Filtern
Dokumenttyp
- Zeitschriftenartikel (11)
- Beitrag zu einem Tagungsband (5)
- Vortrag (4)
- Beitrag zu einem Sammelband (2)
- Posterpräsentation (2)
Sprache
- Englisch (13)
- Deutsch (8)
- Mongolisch (3)
Schlagworte
- Degassing (2)
- Diffusion coefficient (2)
- Hydrogen (2)
- Water (2)
- Aluminoborosilicate glass (1)
- Bulk diffusion (1)
- Carrier gas analysis (1)
- Crystallization (1)
- Debindering (1)
- Diffusion (1)
Eingeladener Vortrag
- nein (4)
Wasserdiffusion in Floatglas
(2005)
Thermo Analytic Investigation of Hydrogen Effusion Behavior - Sensor Evaluation and Calibration
(2011)
The well established carrier gas analysis (CGA) method was used to test different hydrogen detectors comprising a thermal conductivity detector (TCD) and a metal oxide semiconducting (MOx) sensor. The MOx sensor provides high hydrogen sensitivity and selectivity, whereas the TCD exhibits a much shorter response time and a linear hydrogen concentration dependency. Therefore, the TCD was used for quantitative hydrogen concentration measurements above 50?µmol/mol. The respective calibration was made using N2/H2 gas mixtures. Furthermore, the hydrogen content and degassing behaviour of titanium hydride (TiH2-x) was studied. This material turned out to be a potential candidate for a solid sample calibration. Vacuum hot extraction (VHE) coupled with a mass spectrometer (MS) was then calibrated with TiH2-x as transfer standard. The calibration was applied for the evaluation of the hydrogen content of austenitic steel samples (1.4301) and the comparison of CGA-TCD and VHE-MS.