Filtern
Erscheinungsjahr
- 2008 (1)
Dokumenttyp
Sprache
- Englisch (1)
Referierte Publikation
- ja (1)
Schlagworte
- Phosphorus (1) (entfernen)
C, N, O, F and P can be analyzed by instrumental photon activation analysis (IPAA) including decay curve analysis. The interference of 30P (T1/2 = 149.9 s) by 15O (T1/2 = 122.2 s) can be ruled out by direct positron measurement making use of the largely different maximum β+-energies of both nuclides (3.24 MeV and 1.73 MeV, respectively). Interference by carbon (11C) can be avoided by sub-threshold activation with 17 MeV bremsstrahlung. The short half-life of 30P allows a high productivity of the method. Reliability was demonstrated in the range of 0.2%2% P (detection limit = 40 µg/g). Analysis of a certified reference material (BCR-CRM 063) yielded excellent agreement with the certified data.