescence signals are, however, affected by wavelength-, polarization-, and time-dependent instrument-related effects. Thus, at the core of standardization approaches for all fluorescence-based techniques are evaluated fluorescence standards for the consideration of instrument-specific spectral and intensity distortions of measured signals and for instrument performance validation (IPV).
Here, we summarize the research of BAM division 1.2 on development of liquid and solid fluorescence standards for various application-relevant fluorescence parameters and techniques. The portfolio of BAM fluorescence reference materials presently consists of:
i) a Spectral Fluorescence Standard Kit, i.e., a set of liquid fluorescence standards with certified normalized corrected emission spectra, for the determination of a broad variety of fluorescence parameters
ii) a ready-to-use, glass-based multi-emitter fluorescence standard for IPV and the determination of instrument-to-instrument variations
iii) specially adapted calibration tools and validation concepts for microarray-based platforms used in molecular diagnostics and food safety control
iv) quantum yield (Φf) standards for relative determination of the key performance parameter Φf of fluorescent materials, which can be also used for the evaluation of the performance of absolute, standard-free methods utilizing integrating sphere setups or spectrometer accessories. These materials, that will eventually cover the ultraviolet, visible, and near infrared spectral region, are currently under certification.
This toolbox of method-adapted reference materials can perfectly complement existing fluorescence standards. These easy-to-use, reference materials can pave the way to traceable fluorescence measurements to a radiometric scale like the spectral radiance or spectral photon radiance for all users of fluorescence techniques in material sciences and analysis, as well as environmental monitoring and biotechnology. They are particularly useful for customers working in strongly regulated areas like medical diagnostics or pharmaceutical research, where certified standards in conjunction with validated standard operating procedures are mandatory.
Fluorescence microscopic and spectroscopic monitoring of degradation processes upon polymer ageing
(2017)
The majority of all routinely used methods to assess polymer aging are based on destructive tests and methods. Early indicators for the deterioration of polymer materials are e.g., physical or mechanical properties like tensile strength, adhesion, brittleness, and color. It is well-known, however, that predominantly chemical changes are the underlying process of the physical changes that occur in organic materials upon aging over time. Typical initial steps during polymer degradation are crosslinking or chain breaking, alteration of autofluorescence, “yellowing” or bleaching caused by the formation of new functional groups. A straightforward strategy towards the sensitive detection and monitoring of chemical changes in the course of polymer aging is based on non-destructive optical measurements. Luminescence techniques, one of the most sensitive spectroscopic methods are the method of choice. Here, we present first results of luminescence-based monitoring of polymer degradation induced by different environmentally relevant weathering factors (e.g. humidity and UV exposure). Our studies include fluorescence spectroscopy as well as spectral scanning confocal fluorescence microscopy and clearly demonstrate the possibility to follow accelerate-aging processes by luminescence detection.
Luminescence techniques are amongst the most commonly used analytical methods in the life and the material sciences due to their sensitivity and nondestructive character. All photoluminescence signals are, however, affected by wavelength-, polarization- and time-dependent instrument-related effects. Furthermore, substantial challenges to measure absolute luminescence intensities complicate the comparison of data recorded with different instruments and on the same instrument at different times. These problems can be easily resolved with fluorescence standards used for instrument performance validation (IPV) and determination of instrument-to-instrument variations, which allow to measure, quantify, and monitor the wavelength-dependent spectral responsivity for typically used instrument settings. For example, a set of liquid fluorescence standards, the BAM Kit F001-F005, and a ready-to-use glass-based fluorescence standard BAM F-012 developed and certified by BAM enable the characterization of many fluorescence parameters in the UV/vis wavelength range. For the increasingly used near infrared (NIR) region, standards and calibration tools are still very rare. Reliable spectral fluorescence standards and intensity or quantum yield standards are currently not available for the NIR, even though in biology, molecular imaging, and clinical diagnostics fluorescence labels absorbing and emitting in the long wavelength region beyond 650 nm are being increasingly used.
This limitation hampers the reliability and comparability of fluorescence measurements in the NIR and calls for simple fluorescence standards for instrument characterization and for the quantification of fluorescence intensities and efficiencies to improve the comparability of the emission measurements in the NIR. This encouraged us to assess the potential of several NIR-emitting materials as spectral fluorescence standards, thereby extending the BAM Kit from the UV/vis into the NIR up to 950 nm. Moreover, we currently certify quantum yield standards for the UV/vis/NIR to improve the reliability of relative measurements of this spectroscopic key quantity particularly > 650 nm. These tools enable an instrument characterization, signal referencing, quality assurance, traceability, and method validation now also for wavelengths > 650 nm, thereby improving the reliability of fluorescence data in pharmaceutical research, medical and clinical diagnostics, material analysis, and environmental monitoring.
Here, we summerize our efforts concerning new design concepts and examples for fluorescence standards that can provide traceability to radiometric units and present a first step towards a toolbox of fluorescence standards, currently consisting of:
i) A first set of liquid fluorescence standards enables the determination of a broad variety of fluorescence parameters was developed and certified by BAM and is distributed by Sigma-Aldrich.
ii) Ready-to-use, glass-based fluorescence standards for instrument performance validation (IPV) and determination instrument-to-instrument variations can also be used as wavelength standard for fluorescence instruments with low requirements on spectral resolution and allow monitoring of temporal changes of the wavelength-dependent spectral responsivity.
iii) Novel calibration tools and validation concepts for microarray-based platforms used in molecular diagnostics and food safety control.
iv) We currently develop reference materials, which can be used as reliable quantum yield standards for relative methods for the determination of QY and can be valuable in the evaluation of the performance and sources of uncertainty of absolute, standard-free methods using e.g. integrating spheres.